Radar-based Respiratory Rate Monitoring in Standing Position
Authors:
Tassneem Helal,
Fady Aziz,
Omar Metwally,
Marco F. Huber,
Dominik Alscher,
Christoph Wasser,
Urs Schneider
Abstract:
Estimating human vital signs in a contactless non-invasive method using radar provides a convenient method in the medical field to conduct several health checkups easily and quickly. In addition to monitoring while sitting and slee**, the standing position has aroused interest for both the industrial and medical fields. However, it is more challenging due to the micro motions induced by the body…
▽ More
Estimating human vital signs in a contactless non-invasive method using radar provides a convenient method in the medical field to conduct several health checkups easily and quickly. In addition to monitoring while sitting and slee**, the standing position has aroused interest for both the industrial and medical fields. However, it is more challenging due to the micro motions induced by the body for balancing that may cause false respiratory rate estimation. In this work, we focus on the measurement of the respiratory rate of a standing person accurately with the capability of heavy breath detection and estimation. Multiple estimation approaches are presented and compared, including spectral estimation, deep-learning-based approaches, and adaptive peak selection with Kalman filtering. The latest technique is showing the best performance with an absolute error rate of 1.5 bpm, when compared to a Vernier Go Direct\textsuperscript{\textregistered} respiration belt.
△ Less
Submitted 13 June, 2022; v1 submitted 9 March, 2022;
originally announced March 2022.