Statistically Adaptive Filtering for Low Signal Correction in X-ray Computed Tomography
Authors:
Obaidullah Rahman,
Ken D. Sauer,
Charles A. Bouman,
Roman Melnyk,
Brian Nett
Abstract:
Low x-ray dose is desirable in x-ray computed tomographic (CT) imaging due to health concerns. But low dose comes with a cost of low signal artifacts such as streaks and low frequency bias in the reconstruction. As a result, low signal correction is needed to help reduce artifacts while retaining relevant anatomical structures.
Low signal can be encountered in cases where sufficient number of ph…
▽ More
Low x-ray dose is desirable in x-ray computed tomographic (CT) imaging due to health concerns. But low dose comes with a cost of low signal artifacts such as streaks and low frequency bias in the reconstruction. As a result, low signal correction is needed to help reduce artifacts while retaining relevant anatomical structures.
Low signal can be encountered in cases where sufficient number of photons do not reach the detector to have confidence in the recorded data. % NOTE: SNR is ratio of powers, not std. dev. X-ray photons, assumed to have Poisson distribution, have signal to noise ratio proportional to the dose, with poorer SNR in low signal areas. Electronic noise added by the data acquisition system further reduces the signal quality.
In this paper we will demonstrate a technique to combat low signal artifacts through adaptive filtration. It entails statistics-based filtering on the uncorrected data, correcting the lower signal areas more aggressively than the high signal ones. We look at local averages to decide how aggressive the filtering should be, and local standard deviation to decide how much detail preservation to apply. Implementation consists of a pre-correction step i.e. local linear minimum mean-squared error correction, followed by a variance stabilizing transform, and finally adaptive bilateral filtering. The coefficients of the bilateral filter are computed using local statistics. Results show that improvements were made in terms of low frequency bias, streaks, local average and standard deviation, modulation transfer function and noise power spectrum.
△ Less
Submitted 23 September, 2023;
originally announced September 2023.
-
An Edge Alignment-based Orientation Selection Method for Neutron Tomography
Authors:
Diyu Yang,
Shimin Tang,
Singanallur V. Venkatakrishnan,
Mohammad S. N. Chowdhury,
Yuxuan Zhang,
Hassina Z. Bilheux,
Gregery T. Buzzard,
Charles A. Bouman
Abstract:
Neutron computed tomography (nCT) is a 3D characterization technique used to image the internal morphology or chemical composition of samples in biology and materials sciences. A typical workflow involves placing the sample in the path of a neutron beam, acquiring projection data at a predefined set of orientations, and processing the resulting data using an analytic reconstruction algorithm. Typi…
▽ More
Neutron computed tomography (nCT) is a 3D characterization technique used to image the internal morphology or chemical composition of samples in biology and materials sciences. A typical workflow involves placing the sample in the path of a neutron beam, acquiring projection data at a predefined set of orientations, and processing the resulting data using an analytic reconstruction algorithm. Typical nCT scans require hours to days to complete and are then processed using conventional filtered back-projection (FBP), which performs poorly with sparse views or noisy data. Hence, the main methods in order to reduce overall acquisition time are the use of an improved sampling strategy combined with the use of advanced reconstruction methods such as model-based iterative reconstruction (MBIR). In this paper, we propose an adaptive orientation selection method in which an MBIR reconstruction on previously-acquired measurements is used to define an objective function on orientations that balances a data-fitting term promoting edge alignment and a regularization term promoting orientation diversity. Using simulated and experimental data, we demonstrate that our method produces high-quality reconstructions using significantly fewer total measurements than the conventional approach.
△ Less
Submitted 8 March, 2023; v1 submitted 1 December, 2022;
originally announced December 2022.
-
Physics-Based Iterative Reconstruction for Dual Source and Flying Focal Spot Computed Tomography
Authors:
Xiao Wang,
Robert D. MacDougall,
Peng Chen,
Charles A. Bouman,
Simon K. Warfield
Abstract:
For single source helical Computed Tomography (CT), both Filtered-Back Projection (FBP) and statistical iterative reconstruction have been investigated. However for dual source CT with flying focal spot (DS-FFS CT), statistical iterative reconstruction that accurately models the scanner geometry and physics remains unknown to researchers. Therefore, this paper presents a novel physics-based iterat…
▽ More
For single source helical Computed Tomography (CT), both Filtered-Back Projection (FBP) and statistical iterative reconstruction have been investigated. However for dual source CT with flying focal spot (DS-FFS CT), statistical iterative reconstruction that accurately models the scanner geometry and physics remains unknown to researchers. Therefore, this paper presents a novel physics-based iterative reconstruction method for DS-FFS CT and assess its image quality. Our algorithm uses precise physics models to reconstruct from the native cone-beam geometry and interleaved dual source helical trajectory of a DS-FFS CT. To do so, we construct a noise physics model to represent data acquisition noise and a prior image model to represent image noise and texture. In addition, we design forward system models to compute the locations of deflected focal spots, the dimension and sensitivity of voxels and detector units, as well as the length of intersection between X-rays and voxels. The forward system models further represent the coordinated movement between the dual sources by computing their X-ray coverage gaps and overlaps at an arbitrary helical pitch. With the above models, we reconstruct images by using an advanced Consensus Equilibrium (CE) numerical method to compute the maximum a posteriori estimate to a joint optimization problem that simultaneously fits all models. We compared our reconstruction with Siemens ADMIRE, which is the clinical standard hybrid iterative reconstruction (IR) method for DS-FFS CT, in terms of spatial resolution, noise profile and image artifacts through both phantoms and clinical datasets. Experiments show that our reconstruction has a consistently higher spatial resolution than the clinical standard hybrid IR. In addition, our reconstruction shows a reduced magnitude of image undersampling artifacts than the clinical standard.
△ Less
Submitted 25 April, 2021; v1 submitted 26 January, 2020;
originally announced January 2020.
-
Model Based Iterative Reconstruction With Spatially Adaptive Sinogram Weights for Wide-Cone Cardiac CT
Authors:
Amirkoushyar Ziabari,
Dong Hye Ye,
Lin Fu,
Somesh Srivastava,
Ken D. Sauer,
Jean-Baptist Thibault,
Charles A. Bouman
Abstract:
With the recent introduction of CT scanners with large cone angles, wide coverage detectors now provide a desirable scanning platform for cardiac CT that allows whole heart imaging in a single rotation. On these scanners, while half-scan data is strictly sufficient to produce images with the best temporal resolution, acquiring a full 360 degree rotation worth of data is beneficial for wide-cone im…
▽ More
With the recent introduction of CT scanners with large cone angles, wide coverage detectors now provide a desirable scanning platform for cardiac CT that allows whole heart imaging in a single rotation. On these scanners, while half-scan data is strictly sufficient to produce images with the best temporal resolution, acquiring a full 360 degree rotation worth of data is beneficial for wide-cone image reconstruction at negligible additional radiation dose. Applying Model-Based Iterative Reconstruction (MBIR) algorithm to the heart has shown to yield significant enhancement in image quality for cardiac CT. But imaging the heart in large cone angle geometry leads to apparently conflicting data usage considerations. On the one hand, in addition to using the fastest available scanner rotation speed, a minimal complete data set of 180 degrees plus the fan angle is typically used to minimize both cardiac and respiratory motion. On the other hand, a full 360 degree acquisition helps better handle the challenges of missing frequencies and incomplete projections associated with wide-cone half-scan data acquisition. In this paper, we develop a Spatially Adaptive sinogram Weights MBIR algorithm (SAW-MBIR) that is designed to achieve the benefits of both half and full-scan reconstructions in order to maximize temporal resolution over the heart region while providing stable results over the whole volume covered with the wide-area detector. Spatially-adaptive sinogram weights applied to each projection measurement in SAW-MBIR are designed to selectively perform backprojection from the full and half-scan portion of the sinogram based on both projection angle and reconstructed voxel location. We demonstrate with experimental results of SAW-MBIR applied to whole-heart cardiac CT clinical data that overall temporal resolution matches half-scan while full volume image quality is on par with full-scan MBIR.
△ Less
Submitted 20 December, 2018;
originally announced December 2018.
-
Phase-error estimation and image reconstruction from digital-holography data using a Bayesian framework
Authors:
Casey J. Pellizzari,
Mark F. Spencer,
Charles A. Bouman
Abstract:
The estimation of phase errors from digital-holography data is critical for applications such as imaging or wave-front sensing. Conventional techniques require multiple i.i.d. data and perform poorly in the presence of high noise or large phase errors. In this paper we propose a method to estimate isoplanatic phase errors from a single data realization. We develop a model-based iterative reconstru…
▽ More
The estimation of phase errors from digital-holography data is critical for applications such as imaging or wave-front sensing. Conventional techniques require multiple i.i.d. data and perform poorly in the presence of high noise or large phase errors. In this paper we propose a method to estimate isoplanatic phase errors from a single data realization. We develop a model-based iterative reconstruction algorithm which computes the maximum a posteriori estimate of the phase and the speckle-free object reflectance. Using simulated data, we show that the algorithm is robust against high noise and strong phase errors.
△ Less
Submitted 9 June, 2017;
originally announced August 2017.
-
Single Shot Digital Holography Using Iterative Reconstruction with Alternating Updates of Amplitude and Phase
Authors:
Dennis J. Lee,
Charles A. Bouman,
Andrew M. Weiner
Abstract:
We present an image recovery approach to improve amplitude and phase reconstruction from single shot digital holograms, using iterative reconstruction with alternating updates. This approach allows the flexibility to apply different priors to amplitude and phase, improves phase reconstruction in image areas with low amplitudes, and does not require phase unwrap** for regularization. Phantom simu…
▽ More
We present an image recovery approach to improve amplitude and phase reconstruction from single shot digital holograms, using iterative reconstruction with alternating updates. This approach allows the flexibility to apply different priors to amplitude and phase, improves phase reconstruction in image areas with low amplitudes, and does not require phase unwrap** for regularization. Phantom simulations and experimental measurements of a grating sample both demonstrate that the proposed method helps to reduce noise and resolve finer features. The improved image reconstruction from this technique will benefit the many applications of digital holography.
△ Less
Submitted 2 October, 2016; v1 submitted 9 September, 2016;
originally announced September 2016.
-
A Gaussian Mixture MRF for Model-Based Iterative Reconstruction with Applications to Low-Dose X-ray CT
Authors:
Ruoqiao Zhang,
Dong Hye Ye,
Debashish Pal,
Jean-Baptiste Thibault,
Ken D. Sauer,
Charles A. Bouman
Abstract:
Markov random fields (MRFs) have been widely used as prior models in various inverse problems such as tomographic reconstruction. While MRFs provide a simple and often effective way to model the spatial dependencies in images, they suffer from the fact that parameter estimation is difficult. In practice, this means that MRFs typically have very simple structure that cannot completely capture the s…
▽ More
Markov random fields (MRFs) have been widely used as prior models in various inverse problems such as tomographic reconstruction. While MRFs provide a simple and often effective way to model the spatial dependencies in images, they suffer from the fact that parameter estimation is difficult. In practice, this means that MRFs typically have very simple structure that cannot completely capture the subtle characteristics of complex images.
In this paper, we present a novel Gaussian mixture Markov random field model (GM-MRF) that can be used as a very expressive prior model for inverse problems such as denoising and reconstruction. The GM-MRF forms a global image model by merging together individual Gaussian-mixture models (GMMs) for image patches. In addition, we present a novel analytical framework for computing MAP estimates using the GM-MRF prior model through the construction of surrogate functions that result in a sequence of quadratic optimizations. We also introduce a simple but effective method to adjust the GM-MRF so as to control the sharpness in low- and high-contrast regions of the reconstruction separately. We demonstrate the value of the model with experiments including image denoising and low-dose CT reconstruction.
△ Less
Submitted 13 June, 2016; v1 submitted 12 May, 2016;
originally announced May 2016.