Evaluation of a method for time-of-flight, wavelength and distance calibration for neutron scattering instruments by means of a mini-chopper and standard neutron monitors
Authors:
L. Vergara,
M. Arai,
R. Woracek,
M. Olsson,
A. Quintanilla,
S. Alcock,
J. Nilsson,
K. Kanaki,
P. M. Kadletz,
O. Kirstein,
R. Hall-Wilton,
N. Tsapatsaris
Abstract:
Accurate conversion of neutron time-of-flight (TOF) to wavelength, and its uncertainty, is of fundamental importance to neutron scattering measurements. Especially in cases where instruments are highly configurable, the determination of the absolute wavelength after any change must always be performed. Inspired by the manner with which neutron spectrometers determine the absolute wavelength, we ev…
▽ More
Accurate conversion of neutron time-of-flight (TOF) to wavelength, and its uncertainty, is of fundamental importance to neutron scattering measurements. Especially in cases where instruments are highly configurable, the determination of the absolute wavelength after any change must always be performed. Inspired by the manner with which neutron spectrometers determine the absolute wavelength, we evaluate for the first time, in the author's knowledge, a commonly used method for converting TOF to neutron wavelength, the distance of a monitor from the source of neutrons and we analytically calculate the uncertainty contributions that limit the precision of the conversion. The method was evaluated at the V20 test beamline at the Helmholtz Zentrum Berlin (HZB), emulating the ESS source with a pulse of 2.86 ms length and 14 Hz repetition rate, by using a mini-chopper operated at 140 Hz, beam monitors (BMs) and data acquisition infrastructure. The mini-chopper created well-defined neutron pulses and the BM was placed at two positions, enabling the average wavelength of each of the pulses created to be determined. The used experimental setup resulted in absolute wavelength determination at the monitor positions with a $δλ_{mean} / λ_{mean}$ of $\sim$1.8% for $λ>4$ Å. With a modest increase of the distance between the reference monitor positions a $δλ_{mean} / λ_{mean}$ of below 0.5% can be achieved. Further improvements are possible by using a thinner monitor, smaller chopper disc openings and a higher rotational speed chopper. The method requires only two neutron measurements and doesn't necessitate the use of crystals or complex fitting, and could constitute a suitable addition to imaging, diffraction, reflectometers and small angle neutron scattering instruments, at spallation sources, that do not normally utilise fast choppers.
△ Less
Submitted 3 November, 2022;
originally announced November 2022.
-
Spectroscopic Neutron Imaging for Resolving Hydrogen Dynamics Changes in Battery Electrolytes
Authors:
E. R. Carreón Ruiz,
J. Lee,
J. I. Márquez Damián,
M. Strobl,
G. Burca,
R. Woracek,
M. Cochet,
M. -O. Ebert,
L. Höltschi,
P. M. Kadletz,
A. S. Tremsin,
E. Winter,
M. Zlobinski,
L. Gubler,
P. Boillat
Abstract:
We present spectroscopic neutron imaging (SNI), a bridge between imaging and scattering techniques, for the analysis of hydrogenated molecules in lithium-ion cells. The scattering information of CHn-based organic solvents and electrolytes was mapped in two-dimensional space by investigating the wavelength-dependent property of hydrogen atoms through time-of-flight imaging. Our investigation demons…
▽ More
We present spectroscopic neutron imaging (SNI), a bridge between imaging and scattering techniques, for the analysis of hydrogenated molecules in lithium-ion cells. The scattering information of CHn-based organic solvents and electrolytes was mapped in two-dimensional space by investigating the wavelength-dependent property of hydrogen atoms through time-of-flight imaging. Our investigation demonstrates a novel approach to detect physical and chemical changes in hydrogenated liquids, which extends, but not limits, the use of SNI to relevant applications in electrochemical devices, e.g., the study of electrolytes in Li-ion batteries.
△ Less
Submitted 18 July, 2022;
originally announced July 2022.
-
A multimodal operando neutron study of the phase evolution in a graphite electrode
Authors:
Monica-Elisabeta Lăcătuşu,
Luise Theil Kuhn,
Rune E. Johnsen,
Patrick K. M. Tung,
Søren Schmidt,
Takenao Shinohara,
Ryoji Kiyanagi,
Anton S. Tremsin,
Nancy Elewa,
Robin Woracek,
Markus Strobl
Abstract:
Obtaining a complete picture of local processes still poses a significant challenge in battery research. Here we demonstrate an in-situ combination of multimodal neutron imaging with neutron diffraction for spatially resolved operando observations of the lithiation-delithiation of a graphite electrode in a Li-ion battery cell. Throughout the lithiation-delithiation process we image the Li distribu…
▽ More
Obtaining a complete picture of local processes still poses a significant challenge in battery research. Here we demonstrate an in-situ combination of multimodal neutron imaging with neutron diffraction for spatially resolved operando observations of the lithiation-delithiation of a graphite electrode in a Li-ion battery cell. Throughout the lithiation-delithiation process we image the Li distribution based on the local beam attenuation. Simultaneously, we observe the development of the lithiated graphite phases as a function of cycling time and electrode thickness and integral throughout its volume by diffraction contrast imaging and diffraction, respectively. While the conventional imaging data allows to observe the Li uptake in graphite already during the formation of the solid electrolyte interphase, diffraction indicates the onset and development of the Li insertion/extraction globally, which supports the local structural transformation observations by diffraction contrast imaging.
△ Less
Submitted 8 April, 2021;
originally announced April 2021.
-
New high-sensitivity searches for neutrons converting into antineutrons and/or sterile neutrons at the European Spallation Source
Authors:
A. Addazi,
K. Anderson,
S. Ansell,
K. Babu,
J. Barrow,
D. V. Baxter,
P. M. Bentley,
Z. Berezhiani,
R. Bevilacqua,
C. Bohm,
G. Brooijmans,
J. Broussard,
R. Biondi,
B. Dev,
C. Crawford,
A. Dolgov,
K. Dunne,
P. Fierlinger,
M. R. Fitzsimmons,
A. Fomin,
M. Frost,
S. Gardner,
A. Galindo-Uribarri,
E. Golubeva,
S. Girmohanta
, et al. (70 additional authors not shown)
Abstract:
The violation of Baryon Number, $\mathcal{B}$, is an essential ingredient for the preferential creation of matter over antimatter needed to account for the observed baryon asymmetry in the universe. However, such a process has yet to be experimentally observed. The HIBEAM/NNBAR %experiment program is a proposed two-stage experiment at the European Spallation Source (ESS) to search for baryon numbe…
▽ More
The violation of Baryon Number, $\mathcal{B}$, is an essential ingredient for the preferential creation of matter over antimatter needed to account for the observed baryon asymmetry in the universe. However, such a process has yet to be experimentally observed. The HIBEAM/NNBAR %experiment program is a proposed two-stage experiment at the European Spallation Source (ESS) to search for baryon number violation. The program will include high-sensitivity searches for processes that violate baryon number by one or two units: free neutron-antineutron oscillation ($n\rightarrow \bar{n}$) via mixing, neutron-antineutron oscillation via regeneration from a sterile neutron state ($n\rightarrow [n',\bar{n}'] \rightarrow \bar{n}$), and neutron disappearance ($n\rightarrow n'$); the effective $Δ\mathcal{B}=0$ process of neutron regeneration ($n\rightarrow [n',\bar{n}'] \rightarrow n$) is also possible. The program can be used to discover and characterise mixing in the neutron, antineutron, and sterile neutron sectors. The experiment addresses topical open questions such as the origins of baryogenesis, the nature of dark matter, and is sensitive to scales of new physics substantially in excess of those available at colliders. A goal of the program is to open a discovery window to neutron conversion probabilities (sensitivities) by up to three orders of magnitude compared with previous searches. The opportunity to make such a leap in sensitivity tests should not be squandered. The experiment pulls together a diverse international team of physicists from the particle (collider and low energy) and nuclear physics communities, while also including specialists in neutronics and magnetics.
△ Less
Submitted 8 June, 2020;
originally announced June 2020.
-
Time- and energy-resolved effects in the boron-10 based Multi-Grid and helium-3 based thermal neutron detectors
Authors:
A. Backis,
A. Khaplanov,
R. Al Jebali,
R. Ammer,
I. Apostolidis,
J. Birch,
C. -C. Lai,
P. P. Deen,
M. Etxegarai,
N. de Ruette,
J. Freita Ramos,
D. F. Förster,
E. Haettner,
R. Hall-Wilton,
D. Hamilton,
C. Höglund,
P. M. Kadletz,
K. Kanaki,
E. Karnickis,
O. Kirstein,
S. Kolya,
Z. Kraujalyte,
A. Laloni,
K. Livingston,
O. Löhman
, et al. (11 additional authors not shown)
Abstract:
The boron-10 based Multi-Grid detector is being developed as an alternative to helium-3 based neutron detectors. At the European Spallation Source, the detector will be used for time-of-flight neutron spectroscopy at cold to thermal neutron energies. The objective of this work is to investigate fine time- and energy-resolved effects of the Multi-Grid detector, down to a few $μ$eV, while comparing…
▽ More
The boron-10 based Multi-Grid detector is being developed as an alternative to helium-3 based neutron detectors. At the European Spallation Source, the detector will be used for time-of-flight neutron spectroscopy at cold to thermal neutron energies. The objective of this work is to investigate fine time- and energy-resolved effects of the Multi-Grid detector, down to a few $μ$eV, while comparing it to the performance of a typical helium-3 tube. Furthermore, it is to characterize differences between the detector technologies in terms of internal scattering, as well as the time reconstruction of ~ $μ$s short neutron pulses. The data were taken at the Helmholtz Zentrum Berlin, where the Multi-Grid detector and a helium-3 tube were installed at the ESS test beamline, V20. Using a Fermi-chopper, the neutron beam of the reactor was chopped into a few tens of $μ$s wide pulses before reaching the detector, located a few tens of cm downstream. The data of the measurements show an agreement between the derived and calculated neutron detection efficiency curve. The data also provide fine details on the effect of internal scattering, and how it can be reduced. For the first time, the chopper resolution was comparable to the timing resolution of the Multi-Grid detector. This allowed a detailed study of time- and energy resolved effects, as well as a comparison with a typical helium-3 tube.
△ Less
Submitted 14 January, 2021; v1 submitted 2 June, 2020;
originally announced June 2020.
-
Vanadium-based neutron Beam Monitor
Authors:
Vendula Maulerova,
Kalliopi Kanaki,
Peter Kadletz,
Robin Woracek,
Thomas Wilpert,
Kevin Fissum,
Alessio Laloni,
Nicholai Mauritzson,
Fatima Issa,
Richard Hall-Wilton
Abstract:
A prototype quasi-parasitic thermal neutron beam monitor based on isotropic neutron scattering from a thin natural vanadium foil and standard $^3$He proportional counters is conceptualized, designed, simulated, calibrated, and commissioned. The European Spallation Source designed to deliver the highest integrated neutron flux originating from a pulsed source is currently under construction in Lund…
▽ More
A prototype quasi-parasitic thermal neutron beam monitor based on isotropic neutron scattering from a thin natural vanadium foil and standard $^3$He proportional counters is conceptualized, designed, simulated, calibrated, and commissioned. The European Spallation Source designed to deliver the highest integrated neutron flux originating from a pulsed source is currently under construction in Lund, Sweden. The effort to investigate a vanadium-based neutron beam monitor was triggered by a list of requirements for Beam Monitors permanently placed in the ESS neutron beams in order to provide reliable monitoring at complex beamlines: low attenuation, linear response over a wide range of neutron fluxes, near to constant efficiency for neutron wavelengths in a range of 0.6-10 Å, calibration stability and the possibility to place the system in vacuum are all desirable characteristics. The scattering-based prototype, employing a natural vanadium foil and standard $^3$He proportional counters, was investigated at the V17 and V20 neutron beamlines of the Helmholtz-Zentrum in Berlin, Germany, in several different geometrical configurations of the $^3$He proportional counters around the foil. Response linearity is successfully demonstrated for foil thicknesses ranging from 0.04 mm to 3.15 mm. Attenuation lower than 1% for thermal neutrons is demonstrated for the 0.04 mm and 0.125 mm foils. The geometries used for the experiment were simulated allowing for absolute flux calibration and establishing the possible range of efficiencies for various designs of the prototype. The operational flux limits for the beam monitor prototype were established as a dependency of the background radiation and prototype geometry. The herein demonstrated prototype monitors can be employed for neutron fluxes ranging from $10^3-10^{10}$ n/s/cm$^2$.
△ Less
Submitted 24 February, 2020;
originally announced February 2020.
-
First neutron data recorded at the V20 Test Instrument utilizing prototype chopper systems and beam monitors planned for ESS
Authors:
V. Maulerova,
J. Nilsson,
M. Olsson,
S. Alcock,
A. Quintanilla,
D. Zielinski,
A. Mukai,
F. Issa,
S. Kolya,
W. Smith,
D. Broderick,
K. Löki,
J. Sparger,
R. Woracek,
J. Taylor,
T. Richter,
R. Hall-Wilton,
O. Kirstein,
N. Tsapatsaris
Abstract:
The main objective of the European Spallation Source (ESS) is to perform consistently high impact neutron scattering science using the highest neutron flux of any facility in its class. This ambition is naturally accompanied by operational challenges related to the vertical integration of the instruments' hardware and software shared between collaborators from 13 European countries. The present wo…
▽ More
The main objective of the European Spallation Source (ESS) is to perform consistently high impact neutron scattering science using the highest neutron flux of any facility in its class. This ambition is naturally accompanied by operational challenges related to the vertical integration of the instruments' hardware and software shared between collaborators from 13 European countries. The present work will detail integration tests performed at the V20 Test Instrument at Helmholtz-Zentrum Berlin in Germany. An ESS chopper prototype was successfully controlled through the ESS Chopper Integration Controller (CHIC) and Experimental Physics and Industrial Control System (EPICS). Neutron data were successfully collected and time stamped using neutron beam monitors, the ESS prototype detector readout electronics and the Data Management Software Center (DMSC) data acquisition (DAQ) software. Chopper rotation events were timestamped and the chopper disk showed excellent stability and neutron absorption characteristics. The test resulted in (i) the expected time-of-flight (TOF) response, (ii) collected data that compared well with instrument reference data and (iii) revealing the diagnostic power of the performed integration tests. The results of this integration test validate the ESS chopper and detector DAQ architecture choices.
△ Less
Submitted 21 August, 2019;
originally announced August 2019.