Multi-Grid Detector for Neutron Spectroscopy: Results Obtained on Time-of-Flight Spectrometer CNCS
Authors:
M. Anastasopoulos,
R. Bebb,
K. Berry,
J. Birch,
T. Bryś,
J. -C. Buffet,
J. -F. Clergeau,
P. P. Deen,
G. Ehlers,
P. van Esch,
S. M. Everett,
B. Guerard,
R. Hall-Wilton,
K. Herwig,
L. Hultman,
C. Höglund,
I. Iruretagoiena,
F. Issa,
J. Jensen,
A. Khaplanov,
O. Kirstein,
I. Lopez-Higuera,
F. Piscitelli,
L. Robinson,
S. Schmidt
, et al. (1 additional authors not shown)
Abstract:
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demo…
▽ More
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demonstrator has an active area of 0.2 m$^2$. It is specifically tailored to the specifications of CNCS. The detector was installed in June 2016 and has operated since then, collecting neutron scattering data in parallel to the He-3 detectors of CNCS. In this paper, we present a comprehensive analysis of this data, in particular on instrument energy resolution, rate capability, background and relative efficiency. Stability, gamma-ray and fast neutron sensitivity have also been investigated. The effect of scattering in the detector components has been measured and provides input to comparison for Monte Carlo simulations. All data is presented in comparison to that measured by the $^3$He detectors simultaneously, showing that all features recorded by one detector are also recorded by the other. The energy resolution matches closely. We find that the Multi-Grid is able to match the data collected by $^3$He, and see an indication of a considerable advantage in the count rate capability. Based on these results, we are confident that the Multi-Grid detector will be capable of producing high quality scientific data on chopper spectrometers utilising the unprecedented neutron flux of the ESS.
△ Less
Submitted 3 April, 2017; v1 submitted 10 March, 2017;
originally announced March 2017.
Investigation of background in large-area neutron detectors due to alpha emission from impurities in aluminium
Authors:
J. Birch,
J. -C. Buffet,
J. -F. Clergeau,
P. van Esch,
M. Ferraton,
B. Guerard,
R. Hall-Wilton,
L. Hultman,
C. Höglund,
J. Jensen,
A. Khaplanov,
F. Piscitelli
Abstract:
Thermal neutron detector based on films of $^{10}$B$_4$C have been developed as an alternative to $^3$He detectors. In particular, The Multi-Grid detector concept is considered for future large area detectors for ESS and ILL instruments. An excellent signal-to-background ratio is essential to attain expected scientific results. Aluminium is the most natural material for the mechanical structure of…
▽ More
Thermal neutron detector based on films of $^{10}$B$_4$C have been developed as an alternative to $^3$He detectors. In particular, The Multi-Grid detector concept is considered for future large area detectors for ESS and ILL instruments. An excellent signal-to-background ratio is essential to attain expected scientific results. Aluminium is the most natural material for the mechanical structure of of the Multi-Grid detector and other similar concepts due to its mechanical and neutronic properties. Due to natural concentration of $α$ emitters, however, the background from $α$ particles misidentified as neutrons can be unacceptably high. We present our experience operating a detector prototype affected by this issue. Monte Carlo simulations have been used to confirm the background as $α$ particles. The issues have been addressed in the more recent implementations of the Multi-Grid detector by the use of purified aluminium as well as Ni-plating of standard aluminium. The result is the reduction in background by two orders of magnitude. A new large-area prototype has been built incorporating these modifications.
△ Less
Submitted 16 September, 2015; v1 submitted 2 July, 2015;
originally announced July 2015.