Evaluation of a method for time-of-flight, wavelength and distance calibration for neutron scattering instruments by means of a mini-chopper and standard neutron monitors
Authors:
L. Vergara,
M. Arai,
R. Woracek,
M. Olsson,
A. Quintanilla,
S. Alcock,
J. Nilsson,
K. Kanaki,
P. M. Kadletz,
O. Kirstein,
R. Hall-Wilton,
N. Tsapatsaris
Abstract:
Accurate conversion of neutron time-of-flight (TOF) to wavelength, and its uncertainty, is of fundamental importance to neutron scattering measurements. Especially in cases where instruments are highly configurable, the determination of the absolute wavelength after any change must always be performed. Inspired by the manner with which neutron spectrometers determine the absolute wavelength, we ev…
▽ More
Accurate conversion of neutron time-of-flight (TOF) to wavelength, and its uncertainty, is of fundamental importance to neutron scattering measurements. Especially in cases where instruments are highly configurable, the determination of the absolute wavelength after any change must always be performed. Inspired by the manner with which neutron spectrometers determine the absolute wavelength, we evaluate for the first time, in the author's knowledge, a commonly used method for converting TOF to neutron wavelength, the distance of a monitor from the source of neutrons and we analytically calculate the uncertainty contributions that limit the precision of the conversion. The method was evaluated at the V20 test beamline at the Helmholtz Zentrum Berlin (HZB), emulating the ESS source with a pulse of 2.86 ms length and 14 Hz repetition rate, by using a mini-chopper operated at 140 Hz, beam monitors (BMs) and data acquisition infrastructure. The mini-chopper created well-defined neutron pulses and the BM was placed at two positions, enabling the average wavelength of each of the pulses created to be determined. The used experimental setup resulted in absolute wavelength determination at the monitor positions with a $δλ_{mean} / λ_{mean}$ of $\sim$1.8% for $λ>4$ Å. With a modest increase of the distance between the reference monitor positions a $δλ_{mean} / λ_{mean}$ of below 0.5% can be achieved. Further improvements are possible by using a thinner monitor, smaller chopper disc openings and a higher rotational speed chopper. The method requires only two neutron measurements and doesn't necessitate the use of crystals or complex fitting, and could constitute a suitable addition to imaging, diffraction, reflectometers and small angle neutron scattering instruments, at spallation sources, that do not normally utilise fast choppers.
△ Less
Submitted 3 November, 2022;
originally announced November 2022.
-
Time- and energy-resolved effects in the boron-10 based Multi-Grid and helium-3 based thermal neutron detectors
Authors:
A. Backis,
A. Khaplanov,
R. Al Jebali,
R. Ammer,
I. Apostolidis,
J. Birch,
C. -C. Lai,
P. P. Deen,
M. Etxegarai,
N. de Ruette,
J. Freita Ramos,
D. F. Förster,
E. Haettner,
R. Hall-Wilton,
D. Hamilton,
C. Höglund,
P. M. Kadletz,
K. Kanaki,
E. Karnickis,
O. Kirstein,
S. Kolya,
Z. Kraujalyte,
A. Laloni,
K. Livingston,
O. Löhman
, et al. (11 additional authors not shown)
Abstract:
The boron-10 based Multi-Grid detector is being developed as an alternative to helium-3 based neutron detectors. At the European Spallation Source, the detector will be used for time-of-flight neutron spectroscopy at cold to thermal neutron energies. The objective of this work is to investigate fine time- and energy-resolved effects of the Multi-Grid detector, down to a few $μ$eV, while comparing…
▽ More
The boron-10 based Multi-Grid detector is being developed as an alternative to helium-3 based neutron detectors. At the European Spallation Source, the detector will be used for time-of-flight neutron spectroscopy at cold to thermal neutron energies. The objective of this work is to investigate fine time- and energy-resolved effects of the Multi-Grid detector, down to a few $μ$eV, while comparing it to the performance of a typical helium-3 tube. Furthermore, it is to characterize differences between the detector technologies in terms of internal scattering, as well as the time reconstruction of ~ $μ$s short neutron pulses. The data were taken at the Helmholtz Zentrum Berlin, where the Multi-Grid detector and a helium-3 tube were installed at the ESS test beamline, V20. Using a Fermi-chopper, the neutron beam of the reactor was chopped into a few tens of $μ$s wide pulses before reaching the detector, located a few tens of cm downstream. The data of the measurements show an agreement between the derived and calculated neutron detection efficiency curve. The data also provide fine details on the effect of internal scattering, and how it can be reduced. For the first time, the chopper resolution was comparable to the timing resolution of the Multi-Grid detector. This allowed a detailed study of time- and energy resolved effects, as well as a comparison with a typical helium-3 tube.
△ Less
Submitted 14 January, 2021; v1 submitted 2 June, 2020;
originally announced June 2020.
-
First neutron data recorded at the V20 Test Instrument utilizing prototype chopper systems and beam monitors planned for ESS
Authors:
V. Maulerova,
J. Nilsson,
M. Olsson,
S. Alcock,
A. Quintanilla,
D. Zielinski,
A. Mukai,
F. Issa,
S. Kolya,
W. Smith,
D. Broderick,
K. Löki,
J. Sparger,
R. Woracek,
J. Taylor,
T. Richter,
R. Hall-Wilton,
O. Kirstein,
N. Tsapatsaris
Abstract:
The main objective of the European Spallation Source (ESS) is to perform consistently high impact neutron scattering science using the highest neutron flux of any facility in its class. This ambition is naturally accompanied by operational challenges related to the vertical integration of the instruments' hardware and software shared between collaborators from 13 European countries. The present wo…
▽ More
The main objective of the European Spallation Source (ESS) is to perform consistently high impact neutron scattering science using the highest neutron flux of any facility in its class. This ambition is naturally accompanied by operational challenges related to the vertical integration of the instruments' hardware and software shared between collaborators from 13 European countries. The present work will detail integration tests performed at the V20 Test Instrument at Helmholtz-Zentrum Berlin in Germany. An ESS chopper prototype was successfully controlled through the ESS Chopper Integration Controller (CHIC) and Experimental Physics and Industrial Control System (EPICS). Neutron data were successfully collected and time stamped using neutron beam monitors, the ESS prototype detector readout electronics and the Data Management Software Center (DMSC) data acquisition (DAQ) software. Chopper rotation events were timestamped and the chopper disk showed excellent stability and neutron absorption characteristics. The test resulted in (i) the expected time-of-flight (TOF) response, (ii) collected data that compared well with instrument reference data and (iii) revealing the diagnostic power of the performed integration tests. The results of this integration test validate the ESS chopper and detector DAQ architecture choices.
△ Less
Submitted 21 August, 2019;
originally announced August 2019.
-
Tomographic Reconstruction of Triaxial Strain Fields from Bragg-Edge Neutron Imaging
Authors:
J. N. Hendriks,
A. W. T. Gregg,
R. R. Jackson,
C. M. Wensrich,
A. Wills,
A. S. Tremsin,
T. Shinohara,
V. Luzin,
O. Kirstein
Abstract:
This paper presents a proof-of-concept demonstration of triaxial strain tomography from Bragg-edge neutron imaging within a three-dimensional sample. Bragg-edge neutron transmission can provide high-resolution images of the average through thickness strain within a polycrystalline material. This poses an associated rich tomography problem which seeks to reconstruct the full triaxial strain field f…
▽ More
This paper presents a proof-of-concept demonstration of triaxial strain tomography from Bragg-edge neutron imaging within a three-dimensional sample. Bragg-edge neutron transmission can provide high-resolution images of the average through thickness strain within a polycrystalline material. This poses an associated rich tomography problem which seeks to reconstruct the full triaxial strain field from these images. The presented demonstration is an important step towards solving this problem, and towards a technique capable of studying the residual strain and stress within engineering components. A Gaussian process based approach is used that ensures the reconstruction satisfies equilibrium and known boundary conditions. This approach is demonstrated experimentally on a non-trivial steel sample with use of the RADEN instrument at the Japan Proton Accelerator Research Complex. Validation of the reconstruction is provided by comparison with conventional strain scans from the KOWARI constant-wavelength strain diffractometer at the Australian Nuclear Science and Technology Organisation and simulations via finite element analysis.
△ Less
Submitted 28 November, 2019; v1 submitted 20 June, 2019;
originally announced June 2019.
-
Tomographic Reconstruction of Two-Dimensional Residual Strain Fields from Bragg-Edge Neutron Imaging
Authors:
Alexander Gregg,
Johannes Hendriks,
Christopher Wensrich,
Adrian Wills,
Anton Tremsin,
Vladimir Luzin,
Takenao Shinohara,
Oliver Kirstein,
Michael Meylan,
Erich Kisi
Abstract:
Bragg-edge strain imaging from energy-resolved neutron transmission measurements poses an interesting tomography problem. The solution to this problem will allow the reconstruction of detailed triaxial stress and strain distributions within polycrystalline solids from sets of Bragg-edge strain images. Work over the last decade has provided some solutions for a limited number of special cases. In t…
▽ More
Bragg-edge strain imaging from energy-resolved neutron transmission measurements poses an interesting tomography problem. The solution to this problem will allow the reconstruction of detailed triaxial stress and strain distributions within polycrystalline solids from sets of Bragg-edge strain images. Work over the last decade has provided some solutions for a limited number of special cases. In this paper, we provide a general approach to reconstruction of an arbitrary system based on a least squares process constrained by equilibrium. This approach is developed in two- dimensions before being demonstrated experimentally on two samples using the RADEN instrument at the J-PARC spallation neutron source in Japan. Validation of the resulting reconstructions is provided through a comparison to conventional constant wavelength strain measurements carried out on the KOWARI engineering diffractometer within ANSTO in Australia. The paper concludes with a discussion on the range of problems to be addressed in a three-dimensional implementation.
△ Less
Submitted 19 August, 2018;
originally announced August 2018.
-
Multi-Grid Detector for Neutron Spectroscopy: Results Obtained on Time-of-Flight Spectrometer CNCS
Authors:
M. Anastasopoulos,
R. Bebb,
K. Berry,
J. Birch,
T. Bryś,
J. -C. Buffet,
J. -F. Clergeau,
P. P. Deen,
G. Ehlers,
P. van Esch,
S. M. Everett,
B. Guerard,
R. Hall-Wilton,
K. Herwig,
L. Hultman,
C. Höglund,
I. Iruretagoiena,
F. Issa,
J. Jensen,
A. Khaplanov,
O. Kirstein,
I. Lopez-Higuera,
F. Piscitelli,
L. Robinson,
S. Schmidt
, et al. (1 additional authors not shown)
Abstract:
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demo…
▽ More
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demonstrator has an active area of 0.2 m$^2$. It is specifically tailored to the specifications of CNCS. The detector was installed in June 2016 and has operated since then, collecting neutron scattering data in parallel to the He-3 detectors of CNCS. In this paper, we present a comprehensive analysis of this data, in particular on instrument energy resolution, rate capability, background and relative efficiency. Stability, gamma-ray and fast neutron sensitivity have also been investigated. The effect of scattering in the detector components has been measured and provides input to comparison for Monte Carlo simulations. All data is presented in comparison to that measured by the $^3$He detectors simultaneously, showing that all features recorded by one detector are also recorded by the other. The energy resolution matches closely. We find that the Multi-Grid is able to match the data collected by $^3$He, and see an indication of a considerable advantage in the count rate capability. Based on these results, we are confident that the Multi-Grid detector will be capable of producing high quality scientific data on chopper spectrometers utilising the unprecedented neutron flux of the ESS.
△ Less
Submitted 3 April, 2017; v1 submitted 10 March, 2017;
originally announced March 2017.
-
Neutron detectors for the ESS diffractometers
Authors:
I. Stefanescu,
M. Christensen,
J. Fenske,
R. Hall-Wilton,
P. F. Henry,
O. Kirstein,
M. Mueller,
G. Nowak,
D. Pooley,
D. Raspino,
N. Rhodes,
J. Saroun,
J. Schefer,
E. Schooneveld,
J. Sykora,
W. Schweika
Abstract:
The ambitious instrument suite for the future European Spallation Source whose civil construction started recently in Lund, Sweden, demands a set of diverse and challenging requirements for the neutron detectors. For instance, the unprecedented high flux expected on the samples to be investigated in neutron diffraction or reflectometry experiments requires detectors that can handle high counting r…
▽ More
The ambitious instrument suite for the future European Spallation Source whose civil construction started recently in Lund, Sweden, demands a set of diverse and challenging requirements for the neutron detectors. For instance, the unprecedented high flux expected on the samples to be investigated in neutron diffraction or reflectometry experiments requires detectors that can handle high counting rates, while the investigation of sub-millimeter protein crystals will only be possible with large-area detectors that can achieve a position resolution as low as 200 μm. This has motivated an extensive research and development campaign to advance the state-of-the-art detector and to find new technologies that can reach maturity by the time the ESS will operate at full potential. This paper presents the key detector requirements for three of the Time-of-Flight diffraction instrument concepts selected by the Scientific Advisory Committee to advance into the phase of preliminary engineering design. We discuss the available detector technologies suitable for this particular instrument class and their major challenges. The detector technologies selected by the instrument teams to collect the diffraction patterns are briefly discussed. Analytical calculations, Monte-Carlo simulations, and real experimental data are used to develop a generic method to esti- mate the event rate in the diffraction detectors. The proposed approach is based upon conservative assumptions that use information and input parameters that reflect our current level of knowledge and understanding of the ESS project. We apply this method to make predictions for the future diffraction instruments, and thus provide additional information that can help the instrument teams with the optimisation of the detector designs.
△ Less
Submitted 29 November, 2016; v1 submitted 8 July, 2016;
originally announced July 2016.
-
Cost Optimisation of an Instrument Suite at an Accelerator-Driven Spallation Source
Authors:
P. M. Bentley,
C. S. Zendler,
O. Kirstein
Abstract:
This artcile presents an optimisation of performance and cost of neutron scattering instrumentation at the European Spallation Source. This is done by trading detailed cost functions against beam transmission functions in a multi-dimensional, yet simple, parameter space. On the one hand, the neutron guide cost increases as a power of the desired beam divergence, and inversely with the minimum wave…
▽ More
This artcile presents an optimisation of performance and cost of neutron scattering instrumentation at the European Spallation Source. This is done by trading detailed cost functions against beam transmission functions in a multi-dimensional, yet simple, parameter space. On the one hand, the neutron guide cost increases as a power of the desired beam divergence, and inversely with the minimum wavelength, due to the supermirror coating needed. On the other hand, the more neutrons are transported to the instrument the greater are the shielding costs to deal with the gamma rays that result from the eventual absorption of the neutrons. There are additional factors in that many of the parameters defining the neutron guide geometry are continuous variables, and the straightness of the guide increases the transmission of high energy spallation products, which affect the specifications of particularly heavy hardware, such as heavy shutters and additional shielding, beam stops etc. Over the suite of 16 instruments, a maximum potential cost saving is identified in the vicinity of 20 M\euro \,, with negligible penalties on instrument performance and bandwidth, by trading minimum neutron wavelength $λ$ against total optical costs and shielding costs. Finally, these cost calculations are shown to be consistent with those at real world, existing spallation source facilities.
△ Less
Submitted 18 March, 2016; v1 submitted 1 March, 2016;
originally announced March 2016.
-
Hardware Aspects, Modularity and Integration of an Event Mode Data Acquisition and Instrument Control for the European Spallation Source (ESS)
Authors:
Thomas Gahl,
Mark Hagen,
Richard Hall-Wilton,
Scott Kolya,
Mark Koennecke,
Miha Rescic,
Thomas H. Rod,
Iain Sutton,
Garry Trahern,
Oliver Kirstein
Abstract:
The European Spallation Source (ESS) in Lund, Sweden is just entering the construction phase with 3 neutron instruments having started in its design concept phase in 2014. As a collaboration of 17 European countries the majority of hardware devices for neutron instrumentation will be provided in-kind. This presents numerous technical and organisational challenges for the construction and the integ…
▽ More
The European Spallation Source (ESS) in Lund, Sweden is just entering the construction phase with 3 neutron instruments having started in its design concept phase in 2014. As a collaboration of 17 European countries the majority of hardware devices for neutron instrumentation will be provided in-kind. This presents numerous technical and organisational challenges for the construction and the integration of the instruments into the facility wide infrastructure; notably the EPICS control network with standardised hardware interfaces and the facilities absolute timing system. Additionally the new generation of pulsed source requires a new complexity and flexibility of instrumentation to fully exploit its opportunities. In this contribution we present a strategy for the modularity of the instrument hardware with well-defined standardized functionality and control & data interfaces integrating into EPICS and the facilities timing system. It allows for in-kind contribution of dedicated modules for each instrument (horizontal approach) as well as of whole instruments (vertical approach). Key point of the strategy is the time stam** of all readings from the instruments control electronics extending the event mode data acquisition from neutron events to all metadata. This gives the control software the flexibility necessary to adapt the functionality of the instruments to the demands of each single experiment. We present the advantages of that approach for operation and diagnostics and discuss additional hardware requirements necessary.
△ Less
Submitted 7 July, 2015;
originally announced July 2015.
-
Overcoming High Energy Backgrounds at Pulsed Spallation Sources
Authors:
Nataliia Cherkashyna,
Richard J. Hall-Wilton,
Douglas D. DiJulio,
Anton Khaplanov,
Dorothea Pfeiffer,
Julius Scherzinger,
Carsten P. Cooper-Jensen,
Kevin G. Fissum,
Stuart Ansell,
Erik B. Iverson,
Georg Ehlers,
Franz X. Gallmeier,
Tobias Panzner,
Emmanouela Rantsiou,
Kalliopi Kanaki,
Uwe Filges,
Thomas Kittelmann,
Maddi Extegarai,
Valentina Santoro,
Oliver Kirstein,
Phillip M. Bentley
Abstract:
Instrument backgrounds at neutron scattering facilities directly affect the quality and the efficiency of the scientific measurements that users perform. Part of the background at pulsed spallation neutron sources is caused by, and time-correlated with, the emission of high energy particles when the proton beam strikes the spallation target. This prompt pulse ultimately produces a signal, which ca…
▽ More
Instrument backgrounds at neutron scattering facilities directly affect the quality and the efficiency of the scientific measurements that users perform. Part of the background at pulsed spallation neutron sources is caused by, and time-correlated with, the emission of high energy particles when the proton beam strikes the spallation target. This prompt pulse ultimately produces a signal, which can be highly problematic for a subset of instruments and measurements due to the time-correlated properties, and different to that from reactor sources. Measurements of this background have been made at both SNS (ORNL, Oak Ridge, TN, USA) and SINQ (PSI, Villigen, Switzerland). The background levels were generally found to be low compared to natural background. However, very low intensities of high-energy particles have been found to be detrimental to instrument performance in some conditions. Given that instrument performance is typically characterised by S/N, improvements in backgrounds can both improve instrument performance whilst at the same time delivering significant cost savings. A systematic holistic approach is suggested in this contribution to increase the effectiveness of this. Instrument performance should subsequently benefit.
△ Less
Submitted 10 January, 2015;
originally announced January 2015.
-
Neutron Position Sensitive Detectors for the ESS
Authors:
Oliver Kirstein,
Richard Hall-Wilton,
Irina Stefanescu,
Maddi Etxegarai,
Michail Anastasopoulos,
Kevin Fissum,
Anna Gulyachkina,
Carina Höglund,
Mewlude Imam,
Kalliopi Kanaki,
Anton Khaplanov,
Thomas Kittelmann,
Scott Kolya,
Björn Nilsson,
Luis Ortega,
Dorothea Pfeiffer,
Francesco Piscitelli,
Judith Freita Ramos,
Linda Robinson,
Julius Scherzinger
Abstract:
The European Spallation Source (ESS) in Lund, Sweden will become the world's leading neutron source for the study of materials. The instruments are being selected from conceptual proposals submitted by groups from around Europe. These instruments present numerous challenges for detector technology in the absence of the availability of Helium-3, which is the default choice for detectors for instrum…
▽ More
The European Spallation Source (ESS) in Lund, Sweden will become the world's leading neutron source for the study of materials. The instruments are being selected from conceptual proposals submitted by groups from around Europe. These instruments present numerous challenges for detector technology in the absence of the availability of Helium-3, which is the default choice for detectors for instruments built until today and due to the extreme rates expected across the ESS instrument suite. Additionally a new generation of source requires a new generation of detector technologies to fully exploit the opportunities that this source provides. The detectors will be sourced from partners across Europe through numerous in-kind arrangements; a process that is somewhat novel for the neutron scattering community. This contribution presents briefly the current status of detectors for the ESS, and outlines the timeline to completion. For a conjectured instrument suite based upon instruments recommended for construction, a recently updated snapshot of the current expected detector requirements is presented. A strategy outline as to how these requirements might be tackled by novel detector developments is shown. In terms of future developments for the neutron community, synergies should be sought with other disciples, as recognized by various recent initiatives in Europe, in the context of the fundamentally multi-disciplinary nature of detectors. This strategy has at its basis the in-kind and collaborative partnerships necessary to be able to produce optimally performant detectors that allow the ESS instruments to be world-leading. This foresees and encourages a high level of collaboration and interdependence at its core, and rather than each group being all-rounders in every technology, the further development of centres of excellence across Europe for particular technologies and niches.
△ Less
Submitted 23 November, 2014;
originally announced November 2014.
-
Multi-Grid Boron-10 detector for large area applications in neutron scattering science
Authors:
Ken Andersen,
Thierry Bigault,
Jens Birch,
Jean-Claude Buffet,
Jonathan Correa,
Patrick van Esch,
Bruno Guerard,
Richard Hall-Wilton,
Lars Hultman,
Carina Höglund,
Jens Jensen,
Anton Khaplanov,
Oliver Kirstein,
Francesco Piscitelli,
Christian Vettier
Abstract:
The present supply of 3He can no longer meet the detector demands of the upcoming ESS facility and continued detector upgrades at current neutron sources. Therefore viable alternative technologies are required to support the development of cutting-edge instrumentation for neutron scattering science. In this context, 10B-based detectors are being developed by collaboration between the ESS, ILL, and…
▽ More
The present supply of 3He can no longer meet the detector demands of the upcoming ESS facility and continued detector upgrades at current neutron sources. Therefore viable alternative technologies are required to support the development of cutting-edge instrumentation for neutron scattering science. In this context, 10B-based detectors are being developed by collaboration between the ESS, ILL, and Linkö** University. This paper reports on progress of this technology and the prospects applying it in modern neutron scattering experiments. The detector is made-up of multiple rectangular gas counter tubes coated with B4C, enriched in 10B. An anode wire reads out each tube, thereby giving position of conversion in one of the lateral co-ordinates as well as in depth of the detector. Position resolution in the remaining co-ordinate is obtained by segmenting the cathode tube itself. Boron carbide films have been produced at Linkö** University and a detector built at ILL. The characterization study is presented in this paper, including measurement of efficiency, effects of the fill gas species and pressure, coating thickness variation on efficiency and sensitivity to gamma-rays.
△ Less
Submitted 4 September, 2012;
originally announced September 2012.