Time-of-Flight X-ray Measurements for Computed Tomography
Authors:
Gaëtan Lemaire,
Romain Espagnet,
Julien Rossignol,
Louis-Daniel Gaulin,
Edouard Villemure,
Audrey Corbeil Therrien,
Yves Bérubé-Lauzière,
Marc-André Tétrault,
Réjean Fontaine
Abstract:
Time-of-flight (ToF) measurements is a possible alternative to anti-scatter grids in computed tomography (CT). Simulations have shown a possible 75% reduction in the detrimental scattering contribution with a 100 ps FWHM timing resolution. A test bench comprising a pulsed X-ray source and a time-correlated detector has been designed to demonstrate the feasibility of time-of-flight measurements of…
▽ More
Time-of-flight (ToF) measurements is a possible alternative to anti-scatter grids in computed tomography (CT). Simulations have shown a possible 75% reduction in the detrimental scattering contribution with a 100 ps FWHM timing resolution. A test bench comprising a pulsed X-ray source and a time-correlated detector has been designed to demonstrate the feasibility of time-of-flight measurements of X-rays using readout electronics inherited from a ToF-positron emission tomography project. A 86 ps FWHM coincidence time resolution have been achieved with 511 keV annihilation gamma-rays and a 155 ps FWHM timing jitter with a 120 kVp pulsed X-rays source.
△ Less
Submitted 29 November, 2023;
originally announced November 2023.
-
Simulation Study of Photon-to-Digital Converter (PDC) Timing Specifications for LoLX Experiment
Authors:
Nguyen V. H. Viet,
Alaa Al Masri,
Masaharu Nomachi,
Marc-Andre Tétrault,
Soud Al Kharusi,
Thomas Brunner,
Christopher Chambers,
Bindiya Chana,
Austin de St. Croix,
Eamon Egan,
Marco Francesconi,
David Gallacher,
Luca Galli,
Pietro Giampa,
Damian Goeldi,
Jessee Lefebvre,
Chloe Malbrunot,
Peter Margetak,
Juliette Martin,
Thomas McElroy,
Mayur Patel,
Bernadette Rebeiro,
Fabrice Retiere,
El Mehdi Rtimi,
Lisa Rudolph
, et al. (2 additional authors not shown)
Abstract:
The Light only Liquid Xenon (LoLX) experiment is a prototype detector aimed to study liquid xenon (LXe) light properties and various photodetection technologies. LoLX is also aimed to quantify LXe's time resolution as a potential scintillator for 10~ps time-of-flight (TOF) PET. Another key goal of LoLX is to perform a time-based separation of Cerenkov and scintillation photons for new background r…
▽ More
The Light only Liquid Xenon (LoLX) experiment is a prototype detector aimed to study liquid xenon (LXe) light properties and various photodetection technologies. LoLX is also aimed to quantify LXe's time resolution as a potential scintillator for 10~ps time-of-flight (TOF) PET. Another key goal of LoLX is to perform a time-based separation of Cerenkov and scintillation photons for new background rejection methods in LXe experiments. To achieve this separation, LoLX is set to be equipped with photon-to-digital converters (PDCs), a photosensor type that provides a timestamp for each observed photon. To guide the PDC design, we explore requirements for time-based Cerenkov separation. We use a PDC simulator, whose input is the light information from the Geant4-based LoLX simulation model, and evaluate the separation quality against time-to-digital converter (TDC) parameters. Simulation results with TDC parameters offer possible configurations supporting a good separation. Compared with the current filter-based approach, simulations show Cerenkov separation level increases from 54% to 71% when using PDC and time-based separation. With the current photon time profile of LoLX simulation, the results also show 71% separation is achievable with just 4 TDCs per PDC. These simulation results will lead to a specification guide for the PDC as well as expected results to compare against future PDC-based experimental measurements. In the longer term, the overall LoLX results will assist large LXe-based experiments and motivate the assembly of a LXe-based TOF-PET demonstrator system.
△ Less
Submitted 28 October, 2023;
originally announced October 2023.
-
Real-Time Discrete SPAD Array Readout Architecture for Time of Flight PET
Authors:
M. -A. Tétrault,
É. Desaulniers Lamy,
A. Boisvert,
C. Thibaudeau,
F. Dubois,
R. Fontaine,
J. -F. Pratte
Abstract:
Single photon avalanche diode (SPAD) arrays have proven themselves as serious candidates for time of flight positron emission tomography (PET). Discrete readout schemes mitigate the low-noise requirements of analog schemes and offer very fine control over threshold levels and timing pickup strategies. A high optical fill factor is paramount to timing performance in such detectors, and consequently…
▽ More
Single photon avalanche diode (SPAD) arrays have proven themselves as serious candidates for time of flight positron emission tomography (PET). Discrete readout schemes mitigate the low-noise requirements of analog schemes and offer very fine control over threshold levels and timing pickup strategies. A high optical fill factor is paramount to timing performance in such detectors, and consequently space is limited for closely integrated electronics. Nonetheless, a production, daily used PET scanner must minimize bandwidth usage, data volume, data analysis time and power consumption and therefore requires a real-time readout and data processing architecture as close to the detector as possible. We propose a fully digital, embedded real-time readout architecture for SPAD-based detector. The readout circuit is located directly under the SPAD array instead of within or beside it to remove the fill factor versus circuit capabilities tradeoff. The overall real-time engine reduces transmitted data by a factor of 8 in standard operational mode. Combined with small local memory buffers, this significantly reduces overall acquisition dead time. A prototype device featuring individual readout for 6 scintillator channels was fabricated. Timing readout is provided by a first photon discriminator and a 31 ps time to digital discriminator, while energy reading and event packaging is done using standard logic in real-time. The dedicated serial output line supports a sustained rate of 170k counts per second (CPS) in waveform mode, while the standard operational mode supports 2.2 MCPS.
△ Less
Submitted 15 June, 2014;
originally announced June 2014.