Performance and Moli`ere radius measurements using a compact prototype of LumiCal in an electron test beam
Authors:
H. Abramowicz,
A. Abusleme,
K. Afanaciev,
Y. Benhammou,
O. Borysov,
M. Borysova,
I. Bozovic- Jelisavcic,
W. Daniluk,
D. Dannheim,
M. Demichev,
K. Elsener,
M. Firlej,
E. Firu,
T. Fiutowski,
V. Ghenescu,
M. Gostkin,
M. Hempelb,
H. Henschel,
M. Idzik,
A. Ignatenkoc,
A. Ishikawa,
A. Joffe,
G. Kacarevic,
S. Kananov,
O. Karachebanb
, et al. (29 additional authors not shown)
Abstract:
A new design of a detector plane of sub-millimetre thickness for an electromagnetic sampling calorimeter is presented. It is intended to be used in the luminometers LumiCal and BeamCal in future linear $e^+e^-$ collider experiments. The detector planes were produced utilising novel connectivity scheme technologies. They were installed in a compact prototype of the calorimeter and tested at DESY wi…
▽ More
A new design of a detector plane of sub-millimetre thickness for an electromagnetic sampling calorimeter is presented. It is intended to be used in the luminometers LumiCal and BeamCal in future linear $e^+e^-$ collider experiments. The detector planes were produced utilising novel connectivity scheme technologies. They were installed in a compact prototype of the calorimeter and tested at DESY with an electron beam of energy 1-5 GeV. The performance of a prototype of a compact LumiCal comprising eight detector planes was studied. The effective Moli`ere radius at 5 GeV was determined to be (8.1 +/- 0.1 (stat) +/- 0.3 (syst)) mm, a value well reproduced by the Monte Carlo (MC) simulation (8.4 +/- 0.1) mm. The dependence of the effective Moli`ere radius on the electron energy in the range 1-5 GeV was also studied. Good agreement was obtained between data and MC simulation.
△ Less
Submitted 25 October, 2019; v1 submitted 29 December, 2018;
originally announced December 2018.