3D source tracking and error detection in HDR using two independent scintillator dosimetry systems
Authors:
Haydee M. Linares Rosales,
Jacob G. Johansen,
Gustavo Kertzscher,
Kari Tanderup,
Luc Beaulieu,
Sam Beddar
Abstract:
The high dose gradients near the source characteristics of brachytherapy are equivalent to nefarious effects if unnoticed errors take place during the patient treatment. In vivo dosimetry is the only method to quantify the delivered dose. Previous studies to this one, have characterized potential detectors that can be used as in vivo dosimeter. Some of them have focused on the source tracking topi…
▽ More
The high dose gradients near the source characteristics of brachytherapy are equivalent to nefarious effects if unnoticed errors take place during the patient treatment. In vivo dosimetry is the only method to quantify the delivered dose. Previous studies to this one, have characterized potential detectors that can be used as in vivo dosimeter. Some of them have focused on the source tracking topic in HDR brachytherapy. The aim of this study is to perform 3D source position reconstruction by combining in vivo dosimetry measurements from two independent detector systems. The first was based on multiple (three) plastic scintillator detectors and the second on a single inorganic crystal (CsI:Tl). By combining two detector responses, we enabled the determination of the absolute source coordinates in 3D space. The method in this study proposed can be extended to the combination of different systems.
△ Less
Submitted 21 June, 2020;
originally announced June 2020.
-
From conception to clinical trial: IViST -- the first multi-sensor-based platform for real-time In Vivo dosimetry and Source Tracking in HDR brachytherapy
Authors:
Haydee M. Linares Rosales,
Audrey Cantin,
Sylviane Aubin,
Sam Beddar,
Luc Beaulieu
Abstract:
This study aims to introduce IViST (In Vivo Source Tracking), a novel multi-sensors dosimetry platform for real-time treatment monitoring in HDR brachytherapy. IViST is a platform that comprises 3 parts: 1) an optimized and characterized multi-point plastic scintillator dosimeter (3 points mPSD; using BCF-60, BCF-12, and BCF-10 scintillators), 2) a compact assembly of photomultiplier tubes (PMTs)…
▽ More
This study aims to introduce IViST (In Vivo Source Tracking), a novel multi-sensors dosimetry platform for real-time treatment monitoring in HDR brachytherapy. IViST is a platform that comprises 3 parts: 1) an optimized and characterized multi-point plastic scintillator dosimeter (3 points mPSD; using BCF-60, BCF-12, and BCF-10 scintillators), 2) a compact assembly of photomultiplier tubes (PMTs) coupled to dichroic mirrors and filters for high-sensitivity scintillation light collection, and 3) a Python-based graphical user interface used for system management and signal processing. IViST can simultaneously measure dose, triangulate source position, and measure dwell time. By making 100 000 measurements/s, IViST samples enough data to quickly perform key QA/QC tasks such as identifying wrong individual dwell time or interchanged transfer tubes. By using 3 co-linear sensors and planned information for an implant geometry (from DICOM RT), the platform can also triangulate source position in real-time. A clinical trial is presently on-going using the IViST system.
△ Less
Submitted 14 June, 2020;
originally announced June 2020.
-
On the use of machine learning methods for mPSD calibration in HDR brachytherapy
Authors:
Haydee M. Linares Rosales,
Gabriel Couture,
Louis Archambault,
Sam Beddar,
Philippe Despres,
Luc Beaulieu
Abstract:
Purpose: We sought to evaluate the feasibility of using machine learning algorithms for multipoint plastic scintillator detector calibration in high-dose-rate brachytherapy. Methods: The dosimetry system consisted of an optimized 1-mm-core mPSD and a compact assembly of photomultiplier tubes coupled with dichroic mirrors and filters. An $^{192}$Ir source was remotely controlled and sent to various…
▽ More
Purpose: We sought to evaluate the feasibility of using machine learning algorithms for multipoint plastic scintillator detector calibration in high-dose-rate brachytherapy. Methods: The dosimetry system consisted of an optimized 1-mm-core mPSD and a compact assembly of photomultiplier tubes coupled with dichroic mirrors and filters. An $^{192}$Ir source was remotely controlled and sent to various positions in a homemade PMMA holder. Dose measurements covering a range of 0.5 to 12 cm of source displacement were carried out according to TG-43 recommendations. Individual scintillator doses were decoupled using a linear regression model, a random forest estimator, and artificial neural network algorithms. The performance of the different algorithms was evaluated using different sample sizes and distances to the source for the mPSD system calibration. Results: The decoupling methods' deviations from the expected TG-43 dose generally remained below 20%. However, the dose prediction with the three algorithms was accurate to within 7% relative to the dose predicted by the TG-43 formalism for measurements performed in the same range of distances used for calibration. The performance random forest was compromised when the predictions were done beyond the range of distances used for calibration. The dose prediction by the linear regression was less influenced by the calibration conditions than random forest, but with more significant deviations. The number of available measurements for training purposes influenced the random forest and neural network models the most. Their accuracy tended to converge toward deviation values close to 1% from a number of dwell positions greater than 100. Conclusions: In performing HDR brachytherapy dose measurements with an optimized mPSD system, ML algorithms are good alternatives for precise dose reporting and treatment assessment.
△ Less
Submitted 1 May, 2020;
originally announced May 2020.
-
Dosimetric performance of a multi-point plastic scintillator dosimeter as a tool for real-time source tracking in high dose rate brachytherapy
Authors:
Haydee M. Linares Rosales,
Louis Archambault,
Sam Beddar,
Luc Beaulieu
Abstract:
Purpose: To present the performance of a multi-point plastic scintillation detector (mPSD) as a tool in vivo dosimetry in brachytherapy. Methods: A previously optimized three-point sensor system was used for in vivo HDR brachytherapy measurements (using the scintillators BCF-60, BCF-12, and BCF-10). The light detection system of the mPSD consisted of compactly assembled photomultiplier tubes (PMTs…
▽ More
Purpose: To present the performance of a multi-point plastic scintillation detector (mPSD) as a tool in vivo dosimetry in brachytherapy. Methods: A previously optimized three-point sensor system was used for in vivo HDR brachytherapy measurements (using the scintillators BCF-60, BCF-12, and BCF-10). The light detection system of the mPSD consisted of compactly assembled photomultiplier tubes (PMTs) and dichroic mirrors and filters to achieve a highly sensitive scintillation light collection. The PMT signals were recorded using a NI-DAQ board at a rate of 100 kHz. Dose measurements covering a range of 0.5 to 10 cm from the 192Ir source were carried out according to TG-43 U1 recommendations in order to: (1) characterize the system's response in terms of angular dependence; (2) obtain the relative contribution of positioning and measurement uncertainties to the total system uncertainty; (3) assess the system's temporal resolution; and (4) track the source position in real time. Results: The positioning uncertainty dominated close to the source, whereas the measurement uncertainty dominated at larger distances. A maximum measurement uncertainty of 17 % was observed for the BCF-60 scintillator at 10 cm from the source. The average best compromise between positioning and measurement uncertainties were reached at 17.4 mm. The detector further exhibited no angular dependence. The system provided an average location with a standard deviation under 1.7 mm. The maximum observed differences between measured and expected source location was 1.82 mm. Dose deviations remained below 5% in all the explored measurement conditions. With regard to dwell time measurement accuracy, the maximum deviation observed at all distances was 0.56 s. Conclusions: The performance of the system demonstrated that it could be used for real-time dose, position and dwell time measurements during HDR brachytherapy.
△ Less
Submitted 28 October, 2019;
originally announced October 2019.
-
Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy
Authors:
Haydee M. Linares Rosales,
Patricia Duguay-Drouin,
Louis Archambault,
Sam Beddar,
Luc Beaulieu
Abstract:
Purpose: This study aims to optimize and characterize the response of a mPSD for in vivo dosimetry in HDR brachytherapy. Methods: An exhaustive analysis was carried out in order to obtain an optimized mPSD design that maximize the scintillation light collection produced by the interaction of ionizing photons. Several mPSD prototypes were built and tested in order to determine the appropriate order…
▽ More
Purpose: This study aims to optimize and characterize the response of a mPSD for in vivo dosimetry in HDR brachytherapy. Methods: An exhaustive analysis was carried out in order to obtain an optimized mPSD design that maximize the scintillation light collection produced by the interaction of ionizing photons. Several mPSD prototypes were built and tested in order to determine the appropriate order of scintillators relative to the photodetector, as well as their length as a function of the scintillation light emitted. Scintillators BCF-60, BCF-12 and BCF-10 constituted the mPSD sensitive volume.Each scintillator contribution to the total spectrum was determined by irradiations in the low energy range.For the best mPSD design, a numerical optimization was done in order to select the optical components that better match the light emission profile. The optimized dosimetric system was used for HDR brachytherapy dose determination. The system performance was quantified in term of signal to noise ratio and signal to background ratio. Results: It was determined that BCF-60 should be placed at the distal position, BCF-12 in the center and BCF-10 at proximal position with respect to the photodetector.This configuration allowed for optimized light transmission through the collecting fiber, avoiding inter-scintillator excitation and self-absorption effects.The optimized luminescence system allowed for signal deconvolution using a multispectral approach, extracting the dose to each element while taking into account Cerenkov stem effect.Differences between the mPSD measurements and TG-43 remain below 5%. In all measurement conditions, the system was able to properly differentiate the produced scintillation signal from the background one. Conclusions: A mPSD was constructed and optimized for HDR brachytherapy dosimetry, enabling real time dose determination, up to 6.5cm from the 192Ir source.
△ Less
Submitted 31 May, 2019; v1 submitted 24 July, 2018;
originally announced July 2018.