Achieving ultra-low and -uniform residual magnetic fields in a very large magnetically shielded room for fundamental physics experiments
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
D. Bowles,
E. Chanel,
W. Chen,
P. -J. Chiu,
C. B. Crawford,
O. Naviliat-Cuncic,
C. B. Doorenbos,
S. Emmenegger,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch,
V. Kletzl,
J. Krempel,
B. Lauss,
T. Lefort,
A. Lejuez
, et al. (25 additional authors not shown)
Abstract:
High-precision searches for an electric dipole moment of the neutron (nEDM) require stable and uniform magnetic field environments. We present the recent achievements of degaussing and equilibrating the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute. We present the final degaussing configuration that will be used for n2EDM after numerous studies. The optim…
▽ More
High-precision searches for an electric dipole moment of the neutron (nEDM) require stable and uniform magnetic field environments. We present the recent achievements of degaussing and equilibrating the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute. We present the final degaussing configuration that will be used for n2EDM after numerous studies. The optimized procedure results in a residual magnetic field that has been reduced by a factor of two. The ultra-low field is achieved with the full magnetic-field-coil system, and a large vacuum vessel installed, both in the MSR. In the inner volume of ~1.4 m^3, the field is now more uniform and below 300 pT. In addition, the procedure is faster and dissipates less heat into the magnetic environment, which in turn, reduces its thermal relaxation time from 12 h down to ~1.5 h.
△ Less
Submitted 28 September, 2023;
originally announced September 2023.
-
A large 'Active Magnetic Shield' for a high-precision experiment
Authors:
C. Abel,
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
E. Chanel,
J. Chen,
W. Chen,
P. -J. Chiu,
C. B. Crawford,
M. Daum,
C. B. Doorenbos,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujic,
P. Harris,
K. Kirch,
V. Kletzl,
P. A. Koss,
J. Krempel
, et al. (26 additional authors not shown)
Abstract:
We present a novel Active Magnetic Shield (AMS), designed and implemented for the n2EDM experiment at the Paul Scherrer Institute. The experiment will perform a high-sensitivity search for the electric dipole moment of the neutron. Magnetic-field stability and control is of key importance for n2EDM. A large, cubic, 5m side length, magnetically shielded room (MSR) provides a passive, quasi-static s…
▽ More
We present a novel Active Magnetic Shield (AMS), designed and implemented for the n2EDM experiment at the Paul Scherrer Institute. The experiment will perform a high-sensitivity search for the electric dipole moment of the neutron. Magnetic-field stability and control is of key importance for n2EDM. A large, cubic, 5m side length, magnetically shielded room (MSR) provides a passive, quasi-static shielding-factor of about 10^5 for its inner sensitive volume. The AMS consists of a system of eight complex, feedback-controlled compensation coils constructed on an irregular grid spanned on a volume of less than 1000m^3 around the MSR. The AMS is designed to provide a stable and uniform magnetic-field environment around the MSR, while being reasonably compact. The system can compensate static and variable magnetic fields up to +-50muT (homogeneous components) and +-5muT (first-order gradients), suppressing them to a few muT in the sub-Hertz frequency range. The presented design concept and implementation of the AMS fulfills the requirements of the n2EDM experiment and can be useful for other applications, where magnetically silent environments are important and spatial constraints inhibit simpler geometrical solutions.
△ Less
Submitted 14 July, 2023;
originally announced July 2023.
-
Particle Physics at the European Spallation Source
Authors:
H. Abele,
A. Alekou,
A. Algora,
K. Andersen,
S. Baessler,
L. Barron-Palos,
J. Barrow,
E. Baussan,
P. Bentley,
Z. Berezhiani,
Y. Bessler,
A. K. Bhattacharyya,
A. Bianchi,
J. Bijnens,
C. Blanco,
N. Blaskovic Kraljevic,
M. Blennow,
K. Bodek,
M. Bogomilov,
C. Bohm,
B. Bolling,
E. Bouquerel,
G. Brooijmans,
L. J. Broussard,
O. Buchan
, et al. (154 additional authors not shown)
Abstract:
Presently under construction in Lund, Sweden, the European Spallation Source (ESS) will be the world's brightest neutron source. As such, it has the potential for a particle physics program with a unique reach and which is complementary to that available at other facilities. This paper describes proposed particle physics activities for the ESS. These encompass the exploitation of both the neutrons…
▽ More
Presently under construction in Lund, Sweden, the European Spallation Source (ESS) will be the world's brightest neutron source. As such, it has the potential for a particle physics program with a unique reach and which is complementary to that available at other facilities. This paper describes proposed particle physics activities for the ESS. These encompass the exploitation of both the neutrons and neutrinos produced at the ESS for high precision (sensitivity) measurements (searches).
△ Less
Submitted 30 January, 2024; v1 submitted 18 November, 2022;
originally announced November 2022.
-
The `n2EDM MSR' -- a very large magnetically shielded room with an exceptional performance for fundamental physics measurements
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
B. Clement,
E. Chanel,
P. -J. Chiu,
C. B. Crawford,
M. Daum,
C. B. Doorenbos,
S. Emmenegger,
A. Fratangelo,
M. Fertl,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch,
J. Krempel,
B. Lauss,
T. Lefort,
O. Naviliat-Cuncic,
D. Pais,
F. M. Piegsa
, et al. (19 additional authors not shown)
Abstract:
We present the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute which features an interior cubic volume with each side of length 2.92m, thus providing an accessible space of 25m3. The MSR has 87 openings up to 220mm diameter to operate the experimental apparatus inside, and an intermediate space between the layers for sensitive signal processing electronics.…
▽ More
We present the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute which features an interior cubic volume with each side of length 2.92m, thus providing an accessible space of 25m3. The MSR has 87 openings up to 220mm diameter to operate the experimental apparatus inside, and an intermediate space between the layers for sensitive signal processing electronics. The characterization measurements show a remanent magnetic field in the central 1m3 below 100pT, and a field below 600pT in the entire inner volume, up to 4\,cm to the walls. The quasi-static shielding factor at 0.01\,Hz measured with a sinusoidal 2muT peak-to-peak signal is about 100,000 in all three spatial directions and rises fast with frequency to reach 10^8 above 1Hz.
△ Less
Submitted 21 June, 2022;
originally announced June 2022.
-
Map** of the magnetic field to correct systematic effects in a neutron electric dipole moment experiment
Authors:
C. Abel,
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
E. Chanel,
P. -J. Chiu,
B. Clément,
C. B. Crawford,
M. Daum,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
P. Flaux,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujić,
P. G. Harris,
L. Hayen,
N. Hild,
M. Kasprzak,
K. Kirch,
P. Knowles,
H. -C. Koch
, et al. (28 additional authors not shown)
Abstract:
Experiments dedicated to the measurement of the electric dipole moment of the neutron require outstanding control of the magnetic field uniformity. The neutron electric dipole moment (nEDM) experiment at the Paul Scherrer Institute uses a 199Hg co-magnetometer to precisely monitor magnetic field variations. This co-magnetometer, in the presence of field non-uniformity, is responsible for the large…
▽ More
Experiments dedicated to the measurement of the electric dipole moment of the neutron require outstanding control of the magnetic field uniformity. The neutron electric dipole moment (nEDM) experiment at the Paul Scherrer Institute uses a 199Hg co-magnetometer to precisely monitor magnetic field variations. This co-magnetometer, in the presence of field non-uniformity, is responsible for the largest systematic effect of this measurement. To evaluate and correct that effect, offline measurements of the field non-uniformity were performed during map** campaigns in 2013, 2014 and 2017. We present the results of these campaigns, and the improvement the correction of this effect brings to the neutron electric dipole moment measurement.
△ Less
Submitted 3 May, 2022; v1 submitted 16 March, 2021;
originally announced March 2021.
-
Johnson-Nyquist Noise Effects in Neutron Electric-Dipole-Moment Experiments
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
P. -J. Chiu,
B. Clement,
C. B. Crawford,
M. Daum,
S. Emmenegger,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujić,
P. G. Harris,
K. Kirch,
P. A. Koss,
B. Lauss,
T. Lefort,
P. Mohanmurthy,
O. Naviliat-Cuncic,
D. Pais,
F. M. Piegsa,
G. Pignol,
D. Rebreyend
, et al. (15 additional authors not shown)
Abstract:
Magnetic Johnson-Nyquist noise (JNN) originating from metal electrodes, used to create a static electric field in neutron electric-dipole-moment (nEDM) experiments, may limit the sensitivity of measurements. We present here the first dedicated study on JNN applied to a large-scale long-measurement-time experiment with the implementation of a co-magnetometry. In this study, we derive surface- and v…
▽ More
Magnetic Johnson-Nyquist noise (JNN) originating from metal electrodes, used to create a static electric field in neutron electric-dipole-moment (nEDM) experiments, may limit the sensitivity of measurements. We present here the first dedicated study on JNN applied to a large-scale long-measurement-time experiment with the implementation of a co-magnetometry. In this study, we derive surface- and volume-averaged root-mean-square normal noise amplitudes at a certain frequency bandwidth for a cylindrical geometry. In addition, we model the source of noise as a finite number of current dipoles and demonstrate a method to simulate temporal and three-dimensional spatial dependencies of JNN. The calculations are applied to estimate the impact of JNN on measurements with the new apparatus, n2EDM, at the Paul Scherrer Institute. We demonstrate that the performances of the optically pumped $^{133}$Cs magnetometers and $^{199}$Hg co-magnetometers, which will be used in the apparatus, are not limited by JNN. Further, we find that in measurements deploying a co-magnetometer system, the impact of JNN is negligible for nEDM searches down to a sensitivity of $4\,\times\,10^{-28}\,e\cdot{\rm cm}$ in a single measurement; therefore, the use of economically and mechanically favored solid aluminum electrodes is possible.
△ Less
Submitted 9 July, 2021; v1 submitted 2 February, 2021;
originally announced February 2021.
-
The design of the n2EDM experiment
Authors:
N. J. Ayres,
G. Ban,
L. Bienstman,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
E. Chanel,
J. Chen,
P. -J. Chiu,
B. Clément,
C. Crawford,
M. Daum,
B. Dechenaux,
C. B. Doorenbos,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
A. Fratangelo,
P. Flaux,
D. Goupillière,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch
, et al. (36 additional authors not shown)
Abstract:
We present the design of a next-generation experiment, n2EDM, currently under construction at the ultracold neutron source at the Paul Scherrer Institute (PSI) with the aim of carrying out a high-precision search for an electric dipole moment of the neutron. The project builds on experience gained with the previous apparatus operated at PSI until 2017, and is expected to deliver an order of magnit…
▽ More
We present the design of a next-generation experiment, n2EDM, currently under construction at the ultracold neutron source at the Paul Scherrer Institute (PSI) with the aim of carrying out a high-precision search for an electric dipole moment of the neutron. The project builds on experience gained with the previous apparatus operated at PSI until 2017, and is expected to deliver an order of magnitude better sensitivity with provision for further substantial improvements. An overview is given of the experimental method and setup, the sensitivity requirements for the apparatus are derived, and its technical design is described.
△ Less
Submitted 22 January, 2021; v1 submitted 21 January, 2021;
originally announced January 2021.
-
Measurement of the permanent electric dipole moment of the neutron
Authors:
C. Abel,
S. Afach,
N. J. Ayres,
C. A. Baker,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
M. Burghoff,
E. Chanel,
Z. Chowdhuri,
P. -J. Chiu,
B. Clement,
C. B. Crawford,
M. Daum,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
P. Flaux,
B. Franke,
A. Fratangelo,
P. Geltenbort,
K. Green,
W. C. Griffith,
M. van der Grinten
, et al. (59 additional authors not shown)
Abstract:
We present the result of an experiment to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute using Ramsey's method of separated oscillating magnetic fields with ultracold neutrons (UCN). Our measurement stands in the long history of EDM experiments probing physics violating time reversal invariance. The salient features of this experiment were the use of a Hg-19…
▽ More
We present the result of an experiment to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute using Ramsey's method of separated oscillating magnetic fields with ultracold neutrons (UCN). Our measurement stands in the long history of EDM experiments probing physics violating time reversal invariance. The salient features of this experiment were the use of a Hg-199 co-magnetometer and an array of optically pumped cesium vapor magnetometers to cancel and correct for magnetic field changes. The statistical analysis was performed on blinded datasets by two separate groups while the estimation of systematic effects profited from an unprecedented knowledge of the magnetic field. The measured value of the neutron EDM is $d_{\rm n} = (0.0\pm1.1_{\rm stat}\pm0.2_{\rm sys})\times10^{-26}e\,{\rm cm}$.
△ Less
Submitted 31 January, 2020;
originally announced January 2020.