Measurement of the radon concentration in purified water in the Super-Kamiokande IV detector
Authors:
Y. Nakano,
T. Hokama,
M. Matsubara,
M. Miwa,
M. Nakahata,
T. Nakamura,
H. Sekiya,
Y. Takeuchi,
S. Tasaka,
R. A. Wendell
Abstract:
The radioactive noble gas radon can be a serious background source in the underground particle physics experiments studying processes that deposit energy comparable to its decay products. Low energy solar neutrino measurements at Super-Kamiokande suffer from these backgrounds and therefore require precise characterization of the radon concentration in the detector's ultra-pure water. For this purp…
▽ More
The radioactive noble gas radon can be a serious background source in the underground particle physics experiments studying processes that deposit energy comparable to its decay products. Low energy solar neutrino measurements at Super-Kamiokande suffer from these backgrounds and therefore require precise characterization of the radon concentration in the detector's ultra-pure water. For this purpose, we have developed a measurement system consisting of a radon extraction column, a charcoal trap, and a radon detector. In this article we discuss the design, calibration, and performance of the radon extraction column. We also describe the design of the measurement system and evaluate its performance, including its background. Using this system we measured the radon concentration in Super-Kamiokande's water between May 2014 and October 2015. The measured radon concentration in the supply lines of the water circulation system was $1.74\pm0.14~\mathrm{mBq/m^{3}}$ and in the return line was $9.06\pm0.58~\mathrm{mBq/m^{3}}$. Water sampled from the center region of the detector itself had a concentration of $<0.23~\mathrm{mBq/m^{3}}$ ($95\%$ C.L.) and water sampled from the bottom region of the detector had a concentration of $2.63\pm0.22~\mathrm{mBq/m^{3}}$.
△ Less
Submitted 26 June, 2020; v1 submitted 9 October, 2019;
originally announced October 2019.
Measurement of Radon Concentration in Super-Kamiokande's Buffer Gas
Authors:
Y. Nakano,
H. Sekiya,
S. Tasaka,
Y. Takeuchi,
R. A. Wendell,
M. Matsubara,
M. Nakahata
Abstract:
To precisely measure radon concentrations in purified air supplied to the Super-Kamiokande detector as a buffer gas, we have developed a highly sensitive radon detector with an intrinsic background as low as 0.33$\pm$0.07 mBq/m$^{3}$. In this article, we discuss the construction and calibration of this detector as well as results of its application to the measurement and monitoring of the buffer g…
▽ More
To precisely measure radon concentrations in purified air supplied to the Super-Kamiokande detector as a buffer gas, we have developed a highly sensitive radon detector with an intrinsic background as low as 0.33$\pm$0.07 mBq/m$^{3}$. In this article, we discuss the construction and calibration of this detector as well as results of its application to the measurement and monitoring of the buffer gas layer above Super-Kamiokande. In March 2013, the chilled activated charcoal system used to remove radon in the input buffer gas was upgraded. After this improvement, a dramatic reduction in the radon concentration of the supply gas down to 0.08 $\pm$ 0.07 mBq/m$^{3}$. Additionally, the Rn concentration of the in-situ buffer gas has been measured 28.8$\pm$1.7 mBq/m$^{3}$ using the new radon detector. Based on these measurements we have determined that the dominant source of Rn in the buffer gas arises from contamination from the Super-Kamiokande tank itself.
△ Less
Submitted 23 April, 2017;
originally announced April 2017.