Non-maximal entanglement of photons from positron-electron annihilation demonstrated using a novel plastic PET scanner
Authors:
P. Moskal,
D. Kumar,
S. Sharma,
E. Y. Beyene,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Das,
K. Dulski,
M. Gorgol,
B. Jasinska,
K. Kacprzak,
T. Kaplanoglu,
L. Kaplon,
K. Klimaszewski,
T. Kozik,
E. Lisowski,
F. Lisowski,
W. Mryka,
S. Niedzwiecki,
S. Parzych,
E. P. del Rio,
L. Raczynski,
M. Radler
, et al. (7 additional authors not shown)
Abstract:
In the state-of-the-art Positron Emission Tomography (PET), information about the polarization of annihilation photons is not available. Current PET systems track molecules labeled with positron-emitting radioisotopes by detecting the propagation direction of two photons from positron-electron annihilation. However, annihilation photons carry more information than just the site where they originat…
▽ More
In the state-of-the-art Positron Emission Tomography (PET), information about the polarization of annihilation photons is not available. Current PET systems track molecules labeled with positron-emitting radioisotopes by detecting the propagation direction of two photons from positron-electron annihilation. However, annihilation photons carry more information than just the site where they originated. Here we present a novel J-PET scanner built from plastic scintillators, in which annihilation photons interact predominantly via the Compton effect, providing information about photon polarization in addition to information on photon direction of propagation. Theoretically, photons from the decay of positronium in a vacuum are maximally entangled in polarization. However, in matter, when the positron from positronium annihilates with the electron bound to the atom, the question arises whether the photons from such annihilation are maximally entangled. In this work, we determine the distribution of the relative angle between polarization orientations of two photons from positron-electron annihilation in a porous polymer. Contrary to prior results for positron annihilation in aluminum and copper, where the strength of observed correlations is as expected for maximally entangled photons, our results show a significant deviation. We demonstrate that in porous polymer, photon polarization correlation is weaker than for maximally entangled photons but stronger than for separable photons. The data indicate that more than 40% of annihilations in Amberlite resin lead to a non-maximally entangled state. Our result indicates the degree of correlation depends on the annihilation mechanism and the molecular arrangement. We anticipate that the introduced Compton interaction-based PET system opens a promising perspective for exploring polarization correlations in PET as a novel diagnostic indicator.
△ Less
Submitted 11 July, 2024;
originally announced July 2024.
-
Real-time antiproton annihilation vertexing with sub-micron resolution
Authors:
M. Berghold,
D. Orsucci,
F. Guatieri,
S. Alfaro,
M. Auzins,
B. Bergmann,
P. Burian,
R. S. Brusa,
A. Camper,
R. Caravita,
F. Castelli,
G. Cerchiari,
R. Ciuryło,
A. Chehaimi,
G. Consolati,
M. Doser,
K. Eliaszuk,
R. Ferguson,
M. Germann,
A. Giszczak,
L. T. Glöggler,
Ł. Graczykowski,
M. Grosbart,
F. Guatieri,
N. Gusakova
, et al. (42 additional authors not shown)
Abstract:
The primary goal of the AEgIS experiment is to precisely measure the free fall of antihydrogen within Earth's gravitational field. To this end, a cold ~50K antihydrogen beam has to pass through two grids forming a moiré deflectometer before annihilating onto a position-sensitive detector, which shall determine the vertical position of the annihilation vertex relative to the grids with micrometric…
▽ More
The primary goal of the AEgIS experiment is to precisely measure the free fall of antihydrogen within Earth's gravitational field. To this end, a cold ~50K antihydrogen beam has to pass through two grids forming a moiré deflectometer before annihilating onto a position-sensitive detector, which shall determine the vertical position of the annihilation vertex relative to the grids with micrometric accuracy. Here we introduce a vertexing detector based on a modified mobile camera sensor and experimentally demonstrate that it can measure the position of antiproton annihilations with an accuracy of $0.62^{+0.40}_{-0.22}μm$, which represents a 35-fold improvement over the previous state-of-the-art for real-time antiproton vertexing. Importantly, these antiproton detection methods are directly applicable to antihydrogen. Moreover, the sensitivity to light of the sensor enables the in-situ calibration of the moiré deflectometer, significantly reducing systematic errors. This sensor emerges as a breakthrough technology for achieving the \aegis scientific goals and has been selected as the basis for the development of a large-area detector for conducting antihydrogen gravity measurements.
△ Less
Submitted 23 June, 2024;
originally announced June 2024.
-
Kaonic lead feasibility measurement at DAΦNE to solve the charged kaon mass discrepancy
Authors:
D. Bosnar,
L. Abbene,
C. Amsler,
F. Artibani,
M. Bazzi,
M. Bragadireanu,
A. Buttacavoli,
M. Cargnelli,
M. Carminati,
A. Clozza,
F. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
K. Dulski,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
M. Makek,
S. Manti,
J. Marton
, et al. (17 additional authors not shown)
Abstract:
An HPGe detector equipped with a transistor reset preamplifier and readout with a CAEN DT5781 fast pulse digitizer was employed in the measurement of X-rays from kaonic lead at the DA$Φ$NE $e^+e^-$ collider at the Laboratori Nazionali di Frascati of INFN. A thin scintillator in front of a lead target was used to select kaons im**ing on it and to form the trigger for the HPGe detector. We present…
▽ More
An HPGe detector equipped with a transistor reset preamplifier and readout with a CAEN DT5781 fast pulse digitizer was employed in the measurement of X-rays from kaonic lead at the DA$Φ$NE $e^+e^-$ collider at the Laboratori Nazionali di Frascati of INFN. A thin scintillator in front of a lead target was used to select kaons im**ing on it and to form the trigger for the HPGe detector. We present the results of the kaonic lead feasibility measurement, where we show that the resolution of the HPGe detector in regular beam conditions remains the same as that without the beam and that a satisfactory background reduction can be achieved. This measurement serves as a test bed for future dedicated kaonic X-rays measurements for the more precise determination of the charged kaon mass.
△ Less
Submitted 21 May, 2024;
originally announced May 2024.
-
SIDDHARTA-2 apparatus for kaonic atoms research on the DA$Φ$NE collider
Authors:
F. Sirghi,
F. Sgaramella,
L. Abbene,
C. Amsler,
M. Bazzi,
G. Borghi,
D. Bosnar,
M. Bragadireanu,
A. Buttacavoli,
M. Carminati,
M. Cargnelli,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
K. Dulski,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton
, et al. (19 additional authors not shown)
Abstract:
SIDDHARTA-2 represents a state-of-the-art experiment designed to perform dedicated measurements of kaonic atoms, which are particular exotic atom configurations composed of a negatively charged kaon and a nucleus. Investigating these atoms provides an exceptional tool to comprehend the strong interactions in the non-perturbative regime involving strangeness. The experiment is installed at the DA…
▽ More
SIDDHARTA-2 represents a state-of-the-art experiment designed to perform dedicated measurements of kaonic atoms, which are particular exotic atom configurations composed of a negatively charged kaon and a nucleus. Investigating these atoms provides an exceptional tool to comprehend the strong interactions in the non-perturbative regime involving strangeness. The experiment is installed at the DA$Φ$NE electron-positron collider, of the INFN National Laboratory of Frascati (INFN-LNF) in Italy, aiming to perform the first-ever measurement of the 2p$\rightarrow$1s X-ray transitions in kaonic deuterium, a crucial step towards determining the isospin-dependent antikaon-nucleon scattering lengths. Based on the experience gained with the previous SIDDHARTA experiment, which performed the most precise measurement of the kaonic hydrogen 2p$\rightarrow$1s X-ray transitions, the present apparatus has been upgraded with innovative Silicon Drift Detectors (SDDs), distributed around a cryogenic gaseous target placed in a vacuum chamber at a short distance above the interaction region of the collider. We present a comprehensive description of the SIDDHARTA-2 setup including the optimization of its various components during the commissioning phase of the collider.
△ Less
Submitted 6 November, 2023;
originally announced November 2023.
-
CdZnTe detectors tested at the DAΦNE collider for future kaonic atoms measurements
Authors:
A. Scordo,
L. Abbene,
F. Artibani,
M. Bazzi,
M. Bettelli,
D. Bosnar,
G. Borghi,
M. Bragadireanu,
A. Buttacavoli,
M. Cargnelli,
M. Carminati,
A. Clozza,
F. Clozza,
L. De Paolis,
G. Deda,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton
, et al. (20 additional authors not shown)
Abstract:
The SIDDHARTA-2 collaboration at the INFN Laboratories of Frascati (LNF) aims to perform groundbreaking measurements on kaonic atoms. In parallel and beyond the ongoing kaonic deuterium, presently running on the DA$Φ$NE collider at LNF, we plan to install additional detectors to perform further kaonic atoms' studies, taking advantage of the unique low energy and low momentum spread $K^-$ beam deli…
▽ More
The SIDDHARTA-2 collaboration at the INFN Laboratories of Frascati (LNF) aims to perform groundbreaking measurements on kaonic atoms. In parallel and beyond the ongoing kaonic deuterium, presently running on the DA$Φ$NE collider at LNF, we plan to install additional detectors to perform further kaonic atoms' studies, taking advantage of the unique low energy and low momentum spread $K^-$ beam delivered by the at-rest decay of the $φ$ meson. CdZnTe devices are ideal for detecting transitions toward both the upper and lower levels of intermediate-mass kaonic atoms, like kaonic carbon and aluminium, which have an important impact on the strangeness sector of nuclear physics. We present the results obtained in a set of preliminary tests conducted on DA$Φ$NE, in view of measurements foreseen in 2024, with the twofold aim to tune the timing window required to reject the extremely high electromagnetic background, and to quantify the readout electronics saturation effect due to the high rate, when placed close to the Interaction Region (IR). In the first test we used commercial devices and electronics, while for the second one both were customized at the IMEM-CNR of Parma and the University of Palermo. The results confirmed the possibility of finding and matching a proper timing window where to identify the signal events and proved better performances, in terms of energy resolution, of the custom system. In both cases, strong saturation effects were confirmed, accounting for a loss of almost 90\% of the events, which will be overcome by a dedicated shielding structure foreseen for the final experimental setup.
△ Less
Submitted 23 October, 2023;
originally announced October 2023.
-
Feasibility studies for imaging e$^{+}$e$^{-}$ annihilation with modular multi-strip detectors
Authors:
S. Sharma,
L. Povolo,
S. Mariazzi,
G. Korcyl,
K. Kacprzak,
D. Kumar,
S. Niedzwiecki,
J. Baran,
E. Beyene,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
M. Das,
K. Dulski,
K. Eliyan,
A. Gajos,
N. Gupta,
B. C. Hiesmayr,
L. Kaplon,
T. Kaplanoglu,
K. Klimaszewski
, et al. (19 additional authors not shown)
Abstract:
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fu…
▽ More
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fundamental studies. With the ability to produce large quantities of Ps, manipulate them in long-lived Ps states, and image their annihilations after a free fall or after passing through atomic interferometers, this purely leptonic antimatter system can be used to perform inertial sensing studies in view of a direct test of Einstein equivalence principle. It is envisioned that modular multistrip detectors can be exploited as potential detection units for this kind of studies. In this work, we report the results of the first feasibility study performed on a e$^{+}$ beamline using two detection modules to evaluate their reconstruction performance and spatial resolution for imaging e$^{+}$e$^{-}$ annihilations and thus their applicability for gravitational studies of Ps.
△ Less
Submitted 12 September, 2023;
originally announced September 2023.
-
Measurements of high-n transitions in intermediate mass kaonic atoms by SIDDHARTA-2 at DA$\mathrmΦ$NE
Authors:
F. Sgaramella,
M. Tüchler,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
R. Del Grande,
L. De Paolis,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki
, et al. (16 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment installed at the DA$\mathrmΦ$NE collider of INFN-LNF performed, for the first time, measurements of high-n transitions in intermediate mass kaonic atoms during the data taking campaigns of 2021 and 2022. Kaonic carbon, oxygen, nitrogen and aluminium transitions, which occur in the setup materials, were measured by using the kaons stopped in the gaseous helium target cell…
▽ More
The SIDDHARTA-2 experiment installed at the DA$\mathrmΦ$NE collider of INFN-LNF performed, for the first time, measurements of high-n transitions in intermediate mass kaonic atoms during the data taking campaigns of 2021 and 2022. Kaonic carbon, oxygen, nitrogen and aluminium transitions, which occur in the setup materials, were measured by using the kaons stopped in the gaseous helium target cell with aluminium frames and Kapton walls, and are reported in this paper. These new kaonic atoms measurements add valuable input to the kaonic atoms transitions data base, which is used as a reference for theories and models of the low-energy strong interaction between antikaon and nuclei. Moreover, these results pave the way for future dedicated kaonic atoms measurements through the whole periodic table and to a new era for the antikaon-nuclei studies at low energy.
△ Less
Submitted 22 April, 2023;
originally announced April 2023.
-
Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators
Authors:
Meysam Dadgar,
Szymon Parzych,
Jakub Baran,
Neha Chug,
Catalina Curceanu,
Eryk Czerwiński,
Kamil Dulski,
Kavya Elyan,
Aleksander Gajos,
Beatrix Hiesmayr,
Łukasz Kapłon,
Konrad Klimaszewski,
Paweł Konieczka,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Deepak Kumar,
Szymon Niedźwiecki,
Domonik Panek,
Eleną Perez del Rio,
Lech Raczyński,
Sushil Sharma,
Shivani,
Roman Shopa,
Magdalena Skurzok
, et al. (6 additional authors not shown)
Abstract:
Background: Total-Body imaging offers high sensitivity, single-bed position, and low dose, but high construction costs limit worldwide utilization. This study compares existing and develo** tomographs using plastic scintillators via simulations to propose a cost-efficient Total-Body PET scanner.
Methods: Simulations of eight uEXPLORER tomographs with different scintillator materials, axial fie…
▽ More
Background: Total-Body imaging offers high sensitivity, single-bed position, and low dose, but high construction costs limit worldwide utilization. This study compares existing and develo** tomographs using plastic scintillators via simulations to propose a cost-efficient Total-Body PET scanner.
Methods: Simulations of eight uEXPLORER tomographs with different scintillator materials, axial field-of-view, and detector configuration, and eight J-PET scanners with various field-of-view, plastic scintillator cross-sections, and layers were performed. Biograph Vision was also simulated. Two types of simulations were conducted with a central source and a water-filled phantom.
Results: BGO crystal-based scanners showed the best sensitivity (350 cps/kBq at the center). Sparse geometry or LYSO crystals reduced sensitivity. J-PET design showed similar sensitivity to sparse LYSO detectors, with full body coverage and additional gain for brain imaging.
Conclusion: The J-PET tomography system using plastic scintillators could be a cost-efficient alternative for Total-Body PET scanners, overcoming high construction costs while maintaining sensitivity
△ Less
Submitted 12 April, 2023;
originally announced April 2023.
-
Feasibility of the J-PET to monitor range of therapeutic proton beams
Authors:
Jakub Baran,
Damian Borys,
Karol Brzeziński,
Jan Gajewski,
Michał Silarski,
Neha Chug,
Aurélien Coussat,
Eryk Czerwiński,
Meysam Dadgar,
Kamil Dulski,
Kavya V. Eliyan,
Aleksander Gajos Krzysztof Kacprzak,
Łukasz Kapłon,
Konrad Klimaszewski,
Paweł Konieczka,
Renata Kopeć,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Deepak Kumar,
Antony J. Lomax,
Keegan McNamara,
Szymon Niedźwiecki,
Paweł Olko,
Dominik Panek
, et al. (18 additional authors not shown)
Abstract:
Objective: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring. Approach: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries (three dual-heads and three cylindrical). We simulated prot…
▽ More
Objective: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring. Approach: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries (three dual-heads and three cylindrical). We simulated proton irradiation of a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out Bragg Peak (SOBP) of various ranges. The sensitivity and precision of each scanner were calculated, and considering the setup's cost-effectiveness, we indicated potentially optimal geometries for the J-PET scanner prototype dedicated to the proton beam range assessment. Main results: The investigations indicate that the double-layer cylindrical and triple-layer double-head configurations are the most promising for clinical application. We found that the scanner sensitivity is of the order of 10$^{-5}$ coincidences per primary proton, while the precision of the range assessment for both SPB and SOBP irradiation plans was found below 1 mm. Among the scanners with the same number of detector modules, the best results are found for the triple-layer dual-head geometry. Significance: We performed simulation studies demonstrating that the feasibility of the J-PET detector for PET-based proton beam therapy range monitoring is possible with reasonable sensitivity and precision enabling its pre-clinical tests in the clinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometry configurations seem promising for the future clinical application. Experimental tests are needed to confirm these findings.
△ Less
Submitted 28 February, 2023;
originally announced February 2023.
-
Positronium Physics and Biomedical Applications
Authors:
Steven D. Bass,
Sebastiano Mariazzi,
Pawel Moskal,
Ewa Stepien
Abstract:
Positronium is the simplest bound state, built of an electron and a positron. Studies of positronium in vacuum and its decays in medium tell us about Quantum Electrodynamics, QED, and about the structure of matter and biological processes of living organisms at the nanoscale, respectively. Spectroscopic measurements constrain our understanding of QED bound state theory. Searches for rare decays an…
▽ More
Positronium is the simplest bound state, built of an electron and a positron. Studies of positronium in vacuum and its decays in medium tell us about Quantum Electrodynamics, QED, and about the structure of matter and biological processes of living organisms at the nanoscale, respectively. Spectroscopic measurements constrain our understanding of QED bound state theory. Searches for rare decays and measurements of the effect of gravitation on positronium are used to look for new physics phenomena. In biological materials positronium decays are sensitive to the inter- and intra-molecular structure and to the metabolism of living organisms ranging from single cells to human beings. This leads to new ideas of positronium imaging in medicine using the fact that during positron emission tomography (PET) as much as 40% of positron annihilation occurs through the production of positronium atoms inside the patient's body. A new generation of the high sensitivity and multi-photon total-body PET systems opens perspectives for clinical applications of positronium as a biomarker of tissue pathology and the degree of tissue oxidation.
△ Less
Submitted 18 February, 2023;
originally announced February 2023.
-
TOF MLEM Adaptation for the Total-Body J-PET with a Realistic Analytical System Response Matrix
Authors:
R. Y. Shopa,
J. Baran,
K. Klimaszewski,
W. Krzemień,
L. Raczyński,
W. Wiślicki,
K. Brzeziński,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwiński,
M. Dadgar,
K. Dulski,
J. Gajewski,
A. Gajos,
B. C. Hiesmayr,
E. Kavya Valsan,
G. Korcyl,
T. Kozik,
D. Kumar,
Ł. Kapłon,
G. Moskal,
S. Niedźwiecki,
D. Panek,
S. Parzych
, et al. (10 additional authors not shown)
Abstract:
We report a study of the original image reconstruction algorithm based on the time-of-flight maximum likelihood expectation maximisation (TOF MLEM), developed for the total-body (TB) Jagiellonian PET (J-PET) scanners. The method is applicable to generic cylindrical or modular multi-layer layouts and is extendable to multi-photon imaging. The system response matrix (SRM) is represented as a set of…
▽ More
We report a study of the original image reconstruction algorithm based on the time-of-flight maximum likelihood expectation maximisation (TOF MLEM), developed for the total-body (TB) Jagiellonian PET (J-PET) scanners. The method is applicable to generic cylindrical or modular multi-layer layouts and is extendable to multi-photon imaging. The system response matrix (SRM) is represented as a set of analytical functions, uniquely defined for each pair of plastic scintillator strips used for the detection. A realistic resolution model (RM) in detector space is derived from fitting the Monte Carlo simulated emissions and detections of annihilation photons on oblique transverse planes. Additional kernels embedded in SRM account for TOF, parallax effect and axial smearing. The algorithm was tested on datasets, simulated in GATE for the NEMA IEC and static XCAT phantoms inside a 24-module 2-layer TB J-PET. Compared to the reference TOF MLEM with none or a shift-invariant RM, an improvement was observed, as evaluated by the analysis of image quality, difference images and ground truth metrics. We also reconstructed the data with additive contributions, pre-filtered geometrically and with non-TOF scatter correction applied. Despite some deterioration, the obtained results still capitalise on the realistic RM with better edge preservation and superior ground truth metrics. The envisioned prospects of the TOF MLEM with analytical SRM include its application in multi-photon imaging and further upgrade to account for the non-collinearity, positron range and other factors.
△ Less
Submitted 4 May, 2023; v1 submitted 6 February, 2023;
originally announced February 2023.
-
New opportunities for kaonic atoms measurements from CdZnTe detectors
Authors:
L. Abbene,
M. Bettelli,
A. Buttacavoli,
F. Principato,
A. Zappettini,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton,
M. Miliucci
, et al. (20 additional authors not shown)
Abstract:
We present the tests performed by the SIDDHARTA-2 collaboration at the DAΦNE collider with a quasi-hemispherical CdZnTe detector. The very good room-temperature energy resolution and efficiency in a wide energy range show that this detector technology is ideal for studying radiative transitions in intermediate and heavy-mass kaonic atoms. The CdZnTe detector was installed for the first time in an…
▽ More
We present the tests performed by the SIDDHARTA-2 collaboration at the DAΦNE collider with a quasi-hemispherical CdZnTe detector. The very good room-temperature energy resolution and efficiency in a wide energy range show that this detector technology is ideal for studying radiative transitions in intermediate and heavy-mass kaonic atoms. The CdZnTe detector was installed for the first time in an accelerator environment to perform tests on the background rejection capabilities, which were achieved by exploiting the SIDDHARTA-2 Luminosity Monitor. A spectrum with an $^{241}Am$ source has been acquired, with beams circulating in the main rings, and peak resolutions of 6% at 60 keV and of 2.2% at 511 keV have been achieved. The background suppression factor, which turned out to be of the order of $\simeq10^{5-6}$, opens the possibility to plan for future kaonic atom measurements with CdZnTe detectors.
△ Less
Submitted 28 January, 2023;
originally announced January 2023.
-
J-PET detection modules based on plastic scintillators for performing studies with positron and positronium beams
Authors:
S. Sharma,
J. Baran,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
K. Eliyan,
A. Gajos,
B. C. Hiesmayr,
K. Kacprzak,
L. Kaplon,
K. Klimaszewski,
P. Konieczka,
G. Korcyl,
T. Kozik,
W. Krzemien,
D. Kumar,
S. Mariazzi,
S. Niedzwiecki,
L. Panasa,
S. Parzych
, et al. (11 additional authors not shown)
Abstract:
The J-PET detector, which consists of inexpensive plastic scintillators, has demonstrated its potential in the study of fundamental physics. In recent years, a prototype with 192 plastic scintillators arranged in 3 layers has been optimized for the study of positronium decays. This allows performing precision tests of discrete symmetries (C, P, T) in the decays of positronium atoms. Moreover, than…
▽ More
The J-PET detector, which consists of inexpensive plastic scintillators, has demonstrated its potential in the study of fundamental physics. In recent years, a prototype with 192 plastic scintillators arranged in 3 layers has been optimized for the study of positronium decays. This allows performing precision tests of discrete symmetries (C, P, T) in the decays of positronium atoms. Moreover, thanks to the possibility of measuring the polarization direction of the photon based on Compton scattering, the predicted entanglement between the linear polarization of annihilation photons in positronium decays can also be studied. Recently, a new J-PET prototype was commissioned, based on a modular design of detection units. Each module consists of 13 plastic scintillators and can be used as a stand-alone, compact and portable detection unit. In this paper, the main features of the J-PET detector, the modular prototype and their applications for possible studies with positron and positronium beams are discussed. Preliminary results of the first test experiment performed on two detection units in the continuous positron beam recently developed at the Antimatter Laboratory (AML) of Trento are also reported.
△ Less
Submitted 7 January, 2023;
originally announced January 2023.
-
Realistic Total-Body J-PET Geometry Optimization -- Monte Carlo Study
Authors:
Jakub Baran,
Wojciech Krzemien,
Lech Raczyński,
Mateusz Bała,
Aurelien Coussat,
Szymon Parzych,
Neha Chug,
Eryk Czerwiński,
Catalina Oana Curceanu,
Meysam Dadgar,
Kamil Dulski,
Kavya Eliyan,
Jan Gajewski,
Aleksander Gajos,
Beatrix Hiesmayr,
Krzysztof Kacprzak,
Łukasz Kapłon,
Konrad Klimaszewski,
Grzegorz Korcyl,
Tomasz Kozik,
Deepak Kumar,
Szymon Niedźwiecki,
Dominik Panek,
Elena Perez del Rio,
Antoni Ruciński
, et al. (9 additional authors not shown)
Abstract:
Total-Body PET is one of the most promising medical diagnostics modalities. The high sensitivity provided by Total-Body technology can be advantageous for novel tomography methods like positronium imaging. Several efforts are ongoing to lower the price of the TB-PET systems. Among the alternatives, the Jagiellonian PET (J-PET) technology, based on plastic scintillator strips, offers a low-cost alt…
▽ More
Total-Body PET is one of the most promising medical diagnostics modalities. The high sensitivity provided by Total-Body technology can be advantageous for novel tomography methods like positronium imaging. Several efforts are ongoing to lower the price of the TB-PET systems. Among the alternatives, the Jagiellonian PET (J-PET) technology, based on plastic scintillator strips, offers a low-cost alternative. The work aimed to compare five Total-Body J-PET geometries as a possible next generation J-PET scanner design. We present comparative studies of performance characteristics of the cost-effective Total-Body PET scanners using J-PET technology. We investigated in silico five Total-Body scanner geometries. Monte Carlo simulations of the XCAT phantom, the 2-meter sensitivity line source and positronium sensitivity phantoms were performed. We compared the sensitivity profiles for 2-gamma and 3-gamma tomography, relative cost of the setups and performed quantitative analysis of the reconstructed images. The analysis of the reconstructed XCAT images reveals the superiority of the seven-ring scanners over the three-ring setups. However, the three-ring scanners would be approximately 2-3 times cheaper. The peak sensitivity values for two-gamma vary from 20 to 34 cps/kBq. The sensitivity curves for the positronium tomography have a similar shape to the two-gamma sensitivity profiles. The peak values are lower compared to the two-gamma cases, from about 20-28 times, with a maximum of 1.66 cps/kBq. The results show the feasibility of multi-organ imaging of all the systems to be considered for the next generation of TB J-PET designs. The relative cost for all the scanners is about 10-4 times lower compared to the cost of the uExplorer. These properties coupled together with J-PET cost-effectiveness, make the J-PET technology an attractive solution for broad application in clinics.
△ Less
Submitted 16 February, 2024; v1 submitted 5 December, 2022;
originally announced December 2022.
-
Decoherence puzzle in measurements of photons originating from electron-positron annihilation
Authors:
Sushil Sharma,
Deepak Kumar,
Pawel Moskal
Abstract:
Entanglement of photons originating from the electron-positron annihilation has not been proven experimentally. Though the independent experiments performed so far unanimously confirm that correlation between the linear polarizations of back-to-back photons from the electron-positron annihilation is consistent with the assumption that these photons are entangled in polarization. Yet, unexpectedly,…
▽ More
Entanglement of photons originating from the electron-positron annihilation has not been proven experimentally. Though the independent experiments performed so far unanimously confirm that correlation between the linear polarizations of back-to-back photons from the electron-positron annihilation is consistent with the assumption that these photons are entangled in polarization. Yet, unexpectedly, recent experiments differ as regards the correlation of polarization direction of back-to-back photons after the decoherence induced by scattering of one of these photons on the electron in the scattering material. In one of the experiments, the correlation before and after the decoherence of the photon state is the same and in the other experiment the scattering of one photon leads to a significant decrease in this correlation. Here we discuss this puzzle. The decoherent states were ensured by forcing one of the annihilation photons through prior scattering before the Compton kinematics based measurement of the polarization correlation. A comparison between the experimental setups used for the different measurements and the results obtained are briefly discussed, highlighting the parameters that are important for performing such measurements. Finally, the main features of the J-PET detector are presented with the schemes for performing similar measurements so that the results can be used as conclusive remarks for solving this puzzle. Solving the decoherence puzzle will have crucial consequences for basic studies of entanglement, as well as for the proposed application of polarisation of photons in positron emission tomography. In case if the correlation of polarisation of back-to-back photons from the electron-positron annihilation is the same before and after the scattering of these photons, then it will not be useful for the reduction of the scatter fraction in PET diagnostics.
△ Less
Submitted 15 November, 2022; v1 submitted 16 October, 2022;
originally announced October 2022.
-
Studies of the linearity and stability of Silicon Drift Detectors for kaonic atoms X-ray spectroscopy
Authors:
A. Khreptak,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Carminati,
M. Cargnelli,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki,
H. Ohnishi,
K. Piscicchia,
Y. Sada
, et al. (14 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform precision measurements of kaonic atoms X-ray spectroscopy for the investigation of the antikaon-nucleon strong interaction. To achieve this goal, novel large-area Silicon Drift Detectors (SDDs) have been developed. These devices have special geometry, field configuration and readout electronics that ensure excellent performance in…
▽ More
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform precision measurements of kaonic atoms X-ray spectroscopy for the investigation of the antikaon-nucleon strong interaction. To achieve this goal, novel large-area Silicon Drift Detectors (SDDs) have been developed. These devices have special geometry, field configuration and readout electronics that ensure excellent performance in terms of linearity and stability. The paper presents preliminary results for the linearity determination and stability monitoring of the SDDs system during the measurement of kaonic deuterium carried out in the summer of 2022.
△ Less
Submitted 31 August, 2022;
originally announced August 2022.
-
First tests of the full SIDDHARTA-2 experimental apparatus with a 4 He gaseous target
Authors:
A. Scordo,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
P. King,
P. Levi Sandri,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki
, et al. (16 additional authors not shown)
Abstract:
In this paper, we present the first tests performed after the full installation of the SIDDHARTA-2 experimental apparatus on the Interaction Region of the DAΦNE collider at the INFN National Laboratories of Frascati. Before starting the first measurement of the kaonic deuterium 2p{\rightarrow}1s transition, accurate evaluation of the background rejection. mainly achieved with the Kaon Trigger syst…
▽ More
In this paper, we present the first tests performed after the full installation of the SIDDHARTA-2 experimental apparatus on the Interaction Region of the DAΦNE collider at the INFN National Laboratories of Frascati. Before starting the first measurement of the kaonic deuterium 2p{\rightarrow}1s transition, accurate evaluation of the background rejection. mainly achieved with the Kaon Trigger system, was required. This run, performed in the period 04-26/05/2022 with a 4 He gaseous target, confirmed the 10^5 rejection factor obtained with a reduced version of the setup and different machine conditions in 2021. This important outcome motivated the filling of the target cell with deuterium and the starting of the measurement campaign of the kaonic deuterium 2p{\rightarrow}1s transition.
△ Less
Submitted 4 August, 2022;
originally announced August 2022.
-
The SIDDHARTA-2 calibration method for high precision kaonic atoms X-ray spectroscopy measurements
Authors:
F Sgaramella,
M Miliucci,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P King,
P Levi Sandri,
J Marton,
P Moskal,
F Napolitano,
S Niedźwiecki,
K Piscicchia,
A Scordo,
H Shi,
M Silarski
, et al. (7 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform the first kaonic deuterium X-ray transitions to the fundamental level measurement, with a systematic error at the level of a few eV. To achieve this challenging goal the experimental apparatus is equipped with 384 Silicon Drift Detectors (SDDs) distributed around its cryogenic gaseous target. The SDDs developed by the SIDDHARTA-2 c…
▽ More
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform the first kaonic deuterium X-ray transitions to the fundamental level measurement, with a systematic error at the level of a few eV. To achieve this challenging goal the experimental apparatus is equipped with 384 Silicon Drift Detectors (SDDs) distributed around its cryogenic gaseous target. The SDDs developed by the SIDDHARTA-2 collaboration are suitable for high precision kaonic atoms spectroscopy, thanks to their high energy and time resolutions combined with their radiation hardness. The energy response of each detector must be calibrated and monitored to keep the systematic error, due to processes such as gain fluctuations, at the level of 2-3 eV. This paper presents the SIDDHARTA-2 calibration method which was optimized during the preliminary phase of the experiment in the real background conditions of the DA$Φ$NE collider, which is a fundamental tool to guarantee the high precision spectroscopic performances of the system over long periods of data taking, as that required for the kaonic deuterium measurement.
△ Less
Submitted 28 January, 2022;
originally announced January 2022.
-
Kaonic Atoms at the DA$Φ$NE Collider with the SIDDHARTA-2 Experiment
Authors:
F Napolitano,
F Sgaramella,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
L Fabbietti,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P Levi Sandri,
J Marton,
M Miliucci,
P Moskal,
S Niedźwiecki,
K Piscicchia,
A Scordo,
H Shi,
D Sirghi
, et al. (7 additional authors not shown)
Abstract:
Kaonic atoms are a unique tool to explore quantum chromodynamics in the strangeness sector at low energy, with implications reaching neutron stars and dark matter. Precision X-ray spectroscopy can fully unlock the at-threshold isospin dependent antikaon-nucleon scattering lengths, via the atomic transitions to the fundamental level. While the SIDDHARTA experiment at the INFN-LNF DA$Φ$NE collider s…
▽ More
Kaonic atoms are a unique tool to explore quantum chromodynamics in the strangeness sector at low energy, with implications reaching neutron stars and dark matter. Precision X-ray spectroscopy can fully unlock the at-threshold isospin dependent antikaon-nucleon scattering lengths, via the atomic transitions to the fundamental level. While the SIDDHARTA experiment at the INFN-LNF DA$Φ$NE collider successfully measured kaonic hydrogen, its successor SIDDHARTA-2 is starting now its data taking campaign aiming to finally fully disentangle the isoscalar and isovector scattering lengths via the measurement of kaonic deuterium. An overview of the first experimental results from a preparatory run for the SIDDAHARTA-2 experiment is presented.
△ Less
Submitted 27 January, 2022;
originally announced January 2022.
-
A new kaonic helium measurement in gas by SIDDHARTINO at the DAΦNE collider
Authors:
D Sirghi,
F Sirghi,
F Sgaramella,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
L Fabbietti,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P Levi Sandri,
J Marton,
M Miliucci,
P Moskal,
F Napolitano,
S Niedźwiecki,
K Piscicchia,
A Scordo
, et al. (8 additional authors not shown)
Abstract:
The SIDDHARTINO experiment at the DAΦNE Collider of INFN-LNF, the pilot run for the SIDDHARTA-2 experiment which aims to perform the measurement of kaonic deuterium transitions to the fundamental level, has successfully been concluded. The paper reports the main results of this run, including the optimization of various components of the apparatus, among which the degrader needed to maximize the f…
▽ More
The SIDDHARTINO experiment at the DAΦNE Collider of INFN-LNF, the pilot run for the SIDDHARTA-2 experiment which aims to perform the measurement of kaonic deuterium transitions to the fundamental level, has successfully been concluded. The paper reports the main results of this run, including the optimization of various components of the apparatus, among which the degrader needed to maximize the fraction of kaons stopped inside the target, through measurements of kaonic helium transitions to the 2p level. The obtained shift and width values are ε_2p = E_exp-E_e.m = 0.2 {\pm} 2.5(stat) {\pm} 2(syst) eV and Γ_2p = 8 {\pm} 10 eV (stat), respectively. This new measurement of the shift, in particular, represents the most precise one for a gaseous target and is expected to contribute to a better understanding of the kaon-nuclei interaction at low energy.
△ Less
Submitted 24 January, 2022;
originally announced January 2022.
-
Positronium imaging with the novel multiphoton PET scanner
Authors:
Paweł Moskal,
Kamil Dulski,
Neha Chug,
Catalina Curceanu,
Eryk Czerwiński,
Meysam Dadgar,
Jan Gajewski,
Aleksander Gajos,
Grzegorz Grudzień,
Beatrix C. Hiesmayr,
Krzysztof Kacprzak,
Łukasz Kapłon,
Hanieh Karimi,
Konrad Klimaszewski,
Grzegorz Korcyl,
Paweł Kowalski,
Tomasz Kozik,
Nikodem Krawczyk,
Wojciech Krzemień,
Ewelina Kubicz,
Piotr Małczak,
Szymon Niedźwiecki,
Monika Pawlik-Niedźwiecka,
Michał Pędziwiatr,
Lech Raczyński
, et al. (11 additional authors not shown)
Abstract:
In vivo assessment of cancer and precise location of altered tissues at initial stages of molecular disorders are important diagnostic challenges. Positronium is copiously formed in the free molecular spaces in the patient's body during positron emission tomography (PET). The positronium properties vary according to the size of inter- and intramolecular voids and the concentration of molecules in…
▽ More
In vivo assessment of cancer and precise location of altered tissues at initial stages of molecular disorders are important diagnostic challenges. Positronium is copiously formed in the free molecular spaces in the patient's body during positron emission tomography (PET). The positronium properties vary according to the size of inter- and intramolecular voids and the concentration of molecules in them such as, e.g., molecular oxygen, O2; therefore, positronium imaging may provide information about disease progression during the initial stages of molecular alterations. Current PET systems do not allow acquisition of positronium images. This study presents a new method that enables positronium imaging by simultaneous registration of annihilation photons and deexcitation photons from pharmaceuticals labeled with radionuclides. The first positronium imaging of a phantom built from cardiac myxoma and adipose tissue is demonstrated. It is anticipated that positronium imaging will substantially enhance the specificity of PET diagnostics.
△ Less
Submitted 16 December, 2021;
originally announced December 2021.
-
Silicon Drift Detectors spectroscopic response during the SIDDHARTA-2 Kaonic Helium run at the DAΦNE collider
Authors:
Marco Miliucci,
Massimiliano Bazzi,
Damir Bosnar,
Mario Bragadireanu,
Marco Carminati,
Michael Cargnelli,
Alberto Clozza,
Catalina Curceanu,
Griseld Deda,
Luca De Paolis,
Raffaele Del Grande,
Carlo Fiorini,
Carlo Guaraldo,
Mihail Iliescu,
Masahiko Iwasaki,
Pietro King,
Paolo Levi Sandri,
Johann Marton,
Paweł Moskal,
Fabrizio Napolitano,
Szymon Niedźwiecki,
Kristian Piscicchia,
Alessandro Scordo,
Francesco Sgaramella,
Hexi Shi
, et al. (8 additional authors not shown)
Abstract:
A large-area Silicon Drift Detectors (SDDs) system has been developed by the SIDDHARTA2 collaboration for high precision light kaonic atoms X-ray spectroscopy at the DAΦNE collider of Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali di Frascati. The SDDs geometry and electric field configuration, combined with their read-out electronics, make these devices suitable to perform high prec…
▽ More
A large-area Silicon Drift Detectors (SDDs) system has been developed by the SIDDHARTA2 collaboration for high precision light kaonic atoms X-ray spectroscopy at the DAΦNE collider of Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali di Frascati. The SDDs geometry and electric field configuration, combined with their read-out electronics, make these devices suitable to perform high precision light kaonic atoms spectroscopy measurements in the high background of the DAΦNE collider. This work presents the spectroscopic response of the SDDs system during the kaonic helium first exotic atoms run of SIDDHARTA-2, preliminary to the kaonic deuterium data taking campaign
△ Less
Submitted 2 November, 2021;
originally announced November 2021.
-
Optimisation of the event-based TOF filtered back-projection for online imaging in total-body J-PET
Authors:
R. Y. Shopa,
K. Klimaszewski,
P. Kopka,
P. Kowalski,
W. Krzemień,
L. Raczyński,
W. Wiślicki,
N. Chug,
C. Curceanu,
E. Czerwiński,
M. Dadgar,
K. Dulski,
A. Gajos,
B. C. Hiesmayr,
K. Kacprzak,
Ł. Kapłon,
D. Kisielewska,
G. Korcyl,
N. Krawczyk,
E. Kubicz,
Sz. Niedźwiecki,
J. Raj,
S. Sharma,
Shivani,
E. L. Stȩpień
, et al. (2 additional authors not shown)
Abstract:
We perform a parametric study of the newly developed time-of-flight (TOF) image reconstruction algorithm, proposed for the real-time imaging in total-body Jagiellonian PET (J-PET) scanners. The asymmetric 3D filtering kernel is applied at each most likely position of electron-positron annihilation, estimated from the emissions of back-to-back $γ$-photons. The optimisation of its parameters is stud…
▽ More
We perform a parametric study of the newly developed time-of-flight (TOF) image reconstruction algorithm, proposed for the real-time imaging in total-body Jagiellonian PET (J-PET) scanners. The asymmetric 3D filtering kernel is applied at each most likely position of electron-positron annihilation, estimated from the emissions of back-to-back $γ$-photons. The optimisation of its parameters is studied using Monte Carlo simulations of a 1-mm spherical source, NEMA IEC and XCAT phantoms inside the ideal J-PET scanner. The combination of high-pass filters which included the TOF filtered back-projection (FBP), resulted in spatial resolution, 1.5 $\times$ higher in the axial direction than for the conventional 3D FBP. For realistic $10$-minute scans of NEMA IEC and XCAT, which require a trade-off between the noise and spatial resolution, the need for Gaussian TOF kernel components, coupled with median post-filtering, is demonstrated. The best sets of 3D filter parameters were obtained by the Nelder-Mead minimisation of the mean squared error between the resulting and reference images. The approach allows training the reconstruction algorithm for custom scans, using the IEC phantom, when the temporal resolution is below 50 ps. The image quality parameters, estimated for the best outcomes, were systematically better than for the non-TOF FBP.
△ Less
Submitted 27 July, 2021;
originally announced July 2021.
-
Simulating NEMA characteristics of the modular total-body J-PET scanner -- an economic total-body PET from plastic scintillators
Authors:
Paweł Moskal,
Paweł Kowalski,
Roman Shopa,
Lech Raczyński,
Jakub Baran,
Neha Chug,
Catalina Curceanu,
Eryk Czerwiński,
Meysam Dadgar,
Kamil Dulski,
Aleksander Gajos,
Beatrix Hiesmayr,
Krzysztof Kacprzak,
Łukasz Kapłon,
Daria Kisielewska,
Konrad Klimaszewski,
Przemysław Kopka,
Gregorz Korcyl,
Nikodem Krawczyk,
Wojciech Krzemień,
Ewelina Kubicz,
Szymon Niedźwiecki,
Szymon Parzych,
Juhi Raj,
Sushil Sharma
, et al. (4 additional authors not shown)
Abstract:
The purpose of the presented research is estimation of the performance characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET) constructed from plastic scintillators. The characteristics are estimated according to the NEMA NU-2-2018 standards utilizing the GATE package. The simulated detector consists of 24 modules, each built out of 32 plastic scintillator strips (each with…
▽ More
The purpose of the presented research is estimation of the performance characteristics of the economic Total-Body Jagiellonian-PET system (TB-J-PET) constructed from plastic scintillators. The characteristics are estimated according to the NEMA NU-2-2018 standards utilizing the GATE package. The simulated detector consists of 24 modules, each built out of 32 plastic scintillator strips (each with cross section of 6 mm times 30 mm and length of 140 cm or 200 cm) arranged in two layers in regular 24-sided polygon circumscribing a circle with the diameter of 78.6 cm. For the TB-J-PET with an axial field-of-view (AFOV) of 200 cm, a spatial resolutions of 3.7 mm (transversal) and 4.9 mm (axial) are achieved. The NECR peak of 630 kcps is expected at 30 kBq/cc activity concentration and the sensitivity at the center amounts to 38 cps/kBq. The SF is estimated to 36.2 %. The values of SF and spatial resolution are comparable to those obtained for the state-of-the-art clinical PET scanners and the first total-body tomographs: uExplorer and PennPET. With respect to the standard PET systems with AFOV in the range from 16 cm to 26 cm, the TB-J-PET is characterized by an increase in NECR approximately by factor of 4 and by the increase of the whole-body sensitivity by factor of 12.6 to 38. The TOF resolution for the TB-J-PET is expected to be at the level of CRT=240 ps (FWHM). For the TB-J-PET with an axial field-of-view (AFOV) of 140 cm, an image quality of the reconstructed images of a NEMA IEC phantom was presented with a contrast recovery coefficient (CRC) and a background variability parameters. The increase of the whole-body sensitivity and NECR estimated for the TB-J-PET with respect to current commercial PET systems makes the TB-J-PET a promising cost-effective solution for the broad clinical applications of total-body PET scanners.
△ Less
Submitted 3 July, 2021;
originally announced July 2021.
-
Synchronisation and calibration of the 24-modules J-PET prototype with 300~mm axial field of view
Authors:
P. Moskal,
T. Bednarski,
Sz. Niedzwiecki,
M. Silarski,
E. Czerwinski,
T. Kozik,
J. Chhokar,
M. Bała,
C. Curceanu,
R. Del Grande,
M. Dadgar,
K. Dulski,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk
, et al. (14 additional authors not shown)
Abstract:
Research conducted in the framework of the J-PET project aims to develop a cost-effective total-body positron emission tomography scanner. As a first step on the way to construct a full-scale J-PET tomograph from long strips of plastic scintillators, a 24-strip prototype was built and tested. The prototype consists of detection modules arranged axially forming a cylindrical diagnostic chamber with…
▽ More
Research conducted in the framework of the J-PET project aims to develop a cost-effective total-body positron emission tomography scanner. As a first step on the way to construct a full-scale J-PET tomograph from long strips of plastic scintillators, a 24-strip prototype was built and tested. The prototype consists of detection modules arranged axially forming a cylindrical diagnostic chamber with the inner diameter of 360 mm and the axial field-of-view of 300 mm. Promising perspectives for a low-cost construction of a total-body PET scanner are opened due to an axial arrangement of strips of plastic scintillators, wchich have a small light attenuation, superior timing properties, and the possibility of cost-effective increase of the axial field-of-view. The presented prototype comprises dedicated solely digital front-end electronic circuits and a triggerless data acquisition system which required development of new calibration methods including time, thresholds and gain synchronization. The system and elaborated calibration methods including first results of the 24-module J-PET prototype are presented and discussed. The achieved coincidence resolving time equals to CRT = 490 $\pm$ 9 ps. This value can be translated to the position reconstruction accuracy $σ(Δl) =$ 18 mm which is fairly position-independent.
△ Less
Submitted 25 August, 2020;
originally announced August 2020.
-
Characterization of the SIDDHARTA-2 luminosity monitor
Authors:
M. Skurzok,
A. Scordo,
S. Niedzwiecki,
A. Baniahmad,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Carminati,
M. Cargnelli,
A. Clozza,
C. Curceanu,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
P. Levi Sandri,
J. Marton,
M. Miliucci,
P. Moskal,
K. Piscicchia,
F. Sgaramella,
H. Shi
, et al. (7 additional authors not shown)
Abstract:
A luminosity monitor, based on plastic scintillator detectors, has been developed for the SIDDHARTA-2 experiment aiming to perform high precision measurements of kaonic atoms and was installed in 2020 on the DAFNE $e^+e^-$ collider at LNF (Laboratori Nazionali di Frascati, INFN). The main goal of this system is to provide the~instantaneous and integrated luminosity of the DAFNE facility by measuri…
▽ More
A luminosity monitor, based on plastic scintillator detectors, has been developed for the SIDDHARTA-2 experiment aiming to perform high precision measurements of kaonic atoms and was installed in 2020 on the DAFNE $e^+e^-$ collider at LNF (Laboratori Nazionali di Frascati, INFN). The main goal of this system is to provide the~instantaneous and integrated luminosity of the DAFNE facility by measuring the rate of $K^+K^-$ correlated pairs emitted by the phi meson decay. This task requires an accurate timing of the DAQ signals, as well as timing resolution below 1ns, in order to disentangle the $K^\pm$ signals from the background minimum ionizing particles (MIPs) produced during the $e^+e^-$ collisions at DAFNE. In this paper the luminosity monitor concept as well as its laboratory characterization and the first results inside DAFNE are presented.
△ Less
Submitted 15 October, 2020; v1 submitted 12 August, 2020;
originally announced August 2020.
-
The J-PET detector -- a tool for precision studies of ortho-positronium decays
Authors:
K. Dulski,
S. D. Bass,
J. Chhokar,
N. Chug,
C. Curceanu,
E. Czerwiński,
M. Dadgar,
J. Gajewski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
P. Kopka,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz
, et al. (19 additional authors not shown)
Abstract:
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events…
▽ More
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events is based on the measurement of the photons energies. In this article we show that the J-PET tomograph system is capable of exclusive measurements of the decays of ortho-positronium atoms. We present the first positronium production results, its lifetime distribution measurements and discuss estimation of the influence of various background sources. The tomograph s performance demonstrated here makes it suitable for precision studies of positronium decays including entanglement of the final state photons, positron annihilation lifetime spectroscopy plus molecular imaging diagnostics.
△ Less
Submitted 25 May, 2021; v1 submitted 9 June, 2020;
originally announced June 2020.
-
Hit-time and hit-position reconstruction in strips of plastic scintillators using multi-threshold readouts
Authors:
N. G. Sharma,
M. Silarski,
J. Chhokar,
E. Czerwinski,
C. Curceanu,
K. Dulski,
K. Farbaniec,
A. Gajos,
R. Del Grande,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemien,
T. Kozik,
E. Kubicz,
M. Mohammed,
Sz. Niedzwiecki,
M. Palka
, et al. (10 additional authors not shown)
Abstract:
In this article, a new method for the reconstruction of hit-position and hit-time of photons in long scintillator detectors is investigated. This research is motivated by the recent development of the positron emission tomography scanners based on plastic scintillators. The proposed method constitutes a new way of signal processing in Multi-Voltage-Technique. It is based on the determination of th…
▽ More
In this article, a new method for the reconstruction of hit-position and hit-time of photons in long scintillator detectors is investigated. This research is motivated by the recent development of the positron emission tomography scanners based on plastic scintillators. The proposed method constitutes a new way of signal processing in Multi-Voltage-Technique. It is based on the determination of the degree of similarity between the registered signals and the synchronized model signals stored in a library. The library was established for a set of well defined hit-positions along the length of the scintillator. The Mahalanobis distance was used as a measure of similarity between the two compared signals. The method was validated on the experimental data measured using two-strips J-PET prototype with dimensions of 5x9x300 mm$^3$. The obtained Time-of-Flight (TOF) and spatial resolutions amount to 325~ps (FWHM) and 25~mm (FWHM), respectively. The TOF resolution was also compared to the results of an analogous study done using Linear Fitting method. The best TOF resolution was obtained with this method at four pre-defined threshold levels which was comparable to the resolution achieved from the Mahalanobis distance at two pre-defined threshold levels. Although the algorithm of Linear Fitting method is much simpler to apply than the Mahalanobis method, the application of the Mahalanobis distance requires a lower number of applied threshold levels and, hence, decreases the costs of electronics used in PET scanner.
△ Less
Submitted 27 April, 2020;
originally announced April 2020.
-
Technical Design Report for the PANDA Endcap Disc DIRC
Authors:
Panda Collaboration,
F. Davi,
W. Erni,
B. Krusche,
M. Steinacher,
N. Walford,
H. Liu,
Z. Liu,
B. Liu,
X. Shen,
C. Wang,
J. Zhao,
M. Albrecht,
T. Erlen,
F. Feldbauer,
M. Fink,
V. Freudenreich,
M. Fritsch,
F. H. Heinsius,
T. Held,
T. Holtmann,
I. Keshk,
H. Koch,
B. Kopf,
M. Kuhlmann
, et al. (441 additional authors not shown)
Abstract:
PANDA (anti-Proton ANnihiliation at DArmstadt) is planned to be one of the four main experiments at the future international accelerator complex FAIR (Facility for Antiproton and Ion Research) in Darmstadt, Germany. It is going to address fundamental questions of hadron physics and quantum chromodynamics using cooled antiproton beams with a high intensity and and momenta between 1.5 and 15 GeV/c.…
▽ More
PANDA (anti-Proton ANnihiliation at DArmstadt) is planned to be one of the four main experiments at the future international accelerator complex FAIR (Facility for Antiproton and Ion Research) in Darmstadt, Germany. It is going to address fundamental questions of hadron physics and quantum chromodynamics using cooled antiproton beams with a high intensity and and momenta between 1.5 and 15 GeV/c. PANDA is designed to reach a maximum luminosity of 2x10^32 cm^2 s. Most of the physics programs require an excellent particle identification (PID). The PID of hadronic states at the forward endcap of the target spectrometer will be done by a fast and compact Cherenkov detector that uses the detection of internally reflected Cherenkov light (DIRC) principle. It is designed to cover the polar angle range from 5° to 22° and to provide a separation power for the separation of charged pions and kaons up to 3 standard deviations (s.d.) for particle momenta up to 4 GeV/c in order to cover the important particle phase space. This document describes the technical design and the expected performance of the novel PANDA Disc DIRC detector that has not been used in any other high energy physics experiment (HEP) before. The performance has been studied with Monte-Carlo simulations and various beam tests at DESY and CERN. The final design meets all PANDA requirements and guarantees suffcient safety margins.
△ Less
Submitted 29 December, 2019;
originally announced December 2019.
-
Development of J-PEM for breast cancer detection and diagnosis using positronium imaging
Authors:
Shivani,
Elżbieta Luczyńska,
Sylwia Heinze,
Paweł Moskal
Abstract:
The purpose of the presented investigations is to design, construct and establish the characteristic performance of the Jagiellonian Positron Emission Mammography(J-PEM), being designed for the detection and diagnosis of breast cancer. Its construction is based on a novel idea of PET tomography based on plastic scintillators and wavelength shifter (WLS) and a new concept of positronium imaging. We…
▽ More
The purpose of the presented investigations is to design, construct and establish the characteristic performance of the Jagiellonian Positron Emission Mammography(J-PEM), being designed for the detection and diagnosis of breast cancer. Its construction is based on a novel idea of PET tomography based on plastic scintillators and wavelength shifter (WLS) and a new concept of positronium imaging. We have prepared a simulation program based on Monte Carlo methods for optimizing the geometry and material of the J-PEM prototype. Here we present the first results from the simulations and a brief review of the state of art of breast imaging modalities and their characteristics motivating our investigation.
△ Less
Submitted 9 December, 2019;
originally announced December 2019.
-
Estimating relationship between the Time Over Threshold and energy loss by photons in plastic scintillators used in the J-PET scanner
Authors:
S. Sharma,
J. Chhokar,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
J. Gajewski,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
R. Del Grande,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemien,
E. Kubicz,
M. Mohammed
, et al. (14 additional authors not shown)
Abstract:
Time-Over-Threshold (TOT) technique is being used widely due to its implications in develo** the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration methods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in…
▽ More
Time-Over-Threshold (TOT) technique is being used widely due to its implications in develo** the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration methods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in J-PET tomograph which is build from plastic scintillators characterized by fast light signals. The drawback in adopting this technique is lying in the non-linear correlation between input energy loss and TOT of the signal. The main motivation behind this work is to develop the relationship between TOT and energy loss and validate it with the J-PET tomograph.
The experiment was performed using the $^{22}$Na beta emitter source placed in the center of the J-PET tomograph. One can obtain primary photons of two different energies: 511 keV photon from the annihilation of positron (direct annihilation or through the formation of para-Positronim atom or pick-off process of ortho-Positronium atoms), and 1275 keV prompt photon. This allows to study the correlation between TOT values and energy loss for energy range up to 1000 keV. As the photon interacts dominantly via Compton scattering inside the plastic scintillator, there is no direct information of primary photon energy. However, using the J-PET geometry one can measure the scattering angle of the interacting photon. Since, $^{22}$Na source emits photons of two different energies, it is required to know unambiguously the energy of incident photons and its corresponding scattering angle for the estimation of energy deposition. In this work, the relationship between Time Over Threshold and energy loss by interacting photons inside the plastic scintillators used in J-PET scanner is established for a energy deposited range 100-1000 keV
△ Less
Submitted 27 November, 2019;
originally announced November 2019.
-
Performance assessment of the 2$γ$ positronium imaging with the total-body PET scanners
Authors:
P. Moskal,
D. Kisielewska,
Z. Bura,
C. Chhokar,
C. Curceanu,
E. Czerwiński,
M. Dadgar 1,
K. Dulski,
J. Gajewski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
A. Kamińska,
Ł. Kapłon,
H. Karimi,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemień,
T. Kozik,
E. Kubicz,
P. Małczak
, et al. (17 additional authors not shown)
Abstract:
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image ba…
▽ More
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include: (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using $2γ$ annihilations, and (ii) estimation of the spatial and time resolution of the ortho-positronium image as a function of the coincidence resolving time (CRT) of the scanner. Simulations are conducted assuming that radiopharmaceutical is labelled with $^{44}Sc$ isotope emitting one positron and one prompt gamma. The image is reconstructed on the basis of triple coincidence events. The ortho-positronium lifetime spectrum is determined for each voxel of the image. Calculations were performed for cases of total-body detectors build of (i) LYSO scintillators as used in the EXPLORER PET, and (ii) plastic scintillators as anticipated for the cost-effective total-body J-PET scanner. To assess the spatial and time resolution the three cases were considered assuming that CRT is equal to 140ps, 50ps and 10ps. The estimated total-body PET sensitivity for the registration and selection of image forming triple coincidences is larger by a factor of 12.2 (for LYSO PET) and by factor of 4.7 (for plastic PET) with respect to the sensitivity for the standard $2γ$ imaging by LYSO PET scanners with AFOV=20cm.
△ Less
Submitted 15 November, 2019;
originally announced November 2019.
-
Investigations on physical and biological range uncertainties in Krakow proton beam therapy centre
Authors:
Antoni Rucinski,
Jakub Baran,
Giuseppe Battistoni,
Agnieszka Chrostowska,
Marco Durante,
Jan Gajewski,
Magdalena Garbacz,
Kamil Kisielewicz,
Nils Krah,
Vincenzo Patera,
Monika Pawlik-Niedźwiecka,
Ilaria Rinaldi,
Bozena Rozwadowska-Bogusz,
Emanuele Scifoni,
Agata Skrzypek,
Francesco Tommasino,
Angelo Schiavi,
Pawel Moskal
Abstract:
Physical and biological range uncertainties limit the clinical potential of Proton Beam Therapy (PBT). In this proceedings, we report on two research projects, which we are conducting in parallel and which both tackle the problem of range uncertainties. One aims at develo** software tools and the other at develo** detector instrumentation. Regarding the first, we report on our development and…
▽ More
Physical and biological range uncertainties limit the clinical potential of Proton Beam Therapy (PBT). In this proceedings, we report on two research projects, which we are conducting in parallel and which both tackle the problem of range uncertainties. One aims at develo** software tools and the other at develo** detector instrumentation. Regarding the first, we report on our development and pre-clinical application of a GPU-accelerated Monte Carlo (MC) simulation toolkit Fred. Concerning the letter, we report on our investigations of plastic scintillator based PET detectors for particle therapy delivery monitoring. We study the feasibility of Jagiellonian-PET detector technology for proton beam therapy range monitoring by means of MC simulations of the $β^+$ activity induced in a phantom by proton beams and present preliminary results of PET image reconstruction. Using a GPU-accelerated Monte Carlo simulation toolkit Fred and plastic scintillator based PET detectors we aim to improve patient treatment quality with protons.
△ Less
Submitted 25 October, 2019;
originally announced October 2019.
-
Monte Carlo N-Particle simulations of an underwater chemical threats detection system using neutron activation analysis
Authors:
P. Sibczyński,
M. Silarski,
O. Bezshyyko,
V. Ivanyan,
E. Kubicz,
Sz. Niedźwiecki,
P. Moskal,
J. Raj,
S. Sharma,
O. Trofimiuk
Abstract:
In this paper we present Monte Carlo N-Particle (MCNP) simulations of the system for underwater threat detection using neutron activation analysis developed in the SABAT project. The simulated system is based on a D-T neutron generator emitting 14~MeV neutrons without associated $α$ particle detection and equipped with a LaBr$_3$:Ce scintillation detector offering superior energy resolution and al…
▽ More
In this paper we present Monte Carlo N-Particle (MCNP) simulations of the system for underwater threat detection using neutron activation analysis developed in the SABAT project. The simulated system is based on a D-T neutron generator emitting 14~MeV neutrons without associated $α$ particle detection and equipped with a LaBr$_3$:Ce scintillation detector offering superior energy resolution and allowing for precise identification of activation $γ$ quanta. The performed simulations show that using the neutron activation analysis method with the designed geometry we are able to identify $γ$-rays from hydrogen, carbon, sulphur and chlorine originating from mustard gas in a sea water environment. Our results show that the most efficient way of mustard gas detection is to compare the integral peak ratio for Cl and H.
△ Less
Submitted 6 October, 2019;
originally announced October 2019.
-
Simulation studies of annihilation-photon's polarisation via Compton scattering with the J-PET tomograph
Authors:
N. Krawczyk,
B. C. Hiesmayr,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
J. Raj
, et al. (10 additional authors not shown)
Abstract:
J-PET is the first positron-emission tomograph (PET) constructed from plastic scintillators. It was optimized for the detection of photons from electron-positron annihilation. Such photons, having an energy of 511 keV, interact with electrons in plastic scintillators predominantly via the Compton effect. Compton scattering is at most probable at an angle orthogonal to the electric field vector of…
▽ More
J-PET is the first positron-emission tomograph (PET) constructed from plastic scintillators. It was optimized for the detection of photons from electron-positron annihilation. Such photons, having an energy of 511 keV, interact with electrons in plastic scintillators predominantly via the Compton effect. Compton scattering is at most probable at an angle orthogonal to the electric field vector of the interacting photon. Thus registration of multiple photon scatterings with J-PET enables to determine the polarization of the annihilation photons. In this contribution we present estimates on the physical limitation in the accuracy of the polarization determination of $511$~keV photons with the J-PET detector.
△ Less
Submitted 18 January, 2019;
originally announced January 2019.
-
Feasibility studies of the polarization of photons beyond the optical wavelength regime with the J-PET detector
Authors:
P. Moskal,
N. Krawczyk,
B. C. Hiesmayr,
M. Bała,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. Jasińska,
K. Kacprzak,
L. Kapłon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
J. Raj,
Z. Rudy
, et al. (9 additional authors not shown)
Abstract:
J-PET is a detector optimized for registration of photons from the electron-positron annihilation via plastic scintillators where photons interact predominantly via Compton scattering. Registration of both primary and scattered photons enables to determinate the linear polarization of the primary photon on the event by event basis with a certain probability. Here we present quantitative results on…
▽ More
J-PET is a detector optimized for registration of photons from the electron-positron annihilation via plastic scintillators where photons interact predominantly via Compton scattering. Registration of both primary and scattered photons enables to determinate the linear polarization of the primary photon on the event by event basis with a certain probability. Here we present quantitative results on the feasibility of such polarization measurements of photons from the decay of positronium with the J-PET and explore the physical limitations for the resolution of the polarization determination of 511 keV photons via Compton scattering. For scattering angles of about 82 deg (where the best contrast for polarization measurement is theoretically predicted) we find that the single event resolution for the determination of the polarization is about 40 deg (predominantly due to properties of the Compton effect). However, for samples larger than ten thousand events the J-PET is capable of determining relative average polarization of these photons with the precision of about few degrees. The obtained results open new perspectives for studies of various physics phenomena such as quantum entanglement and tests of discrete symmetries in decays of positronium and extend the energy range of polarization measurements by five orders of magnitude beyond the optical wavelength regime.
△ Less
Submitted 27 September, 2018;
originally announced September 2018.
-
Commissioning of the J-PET detector in view of the positron annihilation lifetime spectroscopy
Authors:
K. Dulski,
C. Curceanu,
E. Czerwiński,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak Ł. Kapłon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemień,
T. Kozik,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
J. Raj,
K. Rakoczy,
Z. Rudy
, et al. (8 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET device built from plastic scintillators. It is a multi-purpose detector designed for medical imaging and for studies of properties of positronium atoms in porous matter and in living organisms. In this article we report on the commissioning of the J-PET detector in view of studies of positronium decays. We present results of ana…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET device built from plastic scintillators. It is a multi-purpose detector designed for medical imaging and for studies of properties of positronium atoms in porous matter and in living organisms. In this article we report on the commissioning of the J-PET detector in view of studies of positronium decays. We present results of analysis of the positron lifetime measured in the porous polymer. The obtained results prove that J-PET is capable of performing simultaneous imaging of the density distribution of annihilation points as well as positron annihilation lifetime spectroscopy.
△ Less
Submitted 12 September, 2018;
originally announced September 2018.
-
A feasibility study of the time reversal violation test based on polarization of annihilation photons from the decay of ortho-Positronium with the J-PET detector
Authors:
J. Raj,
A. Gajos,
C. Curceanu,
E. Czerwinski,
K. Dulski,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
L. Kaplon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemien,
E. Kubicz,
M. Mohammed,
Sz. Niedzwiecki,
M. Palka,
M. Pawlik-Niedzwiecka,
L. Raczynski,
K. Rakoczy
, et al. (9 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) is a novel de- vice being developed at Jagiellonian University in Krakow, Poland based on or- ganic scintillators. J-PET is an axially symmetric and high acceptance scanner that can be used as a multi-purpose detector system. It is well suited to pur- sue tests of discrete symmetries in decays of positronium in addition to medical imaging. J-PET…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) is a novel de- vice being developed at Jagiellonian University in Krakow, Poland based on or- ganic scintillators. J-PET is an axially symmetric and high acceptance scanner that can be used as a multi-purpose detector system. It is well suited to pur- sue tests of discrete symmetries in decays of positronium in addition to medical imaging. J-PET enables the measurement of both momenta and the polarization vectors of annihilation photons. The latter is a unique feature of the J-PET detector which allows the study of time reversal symmetry violation operator which can be constructed solely from the annihilation photons momenta before and after the scattering in the detector.
△ Less
Submitted 4 September, 2018;
originally announced September 2018.
-
Estimating the NEMA characteristics of the J-PET tomograph using the GATE package
Authors:
Paweł Kowalski,
Wojciech Wiślicki,
Roman Shopa,
Lech Raczyński,
Konrad Klimaszewski,
Catalina Curceanu,
Eryk Czerwiński,
Kamil Dulski,
Aleksander Gajos,
Marek Gorgol,
Neha Gupta-Sharma,
Beatrix Hiesmayr,
Bożena Jasińska,
Łukasz Kapłon,
Daria Kamińska,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Ewelina Kubicz,
Muhsin Mohammed,
Szymon Niedzwiecki,
Marek Pałka,
Monika Pawlik-Niedźwiecka,
Juhi Raj,
Kamil Rakoczy
, et al. (8 additional authors not shown)
Abstract:
The novel whole-body PET system based on plastic scintillators is developed by the {J-PET} Collaboration. It consists of plastic scintillator strips arranged axially in the form of a cylinder, allowing the cost-effective construction of the total-body PET. In order to determine properties of the scanner prototype and optimize its geometry, advanced computer simulations using the GATE software were…
▽ More
The novel whole-body PET system based on plastic scintillators is developed by the {J-PET} Collaboration. It consists of plastic scintillator strips arranged axially in the form of a cylinder, allowing the cost-effective construction of the total-body PET. In order to determine properties of the scanner prototype and optimize its geometry, advanced computer simulations using the GATE software were performed.
The spatial resolution, the sensitivity, the scatter fraction and the noise equivalent count rate were estimated according to the NEMA norm as a function of the length of the tomograph, number of the detection layers, diameter of the tomographic chamber and for various types of the applied readout. For the single-layer geometry with the diameter of 85 cm, strip length of 100 cm, cross-section of 4 mm x 20 mm and silicon photomultipliers with the additional layer of wavelength shifter as the readout, the spatial resolution (FWHM) in the centre of the scanner is equal to 3 mm (radial, tangential) and 6 mm (axial). For the analogous double-layer geometry with the same readout, diameter and scintillator length, with the strip cross-section of 7 mm x 20 mm, the NECR peak of 300 kcps was reached at 40 kBq/cc activity concentration, the scatter fraction is estimated to about 35% and the sensitivity at the centre amounts to 14.9 cps/kBq. Sensitivity profiles were also determined.
△ Less
Submitted 1 August, 2018;
originally announced August 2018.
-
Evaluation of Single-Chip, Real-Time Tomographic Data Processing on FPGA - SoC Devices
Authors:
G. Korcyl,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
B. Flak,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
M. Kajetanowicz,
D. Kisielewska,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pawlik-Niedźwiecka,
M. Pałka,
L. Raczyński,
P. Rajda
, et al. (14 additional authors not shown)
Abstract:
A novel approach to tomographic data processing has been developed and evaluated using the Jagiellonian PET (J-PET) scanner as an example. We propose a system in which there is no need for powerful, local to the scanner processing facility, capable to reconstruct images on the fly. Instead we introduce a Field Programmable Gate Array (FPGA) System-on-Chip (SoC) platform connected directly to data…
▽ More
A novel approach to tomographic data processing has been developed and evaluated using the Jagiellonian PET (J-PET) scanner as an example. We propose a system in which there is no need for powerful, local to the scanner processing facility, capable to reconstruct images on the fly. Instead we introduce a Field Programmable Gate Array (FPGA) System-on-Chip (SoC) platform connected directly to data streams coming from the scanner, which can perform event building, filtering, coincidence search and Region-Of-Response (ROR) reconstruction by the programmable logic and visualization by the integrated processors. The platform significantly reduces data volume converting raw data to a list-mode representation, while generating visualization on the fly.
△ Less
Submitted 26 July, 2018;
originally announced July 2018.
-
Feasibility study of the positronium imaging with the J-PET tomograph
Authors:
P. Moskal,
D. Kisielewska,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
M. Gorgol,
B. Hiesmayr,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
G. Korcyl,
P. Kowalski,
W. Krzemień,
T. Kozik,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
J. Raj,
S. Sharma,
Shivani,
R. Y. Shopa
, et al. (5 additional authors not shown)
Abstract:
A detection system of the conventional PET tomograph is set-up to record data from e+ e- annihilation into two photons with energy of 511 keV, and it gives information on the density distribution of a radiopharmaceutical in the body of the object. In this paper we explore the possibility of performing the three gamma photons imaging based on ortho-positronium annihilation, as well as the possibili…
▽ More
A detection system of the conventional PET tomograph is set-up to record data from e+ e- annihilation into two photons with energy of 511 keV, and it gives information on the density distribution of a radiopharmaceutical in the body of the object. In this paper we explore the possibility of performing the three gamma photons imaging based on ortho-positronium annihilation, as well as the possibility of positronium mean lifetime imaging with the J-PET tomograph constructed from plastic scintillators. For this purposes simulations of the ortho-positronium formation and its annihilation into three photons were performed taking into account distributions of photons' momenta as predicted by the theory of quantum electrodynamics and the response of the J-PET tomograph. In order to test the proposed ortho-positronium lifetime image reconstruction method, we concentrate on the decay of the ortho-positronium into three photons and applications of radiopharmaceuticals labeled with isotopes emitting a prompt gamma quantum. The proposed method of imaging is based on the determination of hit-times and hit-positions of registered photons which enables the reconstruction of the time and position of the annihilation point as well as the lifetime of the ortho-positronium on an event-by-event basis. We have simulated the production of the positronium in a cylindrical phantom composed of a set of different materials in which the ortho-positronium lifetime varied from 2.0 ns to 3.0 ns, as expected for ortho-positronium created in the human body. The presented reconstruction method for total-body J-PET like detector allows to achieve a mean lifetime resolution of about 40 ps. Recent Positron Annihilation Lifetime Spectroscopy measurements of cancerous and healthy uterine tissues show that this sensitivity may allow to study the morphological changes in cell structures.
△ Less
Submitted 15 January, 2019; v1 submitted 29 May, 2018;
originally announced May 2018.
-
Feasibility study of the time reversal symmetry tests in decays of metastable positronium atoms with the J-PET detector
Authors:
A. Gajos,
C. Curceanu,
E. Czerwiński,
K. Dulski,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
D. Kisielewska,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
J. Raj,
Z. Rudy,
S. Sharma,
Shivani
, et al. (7 additional authors not shown)
Abstract:
This article reports on the feasibility of testing of the symmetry under reversal in time in a purely leptonic system constituted by positronium atoms using the J-PET detector. The present state of T symmetry tests is discussed with an emphasis on the scarcely explored sector of leptonic systems. Two possible strategies of searching for manifestations of T violation in non-vanishing angular correl…
▽ More
This article reports on the feasibility of testing of the symmetry under reversal in time in a purely leptonic system constituted by positronium atoms using the J-PET detector. The present state of T symmetry tests is discussed with an emphasis on the scarcely explored sector of leptonic systems. Two possible strategies of searching for manifestations of T violation in non-vanishing angular correlations of final state observables in the decays of metastable triplet states of positronium available with J-PET are proposed and discussed. Results of a pilot measurement with J-PET and assessment of its performance in reconstruction of three-photon decays are shown along with an analysis of its impact on the sensitivity of the detector for the determination of T -violation sensitive observables.
△ Less
Submitted 19 April, 2018;
originally announced April 2018.
-
Towards 2+1 photon tomography: Energy-based selection of two 511 keV photons and a prompt photon with the J-PET scanner
Authors:
R. Masełek,
W. Krzemien,
K. Klimaszewski,
L. Raczyński,
P. Kowalski,
R. Shopa,
W. Wiślicki,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. Hiesmayr,
B. Jasińska,
D. Kisielewska,
G. Korcyl,
T. Kozik,
N. Krawczyk,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedźwiecka,
S. Niedźwiecki,
M. Pałka
, et al. (9 additional authors not shown)
Abstract:
The possibility to separate signals caused by 511 keV photons created in annihilation of electron-positron pairs and the so-called prompt photons from nuclei de- excitation is investigated. It could potentially be used to improve the quality of reconstructed images in the J-PET scanner in 2+1 photon tomography. Firstly, a research is conducted for several radioisotopes that decay via \b{eta}+ deca…
▽ More
The possibility to separate signals caused by 511 keV photons created in annihilation of electron-positron pairs and the so-called prompt photons from nuclei de- excitation is investigated. It could potentially be used to improve the quality of reconstructed images in the J-PET scanner in 2+1 photon tomography. Firstly, a research is conducted for several radioisotopes that decay via \b{eta}+ decay followed by de-excitation of an excited nucleus. Efficiency, purity and false positive rate are calculated for each isotope as a function of energy deposited threshold, with a hypothesis that signals caused by 511 keV photons deposit smaller values of energy than 1 z 13the selected threshold, while prompt photons deposit larger energy than the threshold. Analysis of the results accompanied with physical properties of radioisotopes suggests using 44 Sc, which is the most promising candidate for medical applications. With the use of GATE and J-POS simulation software, in-phantom scattering was introduced and the best energy deposited threshold value was estimated to be approximately 375 keV. It corresponds to almost 100% efficiency for 511 keV signals, 75% purity for 511 keV photons, and approximately 70% efficiency and purity for prompt photons.
△ Less
Submitted 2 March, 2018;
originally announced March 2018.
-
Underwater detection of dangerous substances: status the SABAT project
Authors:
M. Silarski,
P. Sibczyński,
Sz. Niedźwiecki,
S. Sharma,
J. Raj,
P. Moskal
Abstract:
The Neutron Activation Analysis (NAA) plays an exceptional role in the modern nuclear engineering, especially in detection of hazardous substances. However, in the aquatic environment, there are still many problems to be solved for effective usage of this technique. We present status of SABAT (Stoichiometry Analysis By Activation Techniques), one of the projects aiming at construction of an underw…
▽ More
The Neutron Activation Analysis (NAA) plays an exceptional role in the modern nuclear engineering, especially in detection of hazardous substances. However, in the aquatic environment, there are still many problems to be solved for effective usage of this technique. We present status of SABAT (Stoichiometry Analysis By Activation Techniques), one of the projects aiming at construction of an underwater device for non-invasive threat detection based on the NAA.
△ Less
Submitted 21 November, 2017;
originally announced November 2017.
-
Human Tissues Investigation Using PALS Technique
Authors:
B. Jasińska,
B. Zgardzińska,
G. Chołubek,
M. Gorgol,
K. Wiktor,
K. Wysogld,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
B. Hiesmayr,
B. Jodłowska-Jędrych,
D. Kamińska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedźwiecka,
S. Niedźwiecki,
M. Pałka
, et al. (12 additional authors not shown)
Abstract:
Samples of uterine leiomyomatis and normal tissues taken from patients after surgery were investigated using the Positron Annihilation Lifetime Spectroscopy (PALS). Significant differences in all PALS parameters between normal and diseased tissues were observed. For all studied patients, it was found that the values of the free annihilation and ortho-positronium lifetime are larger for the tumorou…
▽ More
Samples of uterine leiomyomatis and normal tissues taken from patients after surgery were investigated using the Positron Annihilation Lifetime Spectroscopy (PALS). Significant differences in all PALS parameters between normal and diseased tissues were observed. For all studied patients, it was found that the values of the free annihilation and ortho-positronium lifetime are larger for the tumorous tissues than for the healthy ones. For most of the patients, the intensity of the free annihilation and ortho-positronium annihilation was smaller for the tumorous than for the healthy tissues. For the first time, in this kind of studies, the $3γ$ fraction of positron annihilation was determined to describe changes in the tissue porosity during morphologic alteration.
△ Less
Submitted 13 November, 2017;
originally announced November 2017.
-
Introduction of total variation regularization into filtered backprojection algorithm
Authors:
L. Raczyński,
W. Wiślicki,
K. Klimaszewski,
W. Krzemień,
P. Kowalski,
R. Shopa,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. Hiesmayr,
B. Jasińska,
D. Kisielewska-Kamińska,
G. Korcyl,
T. Kozik,
N. Krawczyk,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedźwiecka,
S. Niedźwiecki,
M. Pałka,
Z. Rudy
, et al. (8 additional authors not shown)
Abstract:
In this paper we extend the state-of-the-art filtered backprojection (FBP) method with application of the concept of Total Variation regularization. We compare the performance of the new algorithm with the most common form of regularizing in the FBP image reconstruction via apodizing functions. The methods are validated in terms of cross-correlation coefficient between reconstructed and real image…
▽ More
In this paper we extend the state-of-the-art filtered backprojection (FBP) method with application of the concept of Total Variation regularization. We compare the performance of the new algorithm with the most common form of regularizing in the FBP image reconstruction via apodizing functions. The methods are validated in terms of cross-correlation coefficient between reconstructed and real image of radioactive tracer distribution using standard Derenzo-type phantom. We demonstrate that the proposed approach results in higher cross-correlation values with respect to the standard FBP method.
△ Less
Submitted 31 October, 2017;
originally announced November 2017.
-
J-PET: a new technology for the whole-body PET imaging
Authors:
S. Niedźwiecki,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasińska,
Ł. Kapłon,
D. Kisielewska-Kamińska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedźwiecka,
M. Pałka,
L. Raczyński,
Z. Rudy,
N. G. Sharma,
S. Sharma
, et al. (8 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET built from plastic scintillators. J-PET prototype consists of 192 detection modules arranged axially in three layers forming a cylindrical diagnostic chamber with the inner diameter of 85 cm and the axial field-of-view of 50 cm. An axial arrangement of long strips of plastic scintillators, their small light attenuation, superior…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET built from plastic scintillators. J-PET prototype consists of 192 detection modules arranged axially in three layers forming a cylindrical diagnostic chamber with the inner diameter of 85 cm and the axial field-of-view of 50 cm. An axial arrangement of long strips of plastic scintillators, their small light attenuation, superior timing properties, and relative ease of the increase of the axial field-of-view opens promising perspectives for the cost effective construction of the whole-body PET scanner, as well as construction of MR and CT compatible PET inserts. Present status of the development of the J-PET tomograph will be presented and discussed.
△ Less
Submitted 31 October, 2017;
originally announced October 2017.
-
Commissioning of the J-PET detector for studies of decays of positronium atoms
Authors:
E. Czerwiński,
K. Dulski,
P. Białas,
C. Curceanu,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasińska,
D. Kisielewska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
Z. Rudy,
N. G. Sharma,
S. Sharma,
R. Y. Shopa
, et al. (7 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) is a detector for medical imaging of the whole human body as well as for physics studies involving detection of electron-positron annihilation into photons. J-PET has high angular and time resolution and allows for measurement of spin of the positronium and the momenta and polarization vectors of annihilation quanta. In this article, we present…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) is a detector for medical imaging of the whole human body as well as for physics studies involving detection of electron-positron annihilation into photons. J-PET has high angular and time resolution and allows for measurement of spin of the positronium and the momenta and polarization vectors of annihilation quanta. In this article, we present the potential of the J-PET system for background rejection in the decays of positronium atoms.
△ Less
Submitted 13 November, 2017; v1 submitted 30 October, 2017;
originally announced October 2017.
-
Three-dimensional image reconstruction in J-PET using Filtered Back Projection method
Authors:
R. Y. Shopa,
K. Klimaszewski,
P. Kowalski,
W. Krzemień,
L. Raczyński,
W. Wiślicki,
P. Białas,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. Hiesmayr,
B. Jasińska,
D. Kisielewska-Kamińska,
G. Korcyl,
T. Kozik,
N. Krawczyk,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedźwiecka,
S. Niedźwiecki,
M. Pałka,
Z. Rudy
, et al. (8 additional authors not shown)
Abstract:
We present a method and preliminary results of the image reconstruction in the Jagiellonian PET tomograph. Using GATE (Geant4 Application for Tomographic Emission), interactions of the 511 keV photons with a cylindrical detector were generated. Pairs of such photons, flying back-to-back, originate from e+e- annihilations inside a 1-mm spherical source. Spatial and temporal coordinates of hits were…
▽ More
We present a method and preliminary results of the image reconstruction in the Jagiellonian PET tomograph. Using GATE (Geant4 Application for Tomographic Emission), interactions of the 511 keV photons with a cylindrical detector were generated. Pairs of such photons, flying back-to-back, originate from e+e- annihilations inside a 1-mm spherical source. Spatial and temporal coordinates of hits were smeared using experimental resolutions of the detector. We incorporated the algorithm of the 3D Filtered Back Projection, implemented in the STIR and TomoPy software packages, which differ in approximation methods. Consistent results for the Point Spread Functions of ~5/7,mm and ~9/20, mm were obtained, using STIR, for transverse and longitudinal directions, respectively, with no time of flight information included.
△ Less
Submitted 30 October, 2017;
originally announced October 2017.
-
Analysis procedure of the positronium lifetime spectra for the J-PET detector
Authors:
K. Dulski,
B. Zgardzinska,
P. Bialas,
C. Curceanu E. Czerwinski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
D. Kisielewska-Kaminska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemiez,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedzwiecka,
S. Niedzwiecki,
M. Palka,
L. Raczynski,
J. Raj,
Z. Rudy,
N. G. Sharma,
S. Sharma
, et al. (8 additional authors not shown)
Abstract:
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics…
▽ More
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics and medicine. In this contribution we present a computer program that is compatible with the J-PET software. We compare its performance with the standard program LT 9.0 by using PALS data from hexane measurements at different temperatures. Our program is based on an iterative procedure, and our fits prove that it performs as good as LT 9.0.
△ Less
Submitted 16 October, 2017;
originally announced October 2017.