Systematic effects on a Compton polarimeter at the focus of an X-ray mirror
Authors:
M. Aoyagi,
R. G. Bose,
S. Chun,
E. Gau,
K. Hu,
K. Ishiwata,
N. K. Iyer,
F. Kislat,
M. Kiss,
K. Klepper,
H. Krawczynski,
L. Lisalda,
Y. Maeda,
F. af Malmborg,
H. Matsumoto,
A. Miyamoto,
T. Miyazawa,
M. Pearce,
B. F. Rauch,
N. Rodriguez Cavero,
S. Spooner,
H. Takahashi,
Y. Uchida,
A. T. West,
K. Wimalasena
, et al. (1 additional authors not shown)
Abstract:
XL-Calibur is a balloon-borne Compton polarimeter for X-rays in the $\sim$15-80 keV range. Using an X-ray mirror with a 12 m focal length for collecting photons onto a beryllium scattering rod surrounded by CZT detectors, a minimum-detectable polarization as low as $\sim$3% is expected during a 24-hour on-target observation of a 1 Crab source at 45$^{\circ}$ elevation. Systematic effects alter the…
▽ More
XL-Calibur is a balloon-borne Compton polarimeter for X-rays in the $\sim$15-80 keV range. Using an X-ray mirror with a 12 m focal length for collecting photons onto a beryllium scattering rod surrounded by CZT detectors, a minimum-detectable polarization as low as $\sim$3% is expected during a 24-hour on-target observation of a 1 Crab source at 45$^{\circ}$ elevation. Systematic effects alter the reconstructed polarization as the mirror focal spot moves across the beryllium scatterer, due to pointing offsets, mechanical misalignment or deformation of the carbon-fiber truss supporting the mirror and the polarimeter. Unaddressed, this can give rise to a spurious polarization signal for an unpolarized flux, or a change in reconstructed polarization fraction and angle for a polarized flux. Using bench-marked Monte-Carlo simulations and an accurate mirror point-spread function characterized at synchrotron beam-lines, systematic effects are quantified, and mitigation strategies discussed. By recalculating the scattering site for a shifted beam, systematic errors can be reduced from several tens of percent to the few-percent level for any shift within the scattering element. The treatment of these systematic effects will be important for any polarimetric instrument where a focused X-ray beam is im**ing on a scattering element surrounded by counting detectors.
△ Less
Submitted 23 February, 2024;
originally announced February 2024.
The design and performance of the XL-Calibur anticoincidence shield
Authors:
N. K. Iyer,
M. Kiss,
M. Pearce,
T. -A. Stana,
H. Awaki,
R. G. Bose,
A. Dasgupta,
G. De Geronimo,
E. Gau,
T. Hakamata,
M. Ishida,
K. Ishiwata,
W. Kamogawa,
F. Kislat,
T. Kitaguchi,
H. Krawczynski,
L. Lisalda,
Y. Maeda,
H. Matsumoto,
A. Miyamoto,
T. Miyazawa,
T. Mizuno,
B. F. Rauch,
N. Rodriguez Cavero,
N. Sakamoto
, et al. (9 additional authors not shown)
Abstract:
The XL-Calibur balloon-borne hard X-ray polarimetry mission comprises a Compton-scattering polarimeter placed at the focal point of an X-ray mirror. The polarimeter is housed within a BGO anticoincidence shield, which is needed to mitigate the considerable background radiation present at the observation altitude of ~40 km. This paper details the design, construction and testing of the anticoincide…
▽ More
The XL-Calibur balloon-borne hard X-ray polarimetry mission comprises a Compton-scattering polarimeter placed at the focal point of an X-ray mirror. The polarimeter is housed within a BGO anticoincidence shield, which is needed to mitigate the considerable background radiation present at the observation altitude of ~40 km. This paper details the design, construction and testing of the anticoincidence shield, as well as the performance measured during the week-long maiden flight from Esrange Space Centre to the Canadian Northwest Territories in July 2022. The in-flight performance of the shield followed design expectations, with a veto threshold <100 keV and a measured background rate of ~0.5 Hz (20-40 keV). This is compatible with the scientific goals of the mission, where %-level minimum detectable polarisation is sought for a Hz-level source rate.
△ Less
Submitted 8 December, 2022;
originally announced December 2022.