Physics > Medical Physics
[Submitted on 7 Feb 2022]
Title:Free-breathing motion compensated 4D (3D+respiration) T2-weighted turbo spin-echo MRI for body imaging
View PDFAbstract:Purpose: To develop and evaluate a free-breathing respiratory motion compensated 4D (3D+respiration) $T_2$-weighted turbo spin echo sequence with application to radiology and MR-guided radiotherapy.
Methods: k-space data are continuously acquired using a rewound Cartesian acquisition with spiral profile ordering (rCASPR) to provide matching contrast to the conventional linear phase encode ordering and to sort data into multiple respiratory phases. Low-resolution respiratory-correlated 4D images were reconstructed with compressed sensing and used to estimate non-rigid deformation vector fields, which were subsequently used for a motion compensated image reconstruction. rCASPR sampling was compared to linear and CASPR sampling in terms of point-spread-function (PSF) and image contrast with in silico, phantom and in vivo experiments. Reconstruction parameters for low-resolution 4D-MRI (spatial resolution and temporal regularization) were determined using a grid search. The proposed motion compensated rCASPR was evaluated in eight healthy volunteers and compared to free-breathing scans with linear sampling. Image quality was compared based on visual inspection and quantitatively by means of the gradient entropy.
Results: rCASPR provided a superior PSF (similar in ky and narrower in kz) and showed no considerable differences in images contrast compared to linear sampling. The optimal 4D-MRI reconstruction parameters were spatial resolution=$4.5 mm^3$ and $\lambda_t=10^{-4}$. The groupwise average gradient entropy was 22.31 for linear, 22.20 for rCASPR, 22.14 for soft-gated rCASPR and 22.02 for motion compensated rCASPR.
Conclusion: The proposed motion compensated rCASPR enables high quality free-breathing T2-TSE with minimal changes in image contrast and scan time. The proposed method therefore enables direct transfer of clinically used 3D TSE sequences to free-breathing.
Current browse context:
physics.med-ph
References & Citations
Bibliographic and Citation Tools
Bibliographic Explorer (What is the Explorer?)
Litmaps (What is Litmaps?)
scite Smart Citations (What are Smart Citations?)
Code, Data and Media Associated with this Article
CatalyzeX Code Finder for Papers (What is CatalyzeX?)
DagsHub (What is DagsHub?)
Gotit.pub (What is GotitPub?)
Papers with Code (What is Papers with Code?)
ScienceCast (What is ScienceCast?)
Demos
Recommenders and Search Tools
Influence Flower (What are Influence Flowers?)
Connected Papers (What is Connected Papers?)
CORE Recommender (What is CORE?)
arXivLabs: experimental projects with community collaborators
arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.
Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.
Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.