Measurement of the scintillation resolution in liquid xenon and its impact for future segmented calorimeters
Authors:
C. Romo-Luque,
J. M. Benlloch-Rodríguez,
V. Herrero-Bosch,
N. Salor-Iguiñiz,
R. J. Aliaga,
V. Álvarez,
F. Ballester,
R. Esteve,
R. Gadea,
J. Generowicz,
A. Laing,
F. Monrabal,
M. Querol,
M. Rappaport,
J. Rodríguez,
J. Rodríguez-Ponce,
S. Teruel-Pardo,
J. F. Toledo,
R. Torres-Curado,
P. Ferrario,
J. J. Gómez-Cadenas
Abstract:
We report on a new measurement of the energy resolution that can be attained in liquid xenon when recording only the scintillation light. Our setup is optimised to maximise light collection, and uses state-of-the-art, high-PDE, VUV-sensitive silicon photomultipliers. We find a value of 4.2% +- 0.2% FWHM at 511 keV, a result much better than previous measurements and close to the Poissonian resolut…
▽ More
We report on a new measurement of the energy resolution that can be attained in liquid xenon when recording only the scintillation light. Our setup is optimised to maximise light collection, and uses state-of-the-art, high-PDE, VUV-sensitive silicon photomultipliers. We find a value of 4.2% +- 0.2% FWHM at 511 keV, a result much better than previous measurements and close to the Poissonian resolution that we expect in our setup (5.4% +- 0.8% FWHM at 511 keV). Our results are compatible with a null value of the intrinsic energy resolution in liquid xenon, with an upper bound of 0.4% FWHM at 95% CL at 511 keV, to be compared with 3--4% FWHM in the same region found by theoretical estimations which have been standing for the last twenty years. Our work opens new possibilities for apparatus based on liquid xenon and using scintillation only. In particular it suggests that modular scintillation detectors using liquid xenon can be very competitive as building blocks in segmented calorimeters, with applications to nuclear and particle physics as well as Positron Emission Tomography technology.
△ Less
Submitted 26 January, 2023; v1 submitted 14 October, 2022;
originally announced October 2022.
-
Monte Carlo characterization of PETALO, a full-body liquid xenon-based PET detector
Authors:
J. Renner,
C. Romo-Luque,
R. J. Aliaga,
V. Álvarez,
F. Ballester,
J. M. Benlloch-Rodríguez,
J. V. Carrión,
D. Cubero,
J. Díaz,
R. Esteve,
R. Gadea,
J. Gillam,
J. Generowicz,
J. L. López-Gómez,
A. Martínez,
F. Monrabal,
M. Querol,
M. Rappaport,
J. Rodríguez,
J. Rodríguez-Ponce,
P. Solevi,
S. Teruel-Pardo,
J. F. Toledo,
R. Torres-Curado,
V. Herrero-Bosch
, et al. (2 additional authors not shown)
Abstract:
New detector approaches in Positron Emission Tomography imaging will play an important role in reducing costs, lowering administered radiation doses, and improving overall performance. PETALO employs liquid xenon as the active scintillating medium and UV-sensitive silicon photomultipliers for scintillation readout. The scintillation time in liquid xenon is fast enough to register time-of-flight in…
▽ More
New detector approaches in Positron Emission Tomography imaging will play an important role in reducing costs, lowering administered radiation doses, and improving overall performance. PETALO employs liquid xenon as the active scintillating medium and UV-sensitive silicon photomultipliers for scintillation readout. The scintillation time in liquid xenon is fast enough to register time-of-flight information for each detected coincidence, and sufficient scintillation is produced with low enough fluctuations to obtain good energy resolution. The present simulation study examines a full-body-sized PETALO detector and evaluates its potential performance in PET image reconstruction.
△ Less
Submitted 11 January, 2023; v1 submitted 27 September, 2021;
originally announced September 2021.
-
PETALO read-out: A novel approach for data acquisition systems in PET applications
Authors:
V. Herrero-Bosch,
R. Gadea,
R. J. Aliaga,
J. Rodríguez,
J. F. Toledo,
R. Torres-Curado,
F. Ballester,
R. Esteve,
J. J. Gómez-Cadenas,
P. Ferrario
Abstract:
PETALO (a Positron Emission Tof Apparatus based on Liquid xenOn) is a new approach for Positron Emission Tomography scanners, based on liquid xenon. The PETALO detector aims at capturing the light produced by the scintillation in LXe taking advantage of its uniform response and continuity. This strategy will lead to a geometrical distortion free behavior compared to other PET detectors. To this en…
▽ More
PETALO (a Positron Emission Tof Apparatus based on Liquid xenOn) is a new approach for Positron Emission Tomography scanners, based on liquid xenon. The PETALO detector aims at capturing the light produced by the scintillation in LXe taking advantage of its uniform response and continuity. This strategy will lead to a geometrical distortion free behavior compared to other PET detectors. To this end, the sensors chosen for the light readout are SiPMs, which provide large area, high gain and very low noise. In order to take advantage of the unique PETALO detector characteristics a read-out architecture must be designed to meet the following specifications: Electronics associated to detector (front-end and read-out itself) must be fully expandable in terms of detector size. Read-out scheme must be compatible with the non-segmented structure of the detector. Time of Flight (TOF) capabilities must be assured at the scanner level, that is to say front-end and read-out electronics should not degrade LXe time performance. In these proceedings, a new readout concept is introduced which is compatible with a fully continuous medium detector such as PETALO. Results show that the system is feasible, with the introduction of a fast compression technique to further reduce the data bandwidth requirements.
△ Less
Submitted 1 March, 2021; v1 submitted 25 January, 2021;
originally announced January 2021.
-
Processing of Compton events in the PETALO readout system
Authors:
J. Renner,
J. M. Benlloch-Rodríguez,
J. V. Carrión,
R. Gadea,
V. Herrero-Bosch,
M. Kekic,
C. Romo-Luque,
P. Ferrario,
J. J. Gómez-Cadenas
Abstract:
PETALO (Positron Emission TOF Apparatus based on Liquid xenOn) exploits the unique characteristics of liquid xenon as a scintillator for use in a PET detector. Here initial simulation studies are detailed which highlight the potential of such a detector and outline the steps taken in the reconstruction of 511 keV gamma rays and in the full PET image reconstruction. In particular, a neural network-…
▽ More
PETALO (Positron Emission TOF Apparatus based on Liquid xenOn) exploits the unique characteristics of liquid xenon as a scintillator for use in a PET detector. Here initial simulation studies are detailed which highlight the potential of such a detector and outline the steps taken in the reconstruction of 511 keV gamma rays and in the full PET image reconstruction. In particular, a neural network-based approach is conceived in order to tag gamma rays that are poorly reconstructed due to Compton scattering. It is found that though a significant fraction of events undergo Compton scattering, not all of these events will necessarily be poorly reconstructed. Further study is necessary to determine whether or not an online implementation of the neural network-based event tagging will provide sufficient recovery of the image quality.
△ Less
Submitted 14 January, 2020;
originally announced January 2020.
-
PETALO: Time-of-Flight PET with liquid xenon
Authors:
Paola Ferrario,
Vicente Herrero-Bosch,
José María Benlloch-Rodríguez,
Carmen Romo-Luque,
Jun José Gómez-Cadenas
Abstract:
The fast scintillation decay time and the high scintillation yield of liquid xenon makes it an appropriate material for nuclear medicine. Moreover, being a continuous medium with a uniform response, liquid xenon allows one to avoid most of the geometrical distortions of conventional detectors based on scintillating crystals. In this paper, we describe how these properties have motivated the develo…
▽ More
The fast scintillation decay time and the high scintillation yield of liquid xenon makes it an appropriate material for nuclear medicine. Moreover, being a continuous medium with a uniform response, liquid xenon allows one to avoid most of the geometrical distortions of conventional detectors based on scintillating crystals. In this paper, we describe how these properties have motivated the development of a novel concept for positron emission tomography scanners with Time-Of-Flight measurement, which uses liquid xenon as a scintillating material and silicon photomultipliers as sensors. Monte Carlo studies have indicated that this technology would provide a very good intrinsic time resolution, of around 70 ps. Moreover, being liquid xenon transparent to UV and blue wavelengths, both scintillation and Cherenkov light can be exploited. While the former can be used for energy measurements, the latter is a prompt signal (of a few picoseconds), which provides a very precise time measurement. Monte Carlo simulations point to a time resolution of 30-50 ps obtained using Cherenkov light. A first prototype is being built to demonstrate the high energy, spatial and time resolution of this concept, using a ring of around 30 cm of internal diameter and a depth of 3 cm instrumented with VUV--sensitive silicon photomultipliers.
△ Less
Submitted 22 November, 2019;
originally announced November 2019.