Abstract:
The reliable detection of neutrons in a harsh gamma-ray environment is an important aspect of establishing non-destructive methods for the characterization of spent nuclear fuel. In this study, we present results from extended in-situ monitoring of detector systems consisting of commercially available components: EJ-426, a $^6$Li-enriched solid-state scintillator material sensitive to thermal neut…
▽ More
The reliable detection of neutrons in a harsh gamma-ray environment is an important aspect of establishing non-destructive methods for the characterization of spent nuclear fuel. In this study, we present results from extended in-situ monitoring of detector systems consisting of commercially available components: EJ-426, a $^6$Li-enriched solid-state scintillator material sensitive to thermal neutrons, and two different types of Hamamatsu photomultiplier tubes (PMT). Over the period of eight months, these detectors were operated in close vicinity to spent nuclear fuel stored at the interim storage facility CLAB, Oskarshamn, Sweden. At the measurement position the detectors were continuously exposed to an estimated neutron flux of approx. 280 n/s $\cdot$ cm$^2$ and a gamma-ray dose rate of approx. 6 Sv/h.
Using offline software algorithms, neutron pulses were identified in the data. Over the entire investigated dose range of up to 35 kGr, the detector systems were functioning and were delivering detectable neutron signals. Their performance as measured by the number of identified neutrons degrades down to about 30% of the initial value. Investigations of the irradiated components suggest that this degradation is a result of reduced optical transparency of the involved materials as well as a reduction of PMT gain due to the continuous high currents. Increasing the gain of the PMT through step-ups of the applied high voltage allowed to partially compensate for this loss in detection sensitivity.
The integrated neutron fluence during the measurement was experimentally verified to be in the order of $5 \cdot 10^9$ n/cm$^2$. The results were interpreted with the help of MCNP6.2 simulations of the setup and the neutron flux.
△ Less
Submitted 30 September, 2021; v1 submitted 2 July, 2021;
originally announced July 2021.
Response of a Li-glass/multi-anode photomultiplier detector to focused proton and deuteron beams
Authors:
E. Rofors,
J. Pallon,
R. Al Jebali,
J. R. M. Annand,
L. Boyd,
M. J. Christensen,
U. Clemens,
S. Desert,
M. Elfman,
R. Engels,
K. G. Fissum,
H. Frielinghaus,
R. Frost,
S. Gardner,
C. Gheorghe,
R. Hall-Wilton,
S. Jaksch,
K. Kanaki,
G. Kemmerling,
P. Kristiansson,
K. Livingston,
V. Maulerova,
N. Mauritzson,
R. Montgomery,
H. Perrey
, et al. (4 additional authors not shown)
Abstract:
The response of a position-sensitive Li-glass based scintillation detector to focused beams of 2.5 MeV protons and deuterons has been investigated. The beams were scanned across the detector in 0.5 mm horizontal and vertical steps perpendicular to the beams. Scintillation light was registered using an 8 by 8 pixel multi-anode photomultiplier tube. The signal amplitudes were recorded for each pixel…
▽ More
The response of a position-sensitive Li-glass based scintillation detector to focused beams of 2.5 MeV protons and deuterons has been investigated. The beams were scanned across the detector in 0.5 mm horizontal and vertical steps perpendicular to the beams. Scintillation light was registered using an 8 by 8 pixel multi-anode photomultiplier tube. The signal amplitudes were recorded for each pixel on an event-by-event basis. Several pixels generally registered considerable signals at each beam location. The number of pixels above set thresholds were investigated, with the optimization of the single-hit efficiency over the largest possible area as the goal. For both beams, at a threshold of ~50% of the mean of the full-deposition peak, ~80% of the events were registered in a single pixel, resulting in an effective position resolution of ~5 mm in X and Y.
△ Less
Submitted 18 May, 2020;
originally announced May 2020.