SQUID current sensors with an integrated thermally actuated input current limiter
Authors:
Rainer Körber,
Patryk Krzysteczko,
Monique Klemm,
Tianhao Liu,
Jan-Hendrik Storm,
Jörn Beyer
Abstract:
The development of SQUID current sensors with a thermally actuated input current limiter (TCL) integrated into the input circuit of the sensor is presented. The TCL is based on an unshunted Josephson Junction (JJ) series array, around which meanders a galvanically isolated, but tightly, thermally coupled, resistive heater element. By applying a current to the heater, the JJ critical currents can b…
▽ More
The development of SQUID current sensors with a thermally actuated input current limiter (TCL) integrated into the input circuit of the sensor is presented. The TCL is based on an unshunted Josephson Junction (JJ) series array, around which meanders a galvanically isolated, but tightly, thermally coupled, resistive heater element. By applying a current to the heater, the JJ critical currents can be reduced or completely suppressed, while the other on-chip SQUID circuit elements remain unaffected. The functional parameters of the TCL are determined by direct transport measurements and by static and dynamic flux coupling tests via the input circuit. In liquid Helium, a heater power, which reliably suppresses the critical current of the JJ array to zero, of $\sim$1.5 mW was needed, with the TCL normal state resistance being at 500 $Ω$. In this configuration, typical TCL switching times below 10 $μ$s were observed in the direct transport measurements.
The TCL can be used to temporarily or even permanently disable malfunctioning channels in multichannel SQUID magnetometer systems, when feedback into their otherwise superconducting input circuits is not possible. In doing so, significant signal distortions in neighbouring channels from screening currents in these input circuits are avoided. Furthermore, by completely suppressing and restoring the TCL critical current, dc offset currents in the superconducting input circuits can be prevented. This is relevant, for instance, in SQUID-based spin precession experiments on hyperpolarized noble gases, in which even dc currents of a few $μ$A in the input circuit can lead to significant magnetic field distortions in the adjacent sample region contributing to the transverse spin-spin relaxation rate $1/T_{2}$
△ Less
Submitted 27 February, 2023;
originally announced February 2023.
-
LIGHT-SABRE hyperpolarizes 1-^{13}C-pyruvate continuously, without magnetic field cycling
Authors:
Andrey N. Pravdivtsev,
Kai Buckenmaier,
Nicolas Kempf,
Gabrielle Stevanato,
Klaus Scheffler,
Joern Engelmann,
Markus Plaumann,
Rainer Koerber,
Jan-Bernd Hövener,
Thomas Theis
Abstract:
Nuclear spin hyperpolarization enables real-time observation of metabolism and intermolecular interactions in vivo. 1-13C-Pyruvate is the leading hyperpolarized tracer currently under evaluation in several clinical trials as a promising molecular imaging agent. Still, the quest for a simple, fast, and efficient hyperpolarization technique is ongoing. Here, we describe that continuous, weak irradia…
▽ More
Nuclear spin hyperpolarization enables real-time observation of metabolism and intermolecular interactions in vivo. 1-13C-Pyruvate is the leading hyperpolarized tracer currently under evaluation in several clinical trials as a promising molecular imaging agent. Still, the quest for a simple, fast, and efficient hyperpolarization technique is ongoing. Here, we describe that continuous, weak irradiation in the audio-frequency range of the 13C spin at 121 μT magnetic field (\sim twiceEarth\apos s field) enables spin order transfer from parahydrogen to 13C magnetization of 1-13C-pyruvate. These so-called LIGHT-SABRE pulses couple nuclear spin states of parahydrogen and pyruvate via the J-coupling network of reversibly exchanging Ir-complexes. Using \sim 100% parahydrogen at ambient pressure, we polarized 51 mM of 1-13C-pyruvate in the presence of 5.1 mM Ir-complex continuously and repeatedly to a polarization of 1.1% averaged over free and catalyst-bound pyruvate. The experiments were conducted at -8°C), where almost exclusively bound pyruvate was observed, corresponding to an estimated 11% polarization on bound pyruvate. The obtained hyperpolarization levels closely match those obtained via SABRE-SHEATH under otherwise identical conditions. The creation of three different types of spin orders was observed: transverse 13C magnetization along the applied magnetic field, 13C z-magnetization along the main field B_0, and 13C-1H zz-spin-order. With a superconducting quantum interference device (SQUID) for detection, we found that the generated spin orders result from tiny 1H-13C J-coupling interactions, which are not visible even with our narrow linewidth below 0.3 Hz.
△ Less
Submitted 18 February, 2023;
originally announced February 2023.
-
A tabletop Optically Pumped Magnetometer setup for the monitoring of magnetic nanoparticle clustering and immobilization using Thermal Noise Magnetometry
Authors:
Katrijn Everaert,
Tilmann Sander,
Rainer Körber,
Norbert Loewa,
Bartel Van Waeyenberge,
Jonathan Leliaert,
Frank Wiekhorst
Abstract:
Many characterization techniques for magnetic nanoparticles depend on the usage of external fields. This is not the case in Thermal Noise Magnetometry (TNM), where thermal fluctuations in the magnetic signal of magnetic nanoparticle ensembles are measured without any external excitation. This can provide valuable information about the fundamental dynamical properties of the particles, due to the p…
▽ More
Many characterization techniques for magnetic nanoparticles depend on the usage of external fields. This is not the case in Thermal Noise Magnetometry (TNM), where thermal fluctuations in the magnetic signal of magnetic nanoparticle ensembles are measured without any external excitation. This can provide valuable information about the fundamental dynamical properties of the particles, due to the purely observative experiments of this relatively new technique. Until now, TNM signals have been detected only by a superconducting quantum interference device (SQUID) sensor. We present a tabletop setup using Optically Pumped Magnetometers (OPMs) in a small magnetic shield, offering a flexible and accessible alternative and show the agreement between both measurement systems for two different commercially available nanoparticle samples. We argue that the OPM setup with high accessibility complements the SQUID setup with high sensitivity and bandwidth. Furthermore, because of its excellent sensitivity in the lower frequencies, the OPM tabletop setup is well suited to monitor aggregation processes where the magnetization dynamics of the particles tend to slow down, e.g. in biological processes. As a proof of concept, we show for three different immobilization and clustering processes the changes in the noise spectrum measured in the tabletop setup: 1) the aggregation of particles due to the addition of ethanol, 2) the formation of polymer structures in the sample due to UV exposure, and 3) the cellular uptake of the particles by THP-1 cells. From our results we conclude that the tabletop setup offers a flexible and widely adoptable sensor measurement unit to monitor the immobilization and clustering of magnetic nanoparticles over time for different applications.
△ Less
Submitted 30 September, 2022;
originally announced September 2022.
-
Towards ultra-sensitive SQUIDs based on sub-micrometer-sized Josephson Junctions
Authors:
Jan-Hendrik Storm,
Oliver Kieler,
Rainer Körber
Abstract:
We recently demonstrated a 1$^{\textrm{st}}$-order axial gradiometer SQUID system, which is operated in a liquid He dewar with negligible noise contribution. The achieved close to SQUID-limited measured coupled energy sensitivity $\varepsilon_{c}$ of $\sim 30\,h$ corresponds to a white field noise below 180~aT~Hz${^{-1/2}}$. In order to further improve the SQUID noise performance, the junction cap…
▽ More
We recently demonstrated a 1$^{\textrm{st}}$-order axial gradiometer SQUID system, which is operated in a liquid He dewar with negligible noise contribution. The achieved close to SQUID-limited measured coupled energy sensitivity $\varepsilon_{c}$ of $\sim 30\,h$ corresponds to a white field noise below 180~aT~Hz${^{-1/2}}$. In order to further improve the SQUID noise performance, the junction capacitance was reduced by decreasing its lateral size from $2.5~μ$m to below $1~μ$m. This was realized by extending the fabrication process for submicrometer-sized Josephson Junctions based on the HfTi self-shunted junction technology to an SIS process with AlO$_{\textrm{x}}$ as the insulating layer. We achieved energy sensitivities of 4.7$\,h$ and 20$\,h$ at 4.2~K for uncoupled and coupled SQUIDs, respectively. We also investigated the temperature dependence of the noise of the uncoupled SQUIDs and reached an energy sensitivity of 0.65$\,h$ in the white noise regime at 400~mK.
△ Less
Submitted 5 November, 2023; v1 submitted 21 April, 2020;
originally announced April 2020.
-
Evaluating the Performance of Ultra-Low-Field MRI for In-vivo 3D Current Density Imaging of the Human Head
Authors:
Peter Hömmen,
Antti J. Mäkinen,
Alexander Hunold,
René Machts,
Jens Haueisen,
Koos C. J. Zevenhoven,
Risto J. Ilmoniemi,
Rainer Körber
Abstract:
Magnetic fields associated with currents flowing in tissue can be measured non-invasively by means of zero-field-encoded ultra-low-field magnetic resonance imaging (ULF MRI) enabling current density imaging (CDI) and possibly conductivity map** of human head tissues. Since currents applied to a human are limited by safety regulations and only a small fraction of the current passes through the re…
▽ More
Magnetic fields associated with currents flowing in tissue can be measured non-invasively by means of zero-field-encoded ultra-low-field magnetic resonance imaging (ULF MRI) enabling current density imaging (CDI) and possibly conductivity map** of human head tissues. Since currents applied to a human are limited by safety regulations and only a small fraction of the current passes through the relatively high-resistive skull, a sufficient signal-to-noise ratio (SNR) may be difficult to obtain when using this method. In this work, we study the relationship between the image SNR and the SNR of the field reconstructions from zero-field-encoded data. We evaluate these results for two existing ULF MRI scanners, one ultra-sensitive single-channel system and one whole-head multi-channel system, by simulating sequences necessary for current-density reconstruction. We also derive realistic current-density and magnetic-field estimates from finite-element-method simulations based on a three-compartment head model. We found that existing ULF-MRI systems reach sufficient SNR to detect intra-cranial current distributions with statistical uncertainty below 10%. However, they also reveal that image artifacts influence the reconstruction quality. Further, our simulations indicate that current-density reconstruction in the scalp requires a resolution less than 5 mm and demonstrate that the necessary sensitivity coverage can be accomplished by multi-channel devices.
△ Less
Submitted 20 January, 2020;
originally announced January 2020.
-
Towards direct neuronal current imaging via ultra-low-field MR
Authors:
Nora Höfner,
Rainer Körber,
Martin Burghoff
Abstract:
The feasibility of using ultra-low-field magnetic resonance (ULF MR) for direct neuronal current imaging (NCI) is investigated by phantom measurements. The aim of NCI is to improve the current localization accuracy for neuronal activity of established methods like electroencephalography (EEG) or magnetoencephalography (MEG) (~1 cm). A measurement setup was developed addressing the main challenge o…
▽ More
The feasibility of using ultra-low-field magnetic resonance (ULF MR) for direct neuronal current imaging (NCI) is investigated by phantom measurements. The aim of NCI is to improve the current localization accuracy for neuronal activity of established methods like electroencephalography (EEG) or magnetoencephalography (MEG) (~1 cm). A measurement setup was developed addressing the main challenge of reaching the necessary sensitivity in order to possibly resolve the faint influence of neuronal magnetic fields on 1H nuclear spin precession. Phantom measurements close to physiology conditions are performed simulating a specific long-lasting neuronal activity evoked in the secondary somatosensory cortex showing that the signal-to-noise ratio (SNR) of the setup needs to be further increased by at least a factor of 2.
△ Less
Submitted 26 June, 2019;
originally announced June 2019.
-
Demonstration of full tensor current density imaging using ultra-low field MRI
Authors:
P. Hömmen,
J. -H. Storm,
N. Höfner,
R. Körber
Abstract:
Direct imaging of impressed dc currents inside the head can provide valuable conductivity information, possibly improving electro-magnetic neuroimaging. Ultra-low field magnetic resonance imaging (ULF MRI) at $μ$T Larmor fields can be utilized for current density imaging (CDI). Here, a measurable impact of the magnetic field $B_{J}$, generated by the impressed current density $J$, on the MR signal…
▽ More
Direct imaging of impressed dc currents inside the head can provide valuable conductivity information, possibly improving electro-magnetic neuroimaging. Ultra-low field magnetic resonance imaging (ULF MRI) at $μ$T Larmor fields can be utilized for current density imaging (CDI). Here, a measurable impact of the magnetic field $B_{J}$, generated by the impressed current density $J$, on the MR signal is probed using specialized sequences. In contrast to high-field MRI, the full tensor of $B_{J}$ can be derived without rotation of the subject in the scanner, due to a larger flexibility in the sequence design. We present an ULF MRI setup based on a superconducting quantum nterference device (SQUID), which is operating at a noise level of 380 aT Hz$^{-1/2}$ and capable of switching all imaging fields within a pulse sequence. Thereby, the system enables zero-field encoding, where the full tensor of $B_{J}$ is probed in the absence of other magnetic fields. 3D CDI is demonstrated on phantoms with different geometries carrying currents of approximately 2 mA corresponding to current densities between 0.45 and 8 A/m$^{2}$. By comparison to an in vivo acquired head image, we provide insights to necessary improvements in signal-to-noise ratio.
△ Less
Submitted 8 May, 2019;
originally announced May 2019.
-
Ultra-sensitive SQUID instrumentation for MEG and NCI by ULF MRI
Authors:
Rainer Körber
Abstract:
The requirements for the construction of ultra-sensitive SQUID instrumentation as used in biomagnetism are presented. Typically, SQUIDs are inductively coupled to pick-up coils and for this arrangement one can improve the noise performance by increasing the sensing coil area. To achieve optimum sensitivity one has to consider the signal-to-noise ratio (SNR), which is an intricate interplay between…
▽ More
The requirements for the construction of ultra-sensitive SQUID instrumentation as used in biomagnetism are presented. Typically, SQUIDs are inductively coupled to pick-up coils and for this arrangement one can improve the noise performance by increasing the sensing coil area. To achieve optimum sensitivity one has to consider the signal-to-noise ratio (SNR), which is an intricate interplay between source characteristics and noise origin. It turns out that separate pick-up coil designs are needed for various noise characteristics even for an identical source. Hence, a hybrid system with differently sized pick-up coils presents the best option for multipurpose applications. A single channel system with close to SQUID intrinsic noise level is also described. This is possible by utilizing a special dewar design and thereby enabling a further increase in SNR. Such a system might be used for current density imaging and neuronal current imaging by ultra-low-field magnetic resonance where it also must be able to be field-tolerant to up to 100-200 mT.
△ Less
Submitted 22 March, 2019;
originally announced March 2019.
-
Ultra-sensitive SQUID systems for pulsed fields - Degaussing superconducting pick-up coils
Authors:
Eva Al-Dabbagh,
Jan-Hendrik Storm,
Rainer Körber
Abstract:
SQUID systems for ultra-low-field magnetic resonance (ULF MR) feature superconducting pick-up coils which must tolerate exposure to pulsed fields of up to 100 mT. Using type-II superconductor niobium (Nb) field distortions due to trapped vortices in the wire result. In addition, their rearrangement after quick removal of the pulsed field leads to excess low frequency noise which limits the signal-…
▽ More
SQUID systems for ultra-low-field magnetic resonance (ULF MR) feature superconducting pick-up coils which must tolerate exposure to pulsed fields of up to 100 mT. Using type-II superconductor niobium (Nb) field distortions due to trapped vortices in the wire result. In addition, their rearrangement after quick removal of the pulsed field leads to excess low frequency noise which limits the signal-to-noise ratio. In contrast, type I superconductors, such as lead (Pb), do not exhibit vortices but form an intermediate state with the coexistence of normal and superconducting domains. We measured the magnetization loops of superconducting wire samples of Nb and Pb together with their noise behavior after pulsed fields. Pb also exhibits significant excess low frequency noise once the wire has been driven into the intermediate state. To avoid this problem, we removed the field not abruptly but in a linearly decaying sinusoidal manner thereby degaussing the wire. After application of 57 mT, we found that Nb can be degaussed within at least 50 ms, the shortest time used in this study. Pb can also be degaussed, albeit within 100 ms and a more complex dependency on the degaussing parameters. After successful degaussing, negligible excess low frequency noise is observed.
△ Less
Submitted 23 January, 2018;
originally announced January 2018.
-
A modular, extendible and field-tolerant multichannel vector magnetometer based on current sensor SQUIDs
Authors:
Jan-Hendrik Storm,
Dietmar Drung,
Martin Burghoff,
Rainer Körber
Abstract:
We present the prototype module of our extendible and robust multichannel SQUID magnetometer system. A multi-module arrangement can be implemented by using up to 7 modules. It is intended for high-precision measurements of biomagnetism and spin precession. Further demanding applications are magnetorelaxometry and ultra-low-field nuclear magnetic resonance (ULF NMR), where pulsed fields of up to 10…
▽ More
We present the prototype module of our extendible and robust multichannel SQUID magnetometer system. A multi-module arrangement can be implemented by using up to 7 modules. It is intended for high-precision measurements of biomagnetism and spin precession. Further demanding applications are magnetorelaxometry and ultra-low-field nuclear magnetic resonance (ULF NMR), where pulsed fields of up to 100 mT are typically applied. The system is operated inside the Berlin Magnetically Shielded Room and equipped with 18 magnetometers consisting of niobium (Nb) wire-wound pick-up coils. A total of 16 small pick-up coils with 17.1 mm diameter form a regular grid with individual channels arranged to ensure system sensitivity covers all three orthogonal spatial directions. Two large hexagonal pick-up coils with an equivalent diameter of 74.5 mm sensitive in z-direction surround the grid at two different heights and are suitable for the detection of deep sources. Each coil is connected to the input of a thin-film Nb SQUID current sensor via a detachable superconducting contact. The SQUIDs are equipped with integrated input current limiters. Feedback into the pick-up coils minimizes crosstalk between channels. The current sensor chip package includes a superconducting shield of Nb. The field distortion of the prototype and a multi-module arrangement was analysed by numerical simulation. The measured noise of the small magnetometers was between 0.6 and 1.5 fT/Hz$^{1/2}$, and well below 1 fT/Hz$^{1/2}$ for the large ones. Using a software gradiometer, we achieved a minimum noise level of 0.54 fT/Hz$^{1/2}$. We performed ULF NMR experiments, verifying the system's robustness against pulsed fields, and magnetoencephalographgy (MEG) on somatosensory evoked neuronal activity. The low noise performance of our 18-channel prototype enabled the detection of high-frequency components at around 1 kHz by MEG.
△ Less
Submitted 28 August, 2017;
originally announced August 2017.
-
An ultra-sensitive and wideband magnetometer based on a superconducting quantum interference device
Authors:
Jan-Hendrik Storm,
Peter Hömmen,
Dietmar Drung,
Rainer Körber
Abstract:
The magnetic field noise in superconducting quantum interference devices (SQUIDs) used for biomagnetic research such as magnetoencephalography or ultra-low-field nuclear magnetic resonance is usually limited by instrumental dewar noise. We constructed a wideband, ultra-low noise system with a 45 mm diameter superconducting pick-up coil inductively coupled to a current sensor SQUID. Thermal noise i…
▽ More
The magnetic field noise in superconducting quantum interference devices (SQUIDs) used for biomagnetic research such as magnetoencephalography or ultra-low-field nuclear magnetic resonance is usually limited by instrumental dewar noise. We constructed a wideband, ultra-low noise system with a 45 mm diameter superconducting pick-up coil inductively coupled to a current sensor SQUID. Thermal noise in the liquid helium dewar is minimized by using aluminized polyester fabric as superinsulation and aluminum oxide strips as heat shields, respectively. With a magnetometer pick-up coil in the center of the Berlin magneti- cally shielded room 2 (BMSR2) a noise level of around 150 aT Hz$^{-1/2}$ is achieved in the white noise regime between about 20 kHz and the system bandwidth of about 2.5 MHz. At lower frequencies, the resolution is limited by magnetic field noise arising from the walls of the shielded room. Modeling the BMSR2 as a closed cube with continuous μ-metal walls we can quantitatively reproduce its measured field noise.
△ Less
Submitted 17 February, 2017;
originally announced February 2017.